This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF) Free
Right arrow Letters to the Editor: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when Letters to the Editor are posted
Right arrow Alert me if a correction is posted
Services
Right arrow E-mail this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrow Rights and Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by SCHMALZRIED, T. P.
Right arrow Articles by CALLAGHAN, J. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by SCHMALZRIED, T. P.
Right arrow Articles by CALLAGHAN, J. J.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Facebook   Add to Technorati   Add to Twitter  
What's this?
The Journal of Bone and Joint Surgery 81:115-136 (1999)
© 1999 The Journal of Bone and Joint Surgery, Inc.


Current Concepts Review

Current Concepts Review - Wear in Total Hip and Knee Replacements*

THOMAS P. SCHMALZRIED, M.D.{dagger}, LOS ANGELES, CALIFORNIA and JOHN J. CALLAGHAN, M.D.{ddagger}, IOWA CITY, IOWA


    Introduction
 Top
 Introduction
 Wear Modes
 Friction and Frictional Torque
 Polyethylene Wear
 Periprosthetic Bone Loss
 The Effective Joint Space
 Overview
 References
 
Before the advent of total joint replacement, patients who had end-stage arthritis of the lower extremities had unremitting pain and a greatly decreased functional capacity. In addition, they often were confined to a wheelchair and were dependent on the care of others. Today, the outcomes of primary total hip and knee replacement are predictable and usually excellent35,45,57,64,212,221. Prosthetic joint replacement has dramatically improved the lives of millions of people worldwide.

As the fixation of total joint implants has become more reliable and durable and as the technology of total joint replacement has been applied to younger and more active patients, the current limitations of total joint arthroplasty are related to the wear of the components248. Wear is the removal of material, with the generation of wear particles, that occurs as a result of the relative motion between two opposing surfaces under load. In complex mechanical-biological systems such as total hip and knee replacements, there can be many types of wear. Although the mechanical consequences of wear, such as progressive thinning of polyethylene components, can limit the functional life of a joint replacement, the clinical problems from wear more frequently are due to the release of an excessive number of wear particles into a biological environment. When particles within a certain size-range are phagocytized in sufficient amounts, the macrophages enter into an activated state of metabolism, releasing substances that can result in periprosthetic bone resorption. Progressive loss of periprosthetic bone can necessitate a reoperation, which is the definitive measure of clinical failure of a joint arthroplasty.


    Wear Modes
 Top
 Introduction
 Wear Modes
 Friction and Frictional Torque
 Polyethylene Wear
 Periprosthetic Bone Loss
 The Effective Joint Space
 Overview
 References
 
It is important to distinguish among the fundamental mechanisms of wear (adhesion, abrasion, and fatigue); the changes in the appearance (the morphological characteristics) of the bearing surfaces, which are referred to as wear damage; and the conditions under which the prosthesis was functioning when the wear occurred, which have been termed the wear modes156. Adhesion involves bonding of the surfaces when they are pressed together under load. Sufficient relative motion results in material being pulled away from one or more surfaces, usually from the weaker material. Abrasion is a mechanical process wherein asperities on the harder surface cut and plow through the softer surface, resulting in removal of material. When local stresses exceed the fatigue strength of a material, that material then fails after a certain number of loading cycles, releasing material from the surface. One or more of the classic mechanisms of wear may be operating on the prosthesis in a particular wear mode, and a prosthesis may function in several wear modes over its in vivo service life. The predominant type of wear of one prosthetic joint can differ from that of another. Furthermore, in a specific joint, there may be different types of wear occurring at different times over the service life of the implant. The damage to an implant is a result of all of the mechanisms of wear that have acted on it over its service life, with the most recent events having the greatest influence.

Mode-1 wear results from the motion that is intended to occur between the two primary bearing surfaces, such as the motion of the prosthetic femoral against the polyethylene acetabular bearing surface. Mode-2 wear refers to the condition of a primary bearing surface that moves against a secondary surface that it is not intended to move against. For example, mode-2 wear occurs when a femoral component penetrates through a modular polyethylene bearing and rubs against the metallic tibial base-plate or acetabular shell8,22,67,131,148,231. Mode-3 wear refers to the condition of the primary surfaces as they move against each other but with the interposition of third-body particles. In mode-3 wear, the contaminant particles directly abrade one or both of the primary bearing surfaces. This type of wear is known as three-body abrasion or three-body wear. The primary bearing surfaces may be transiently or permanently roughened by this interaction.

Mode-4 wear refers to two secondary (nonprimary) surfaces rubbing together. Examples of mode-4 wear include impingement of the prosthetic femoral neck on the rim of the acetabular component249; motion at the stem-cement or bone-cement interface or relative motion of a porous coating, or other metallic surface, against bone2,6,30,90,115,164; relative motion of the external surface of a modular polyethylene component against the metal support (so-called backside wear); fretting between a metallic substrate and a fixation screw67,110,142,188; and fretting and corrosion of modular taper connections232 as well as that of extra-articular sources104,133,218. The particles that are produced by these types of wear may be composed of bone, polymethylmethacrylate, metal alloys, metallic corrosion products, or hydroxyapatite2,24,214,232. Wear particles produced by mode-4 wear can induce an inflammatory reaction and can be transported to the bearing surfaces and induce three-body wear (mode-3 wear)104,181,232.

Mode-1 wear is necessary for the joint prosthesis to function, whereas modes 2, 3, and 4 are generally unintended. The operating conditions in vivo are variable, and several types of wear can occur simultaneously. The clinical importance and interaction of wear modes may be clarified by a hypothetical example. A total hip or knee prosthesis that is well fixed and well functioning has a low rate of mode-1 wear of the polyethylene articular surface or surfaces. The gradual release of polyethylene particles into the periprosthetic tissues can result in a slow rate of interfacial bone resorption201, which can lead to an increase in the relative motion between the implant and the adjacent bone. Such relative motion causes mode-4 wear, which, depending on the type of joint prosthesis, can generate particles of bone, cement, metal, or hydroxyapatite. These hard particles can effect the baseline mode-1 wear by passing through the articulation and resulting in a transient three-body wear mechanism (mode-3 wear). The femoral component can be scratched by this interaction. Additionally or independently, hard particles may become embedded in the polyethylene and act as an ongoing abrasive source. The increased roughness of the bearing surface of the femoral component can then increase the rate of polyethylene wear in mode-1 because of the increased two-body abrasive wear42,64,66,153,171,238. This wear of the polyethylene can increase the rate of production of polyethylene particles, which increases the rate of periprosthetic bone resorption, leading to additional relative motion and eventually to the possible clinical failure of the implant. The time-course for such a sequence of events is variable. From a practical perspective, a problem with fixation leads to a prblem with wear and vice-versa.


    Friction and Frictional Torque
 Top
 Introduction
 Wear Modes
 Friction and Frictional Torque
 Polyethylene Wear
 Periprosthetic Bone Loss
 The Effective Joint Space
 Overview
 References
 
Friction is the resistance to movement between two surfaces in contact. The degree of resistance is proportional to the load. The ratio between frictional force and load (friction/load) is the coefficient of friction (µ)19. Frictional torque is the force created as a result of the friction of bearing. Charnley et al. initially selected a stainless steel-on-polytetrafluoroethylene bearing couple because of a low coefficient of friction39,43. The small, 22.25-millimeter-diameter head (subsequently referred to as a twenty-two-millimeter head) was selected to minimize the moment arm of the frictional forces and, thus, to minimize the frictional torque (coefficient of friction [µ x r2). Thick walls of polyethylene increased the outer diameter of the Charnley socket, which distributed the frictional torque over a large fixation area. Unfortunately, Charnley hip components with the polytetrafluoroethylene bearing uniformly failed because of rapid wear with the release of polytetrafluoroethylene waear particles, formation of granulomas, and loosening of the component86,238. Thus, despite a sound theoretical premise supported by laboratory investigations, this early clinical experience demonstrated the difficulty of predicting wear resistance in vivo as well as the adverse effects of wear particles.

Contrary to theoretical considerations, frictional torque has not been demonstrated to be important in the initiation of aseptic loosening of either femoral or acetabluar components161,207. Accumulating evidence indicates that wear particles have a far greater effect on the durability of the fixation of the implant than does frictional torque. From this perspective, the subsequent success of the Charnley low-friction arthroplasty with a polyethylene acetabular component is primarily a function of the low volumetric wear of the twenty-two-millimeter metal bearing, not of low frictional torque. Bearings with a larger diameter can be successful if the rate of wear is low161. This is an important consideration as methods to reduce polyethylene wear, such as extensive cross-linking of polyethylene or utilization of hard-on-hard bearings (such as ceramic-on-ceramic and metal-on-metal), are being investigated. The coefficient of friction for the metal-on-metal bearing of the McKee-Farrar hip prosthesis is roughly two to three times greater that that of the Charnley prosthesis. The larger diameter of the McKee-Farrar bearing couple, which is approximately forty millimeters, amplifies this difference, and as a result the frictional torque is as much as ten times greater that that of the Charnley bearing couple207. This value is stil only about one-tenth the static torque to failure of an acetabular component immediately after it has been freshly implanted with cement5,63,160,235.


    Polyethylene Wear
 Top
 Introduction
 Wear Modes
 Friction and Frictional Torque
 Polyethylene Wear
 Periprosthetic Bone Loss
 The Effective Joint Space
 Overview
 References
 
Most total hip replacements and effectively all total knee replacements have one primary bearing surface made of ultra-high molecular weight polyethylene (referred to simply as polyethylene). Although there are many potential sources of wear particles, in most total hip and knee replacements the greatest contribution is from mode-1 wear — that is, polyethylene particles generated by the intended motion of the joint at the primary bearing surfaces. For practical purposes, wear of the much harder bearing surface of the femoral component is negligible in mode 1. Polyethylene wear is distinct from creep, which is plastic deformation due to loading. Creep contributes to the deformation of a polyethylene bearing but does not produce wear particles. The rate of creep decreases rapidly over time, becoming negligible by the first twelve to eighteen months after implantation113,253. Wear accounts for most of the change in the surface of a polyethylene bearing over the longer term113,253.

There are a myriad of variables that affect the wear of a polyethylene bearing in vivo, including the wear resistance of the materials as well as the loads, lubrication, sliding distance, motion pattern, specifics of the design and manufacturing of the polyethylene component, implantation techniques, type of wear, and amount and type of use of the joint. The wear resistance of polyethylene is a function of the base resin, the manufacturing, and the method of sterilization of the polyethylene component48,143,223. Oxidation reduces the static strength and elongation properties of polyethylene and decreases the resistance of polyethylene bearings to fatigue, leading to higher in vivo rates of wear. A time-dependent increase in the amount of oxidation can result from gamma irradiation in air, which until recently was the most common method of sterilization in the orthopaedic industry. Gamma irradiation breaks molecular bonds in the long polyethylene chains, giving rise to free radicals. When present, oxygen can combine with these free radicals. Peak levels of oxidation typically occur about one to two millimeters below the surface of a polyethylene component. As the degree of this subsurface oxidation increases, so does the occurrence of fatigue cracking and delamination, as has been observed in retrieved tibial components46-48. Components that have been sterilized less than one year before implantation exhibit lower in vivo oxidation and better in vivo performance than components with a longer so-called shelf life before implantation52. In laboratory wear tests, polyethylene that had been sterilized with gamma irradiation in air and had been aged exhibited a higher rate of wear than material that had not been irradiated71,159.

When free radicals are formed in polyethylene, such as by gamma irradiation, cross-linking of polyethylene molecules is a competing reaction to oxidation. Cross-linking also changes the material properties of polyethylene, but it can improve the resistance to wear. The relative amount of oxidation and cross-linking varies with the depth from the surface of the component and results in a corresponding variation in the resistance of the material to wear as a function of the depth from the surface. In general, as oxidation increases, cross-linking decreases and vice versa159. Methods for controlled cross-linking include the use of chemicals (peroxide), variable-dose gamma irradiation, and electron-beam irradiation. Clinical studies have indicated a substantial reduction in wear associated with cross-linked polyethylene185,253. Recent studies with use of a wear simulator have indicated that, with optimum cross-linking, the type of wear that occurs in acetabular cups can be reduced by more than 95 percent120,159,236. However, cross-linking may not have the same degree of benefit with regard to the types of wear that occur in total knee replacement236.

Polyethylene contact stresses are a function of the thickness of the component as well as the load and the contact area12. A minimum thickness of six millimeters is recommended for a conforming articulation such as a total hip replacement. In general, from a materials standpoint, a thickness of more than six millimeters is preferred for bearings with less conformity13. Polyethylene wear is also a function of the motion pattern. In wear tests that use a linear motion path, such as a reciprocating pin-on-disk, the rate of polyethylene wear for a given set of test conditions is ten to 100 times lower than that in wear tests that use crossing motion paths151. Wear tests that use crossing motion paths have been shown to produce more closely the type and amount of wear occurring in vivo in total hip replacements156,191.

The clinical manifestations of polyethylene wear in total hip and total knee replacements include the removal of material, which results in progressive penetration of the femoral head into the polyethylene acetabular component and a reduction in the thickness of the polyethylene tibial bearing. A reoperation may be necessary because of thinning or wear-through of a polyethylene bearing or because of frank failure of the component8,22,67,231. Progressive penetration of the femoral head into the polyethylene acetabular component can lead to impingement between the femoral neck and the socket, which may contribute to loosening of the acetabular component43,249. Asymmetrical wear of a tibial polyethylene bearing can alter the mechanical axis of the knee and thereby increase the rate of wear in that compartment because of the increased load33,190,237.

Roughness of the Countersurface
When a joint prosthesis has one polyethylene bearing surface, the other bearing surface is commonly referred to as the countersurface. The base material and the specifics of manufacturing, such as the method of polishing, determine the initial (as-manufactured) surface characteristics of a femoral head or those of a femoral component of a total knee prosthesis. The microtopography of the surface determines its roughness. A contact or noncontact (laser) stylus can be used to scan the surface and make an analog recording of the peaks and valleys over a specified length of the surface. As a result of mode-3 wear, the original surface can be damaged in vivo, resulting in a rougher surface.

Increased roughness of the femoral countersurface may dramatically accelerate two-body abrasive wear of polyethylene61. Experimental studies have indicated that a threefold increase in the roughness of the femoral countersurface can cause at least a tenfold increase in the rate of polyethylene wear50,70. Polyethylene wear is sensitive to the specific type of surface damage. A scratch of two micrometers in depth with an average lip height of one micrometer may not substantially increase the measured average roughness of the surface, but in laboratory wear tests such scratches increased polyethylene wear thirty to seventyfold, depending on the motion pattern72. The combined effects of increased roughness of the countersurface and increased oxidation on wear may be synergistic rather than simply additive16.

Damage to the countersurface is common and was found on 89 percent of fifty-four cobalt-chromium-alloy femoral heads after a service life that ranged from eight months to nineteen years119. The affected areas were usually discrete, measuring from one to twenty square millimeters. Similarly, in an analysis of retrieved Charnley femoral components, 76 percent of seventy-one stainless-steel heads had an average surface roughness (Ra) that was greater than the specification of the as-manufactured surface112. Increased roughness of the countersurface of the femoral head has been associated with substantially higher rates of polyethylene wear in vivo, with the magnitude of the increase dependent on the nature of the damage to the countersurface89.

The mechanisms for generating an increased roughness of the countersurface in vivo apply to all materials, but the susceptibility to scratching is a function of the hardness of the material. The decreased hardness of titanium alloy, compared with stainless-steel and cobalt-based metal alloys, results in decreased resistance to abrasion. Although the initial surface roughness of a titanium-alloy femoral head may be equivalent to that of other bearing materials, there is a greater potential for its surface roughness to increase in vivo. In an environment with few or no hard third bodies, the wear performance of titanium alloy against polyethylene can be comparable with that of other metals152,153,158; however, the presence of hard third bodies, such as particles of cement or metal, has a greater adverse effect on the performance of titanium alloy against polyethylene148,153,181.

Ceramics are harder than stainless-steel and cobalt-based metal alloys and are therefore more resistant to damage by third-body particles than are metal countersurfaces49. For this reason, the increased hardness of ceramic materials is advantageous. Ceramic materials did not demonstrate a substantial advantage in a laboratory joint simulator in which there were few or no third bodies154,155. However, in clinical comparisons, in which the operating environment of the articulation is more variable, ceramic heads have demonstrated rates of wear that are lower than those of metallic heads62.

Polyethylene Wear Particles
Not until the 1990s was it recognized that most polyethylene wear particles are less than one micrometer in size and are produced in very large numbers, even by well functioning joints. Because the resolution of light microscopy is limited by the wavelength of visible light (0.4 to 0.7 micrometer), objects that are of submicrometer size cannot be clearly seen with a light microscope. With polarized light microscopy, submicrometer polyethylene particles appear as a fine, diffuse, background birefringence in the cytoplasm of macrophages and giant cells86,198. For this reason, polyethylene wear particles may have been present but not appreciated in some early studies of tissue from osteolytic lesions198. Techniques have been developed to analyze wear particles generated in vivo by retrieving them from periprosthetic tissues33,100,156,166,168,198,214. The concentration of wear particles from prosthetic joints can extend into the billions per gram of tissue100,166,168,230. As suggested by light microscopic observations, most of these are polyethylene wear particles166,214 that have linear dimensions that are less than one micrometer34,100,166,168,199,213,230.

Wear particles are a function of the type of wear that produces them87,205. For example, a smooth, polished femoral component moving against polyethylene in the absence of third bodies, as in a laboratory joint simulator or in an optimally functioning total joint replacement, produces a visually smooth, so-called burnished or polished-appearing polyethylene surface. This intended type of wear is the source of numerous submicrometer polyethylene wear particles156,198,201,205. There are case-to-case differences in the mode and the frequency distribution of these small particles in total hip replacements, and some data have indicated that the distribution is influenced by the type and amount of surface damage of the femoral head230.

Differences in the articulating surfaces and motion patterns of total knee replacements as compared with those of total hip replacements have important effects on the wear of the polyethylene. Decreased conformity can result in substantially increased contact stresses that can exceed the yield strength of polyethylene4,13,46. Furthermore, in a total knee replacement, the motion pattern can include rolling, sliding, and rotation on the same surface. The combination of these factors results in differences in the mechanisms of wear for total knee replacement compared with those for total hip replacement. In total hip replacement, the predominant wear mechanisms appear to involve microadhesion and microabrasion with the generation of many polyethylene particles less than one micrometer in length156. In contrast, subsurface delamination, pitting, and fatigue cracking, with the release of much larger particles of polyethylene, have been recognized as important mechanisms of wear in total knee replacement46,140,247. These mechanisms result in the visually striking surface damage of some retrieved tibial polyethylene bearings140,245.

More variety is seen in the size, shape, and texture of polyethylene particles associated with posterior-cruciate-retaining total knee replacements compared with those associated with total hip replacements100,105,205. Submicrometer granules are less prevalent in specimens obtained from the tissues around total knee replacements than in specimens taken from the tissues around total hip replacements. Larger flake-shaped particles, measuring several micrometers in length and width, are relatively common in association with total knee replacements but are not common in association with total hip replacements. The overall average area of particles generated by total knee replacements has been reported to be about twice that of particles generated by total hip replacements209. There are data indicating that specific features of the type of total knee prosthesis can influence the size of the polyethylene particles that are produced100. Furthermore, modular total hip and knee components have several surfaces, such as the so-called backside of the polyethylene, which can contribute polyethylene wear particles to the periprosthetic tissues. In one study, tissues adjacent to failed total knee replacements had more particles per gram than those adjacent to failed total hip replacements. However, the knee replacements had a longer service life and there was no difference in the apparent rate of particle deposition100.

Polyethylene Wear in Vivo
The clinical assessment of the wear of a polyethylene bearing has traditionally been based on radiographic studies. As measured on standard radiographs, the degree of penetration of the femoral component into the polyethylene component is the linear wear of the bearing. Because of practical issues related to the alignment and reproduction of x-ray-beam projection, it is more difficult to assess wear routinely on radiographs of total knee replacements than it is on radiographs of total hip replacements. For this reason, there is less information in the literature on the radiographic assessment of wear of total knee replacements45.

For total hip replacements, the method of radiographic measurement that is most commonly referenced is a variation of the duoradiographic technique originally described by Charnley and Halley42 and reported by Livermore et al.144. On standard anteroposterior radiographs of the pelvis, a compass is used to identify the shortest distance from the center of the femoral head to a reference point on the acetabular cup on the follow-up radiographs. A measurement is then made between the same reference points on the initial postoperative radiograph. After correction for magnification, the difference between the measurements on the initial postoperative and follow-up radiographs is the linear wear, which is conventionally expressed in millimeters. The linear wear rate is then calculated by dividing the linear wear by the duration of time after implantation. The linear wear rate is conventionally expressed as millimeters per year. This method only measures wear that occurs in the plane of the radiograph, and it cannot detect any component of the wear vector that occurs out of the plane of the radiograph. The maximum variation in the measurements made by a single observer with use of the Livermore technique was initially reported to be 0.1 millimeter144. However, other investigators have found the variability to be between three and four millimeters32. Electronic digital calipers can measure to the nearest 0.05 millimeter, and the mean intraobserver error with this device is reported to about 0.08 millimeter243.

A more accurate term for this measurement is linear penetration. The term linear wear has historically been used synonymously with the term linear penetration in discussions of polyethylene bearings. However, the measurement of linear penetration includes several factors in addition to wear, which is the removal of material, with the generation of wear particles, from the articulation of the femoral and acetabular components. Linear penetration includes creep (plastic deformation) and wear113. The contribution of creep to linear penetration is greatest in the early postoperative period and decreases with time, becoming negligible by twelve to eighteen months. Additionally, there is an initial running-in of the bearing, which results in better conformity, lower contact stresses, and a lower rate of wear12,13,253. In longer-term studies, decreasing patient activity over time can result in decreased wear; also, there may be a survival selection for the better-functioning implants (those with lower wear rates)210. For these reasons, short-term rates of linear penetration are higher than long-term rates226,253. Modularity is another factor that can contribute to linear penetration. Because of issues related to creep, manufacturing tolerances, and the in vivo assembly of modular acetabular components, the acetabular liner may change position relative to the acetabular shell, resulting in a change in the position of the femoral head and an increase in the short-term rate of linear penetration226. So-called backside wear of the modular polyethylene liner could also contribute to higher rates of linear penetration87,110.

Volumetric wear is a measure of the amount of material removed from the bearing surface. The simple formula v = {pi}r2w, in which v is the change in the volume of the polyethylene bearing, r is the radius of the femoral head, and w is the measured linear wear, has commonly been used to calculate volumetric wear144. This formula assumes a single, cylindrical wear track, but this assumption is not supported by some retrieval studies254.

In the 1990s, techniques for computer-assisted measurement of wear have been developed54,169,216,227. Standard radiographs can be digitized to create a computer model of the femoral head and the acetabular component. Use of both anteroposterior and lateral radiographs allows construction of a three-dimensional model. Comparison of serial radiographs gives both the magnitude and the direction of the femoral head displacement. Such computer-assisted techniques can reduce measurement variability that is due to difficulties related to the identification of a single reference point, the angle of the radiographic beam, and positioning of the patient. Edge-detection techniques that infer the margins of the components by evaluating gray-scale intensity on digitized images of radiographs have been developed. These techniques minimize the potential for intraobserver and interobserver variability and should enable more accurate comparisons of results between institutions169,216.

Assuming a negligible contribution by creep in the long term, the wear of retrieved polyethylene acetabular components has commonly been measured with variations of the so-called shadowgraph technique132,249. With this technique, a cast of the acetabular bearing surface is made and the profile (the so-called shadow) of the cast is used to measure the wear track. This method allows determination of the angle of wear relative to the mouth of the component so that a wear vector (magnitude and direction) can be determined. Single-plane radiographic measurements have slightly underestimated the linear wear measured on the corresponding retrieved implants. This may be due, at least in part, to the fact that the wear vector is not consistently in the plane of the radiographs. Additionally, the wear track is not a complete cylinder. Volumetric wear calculated from linear wear with use of a formula for a hemicylinder with a correction factor for the direction of wear is about 47 percent less than that based on the simple cylindrical formula132. Multiple wear vectors have been identified in 30 percent of one type of polyethylene acetabular component. The vectors do not appear to be a result of loosening of the cup, but impingement between the edge of the cup and the neck of the femoral component may be a factor254.

Fluid-displacement methods have also been used to measure the wear of retrieved polyethylene acetabular components121,225. A femoral head of appropriate size is placed into the original articulating contour, and the volume of fluid that is required to fill the remaining contour (the worn area) is measured. The interobserver variability is reportedly within five cubic millimeters, and the accuracy is reportedly within fifteen cubic millimeters121.

The amount of wear of hard-on-hard bearing surfaces, such as metal-on-metal or ceramic-on-ceramic, is typically so small that it cannot be measured on routine clinical radiographs. Furthermore, the amount of wear can also be too small to be measured on retrieved specimens with use of the shadowgraph technique207. Consequently, computerized coordinate measuring devices have been employed to quantify the amount of wear. These devices can be used to assess the sphericity of the bearing surfaces by comparing the measured dimensions in multiple planes with the best-fit sphere. The volume of wear can be calculated by integrating the depth of multiple individual wear points on the worn surface157.

Studies of Wear in Vivo
The rates of polyethylene wear in total hip replacements have been reported in a number of previous studies10,11,32,36,39-41,55,56,84,98,104,112,121,132,144,145,180,183,184,201,216,225-227,243,249,250,252,253. (The findings of those studies were summarized in table form by one of us [T. P. S.] and colleagues in a previous issue of The Journal of Bone and Joint Surgery211.) Many variables influence polyethylene wear in vivo and, consequently, the rates are highly variable. These studies have demonstrated substantial variability not only with regard to the average rates of linear wear but also with regard to the ranges (when reported). Regardless of the duration of follow-up, a number of hip replacements have no radiographically measurable wear and some demonstrate wear that is several times the average for the study. These large patient-to-patient variations in rates of wear have not been explained by differences in the wear resistance of the polyethylene239. This is not surprising considering the number of variables that contribute to wear in vivo.

Patient-related variables include age, gender, weight, general health, and activity as it relates to the use of the hip prosthesis210. Variables related to the hip reconstruction include the implanted materials (including, but not limited to, the polyethylene bearing material); the design and manufacturing of the prosthesis; and the characteristics of the implantation procedure, such as the operative techniques, biomechanical considerations, and the initial as well as the long-term fixation of the implants. These variables are important with respect to wear as they can affect the loads and the motions of the bearing and the degree of three-body-wear mechanisms. There is also variability in wear measurements because of differences in the methods of assessment and limitations of the measurement techniques, as already discussed. For these reasons, the strength of comparisons made between different studies is limited. Furthermore, in cohort studies, the rates of wear do not follow a Gaussian distribution, and appropriate methods should be used in statistical analysis to account for these distributions186.

In one study, in which serial radiographs of each hip replacement were analyzed and compared over a minimum of five years, the rate of penetration of the femoral head into the polyethylene liner tended to decrease with time, reaching a steady state after the sixth postoperative year226. Despite substantial differences in the rates of penetration over the first three years in subsets of hip replacements with high and low rates of penetration, the rates became similar over time. These results indicate that, for individual patients and prosthetic hip systems, multiple assessments of wear over time are more valuable than a single measurement226 and that caution should be exercised when comparing rates of penetration in hip replacements after different durations of follow-up113.

Theoretical models and retrieval analyses have shown that the rate of volumetric wear of polyethylene components increases with an increase in the diameter of the femoral head. This is because of an increase in the contact surface and because, during the same gait cycle, the sliding distance (the motion of the surface of the femoral head relative to the surface of the socket) increases with an increase in the size of the bearing171. With use of the simple cylindrical formula, v = {pi}r2w, it is shown that for any given amount of linear wear the volumetric wear increases exponentially with increases in the radius of the bearing. In one retrieval study, for each millimeter increase in the diameter of the head, there was an increase of 6.3 cubic millimeters per year in volumetric wear132. In another report, the rate of volumetric wear increased from 7.5 to 10 percent for each millimeter increase in the diameter of the head121.

In clinical studies, the linear wear rates of polyethylene against thirty-two-millimeter-diameter femoral heads have been equivalent to or greater than those of polyethylene against heads with a smaller diameter115,144, and they have been associated with bone resorption and loosening of the acetabular component175,201,204. Strictly with respect to wear, heads that have a thirty-two-millimeter diameter do not compare favorably with heads that have a smaller diameter. Because of their large diameters, surface-replacement components have rates of volumetric polyethylene wear that are four to ten times higher than those of conventional total hip replacements with twenty-eight-millimeter heads132. It had been hoped that the reduction in polyethylene stresses due to the larger contact area would result in reduced linear wear, but this has not been the case. Within the range of contact areas for metal-on-polyethylene prosthetic hip bearings, it appears that the stresses in these conforming bearings are all so low that any differences do not appreciably affect wear in vivo. The increased contact area and sliding distance of larger heads result in increased volumetric wear. The relationship between the size of the head and the rate of wear may be confounded by the use of a relatively thin polyethylene component with some thirty-two-millimeter bearings and in surface replacements.

The association between volumetric wear and periprosthetic bone resorption is related to the number of polyethylene wear particles that have been released into the so-called effective joint space, which includes all periprosthetic spaces and tissues in communication with joint fluid198. A twenty-eight-millimeter-diameter bearing with a conservative linear wear rate of 0.05 millimeter per year (a volumetric wear rate of about thirty cubic millimeters), with individual wear particles equal in volume to a 0.5-micrometer-diameter sphere, generates a total of 500 billion particles. Assuming that a patient takes one million steps per year, this translates to 500,000 particles per step. Such estimates are very sensitive to particle size. The number of particles produced by a given volume of wear varies with the cube of the diameter of the particle. A single ten-micrometer-diameter spherical particle has the same volume as 8000 0.5-micrometer-diameter particles. There would be only about sixty-three million ten-micrometer-diameter particles for the same amount of volumetric wear (thirty cubic millimeters). At one million steps per year, this translates to only sixty-three wear particles per step156.

Studies of wear particles retrieved from periprosthetic tissues and analyses of worn polyethylene surfaces have demonstrated findings that are consistent with an average particle size in the 0.5-micrometer-diameter range34,100,156,166,168,205,214,230. If the size of the wear particles is constant, then increases in volumetric wear lead to increased numbers of polyethylene wear particles. From a combined mechanical and biological perspective, optimization of in vivo wear requires not only a reduction of wear volume but also a reduction in the generation of the most biologically active wear particles. A lower rate of wear may not necessarily be preferred clinically if a higher number of biologically active wear particles is generated.

Compared with the assessment of wear on the hemispherical bearing of a total hip replacement, the assessment of wear in a total knee replacement is more complex because of differences in the geometry of the numerous femoral and tibial components and because of differences in kinematics. Studies have indicated that the rates of polyethylene wear in total knee replacements are even more variable than the rates of wear in total hip replacements and that they are a function of design, including the conformity of the articulation7,190,245; operative technique, such as the mechanical alignment of the knee and the fixation of the components33,190,237; the presence of abrasive third bodies103,140; the thickness of the polyethylene component; and the polyethylene manufacturing process13,67,143,245,246.

The geometry of an articulation can be generally described as a convex surface on a concave surface. In total knee replacement, the degree of conformity between the two surfaces can be described as the ratio of the radius of curvature of the tibial component (R2) to the radius of curvature of the femoral component (R1), or R2/R1. Such an analysis can be done for both the sagittal and the coronal plane geometry. As the value of this ratio approaches one, the conformity of the articulation increases. Thus, the most conforming articulation (such as in a total hip replacement or a flat-on-flat surface) would have matched radii and a ratio of one. As R2 becomes larger than R1, the conformity decreases and contact stress increases. For a metal-on-polyethylene articulation, the contact stress is roughly doubled when this ratio increases from one to five220. Because the radius of curvature of a femoral component may not be constant over the entire sagittal profile, the contact stresses for the same load may vary throughout the range of motion of the knee. Constraint is distinct from conformity. Constraint is the restriction of motion. A flat-on-flat articulation is completely conforming and has no constraint to motion in any plane, whereas a dished articulation of matched radii is completely conforming but motion is constrained to one plane.

Condylar designs with a conforming tibiofemoral articulation have large contact areas, lower contact stresses, and more favorable wear characteristics45,111, but they may not allow physiological translational and rotational movements. Relatively flat tibial articulations can accommodate such motions, but they have smaller contact areas, higher contact stresses, and higher rates of wear46,80,81,95. Experience indicates that, when an attempt is made to maximize function through design innovations, priority must be given to the material limitations of the prosthetic implants18. The apparent dilemma of balancing the goals of conformity and multidirectional motion is addressed by designs with mobile tibial polyethylene bearings. These designs have a high degree of tibiofemoral conformity and allow for rotation (a so-called rotating platform) or rotation combined with anterior-posterior translation (so-called meniscal bearings)7,25-27,80. Dislocation of the mobile bearings was reported in 2 percent (three) of 123 knees and was associated with instability in flexion and with revision total knee replacement25.

Rather than quantifying a change in dimensions, as is done in assessments of wear in total hip replacements, most investigators who have performed retrieval studies of total knee components have analyzed the relative amounts of several types of surface damage, including pitting, scratching, abrasion, burnishing, delamination, and embedded third-body abrasive particles103,140. The degree of surface damage is related to the weight of the patient and the duration of time after implantation103,245. In total knee replacements, peak stresses occur at about one to two millimeters below the surface. The combination of high subsurface stresses and decreased mechanical properties due to oxidation at roughly this same depth predisposes to delamination, which is a frequent type of surface damage in tibial bearings with lower conformity47,140,245 and is associated with rapid rates of wear66,67,131,172,231.

Pitting was the most common type of surface damage seen in association with a conforming total condylar design; only two of forty-eight components showed evidence of delamination103. The highest scores for damage were in the central, weight-bearing portions of the medial and lateral tibial concavities. There was greater damage on the anterior aspect than on the posterior aspect, which is probably due to the kinematics of this design. No delamination was seen in the standard tibial polyethylene components of twenty retrieved posterior cruciate-substituting total knee replacements. Burnishing was the most common type of surface damage reported for patellar components103. It was roughly equal on the medial and lateral aspects of the patellar component, where contact occurred with the femoral component, and the severity of the damage was related to the weight of the patient. In metal-backed patellar components, the polyethylene may be quite thin and demonstrate rapid wear, fracture, or dissociation from the metal backing147.

In a retrieval study190 of a posterior cruciate ligament-sacrificing total knee replacement with a high degree of tibiofemoral conformity in the sagittal plane and with polyethylene that had a minimum thickness of six millimeters, the rate of linear penetration of the femoral component averaged 0.025 millimeter per year, which is considerably lower than that reported for the Charnley total hip prosthesis249,250,252,253. The calculated rate of volumetric wear was sixteen cubic millimeters per year, which is also lower than that reported for the Charnley total hip prosthesis245,249,252,253. The relative amount of penetration on the medial side was increased in varus knees and decreased in valgus knees. Symmetrical wear was most likely to occur when the knee was in about 5 degrees of tibiofemoral valgus. In a study of one design of a so-called meniscal-bearing knee replacement, which has a high degree of tibiofemoral conformity in the sagittal plane, bearings retrieved at reoperation because of loosening or dislocation had an average rate of penetration of only 0.026 millimeter per year, which is quite low7. The contact area available for load transmission in each tibiofemoral compartment was 5.7 square centimeters, which is larger than that of the Charnley hip prosthesis. Such low rates of polyethylene wear are consistent with a linear motion path.

Higher rates of wear in total knee replacements have been associated with some designs characterized by lower conformity, especially when the tibial polyethylene is less than six millimeters thick46,245 and when so-called heat-pressing of the tibial articular surface has been used131. Younger, more active, and larger (predominantly male) patients are at risk for accelerated wear in association with these prosthetic variables67,134,172,231,237. This type of rapid wear tends to involve only the medial or lateral compartment, suggesting a role of asymmetrical loading and high localized stresses from a suboptimum mechanical environment. The rate of progression of the femoral component into the polyethylene can be dramatic, as much as one millimeter per year, resulting in wear-through of a four-millimeter-thick component in four years66,131. The type of wear that occurs is predominantly due to subsurface fatigue resulting in delamination (and the production of larger particles of polyethylene)205. Wear-through or gross mechanical failure of the polyethylene component can result in mode-2 wear with abrasive damage to the femoral component and substantial generation of metal particles.

The clinical triad of effusion, pain, and progressive change in the coronal alignment of the knee (most commonly into varus alignment) is characteristic of accelerated polyethylene wear131. The joint fluid is laden with polyethylene particles of various sizes, and aspiration can confirm the diagnosis of polyethylene-induced synovitis23,88. Arthroscopy can be helpful in assessing the degree of polyethylene wear and the damage of the femoral component as well as in planning a revision operation. Arthroscopic débridement may provide temporary relief, but most patients who have this triad need a revision172.

In clinical studies, wear rates generally have been retrospectively compared on the basis of a specific variable, such as the type of femoral component or the type of fixation (with or without cement). The tremendous number of potentially confounding variables in such clinical studies is a fundamental limitation to this type of comparison. Furthermore, great care should be taken when extrapolating the results to other reconstructions with the same generic variable. An example is the issue of metal-backed acetabular components. One specific type of metal-backed acetabular component inserted with cement has been reported to have an increased rate of polyethylene wear compared with an all-polyethylene component inserted with cement36. However, a different type of metal-backed acetabular component has been reported to be associated with a lower rate of polyethylene wear in other studies32,104. The issue cannot be as simple as the presence or absence of metal backing. Specific details of the design, materials, and manufacturing of the components as well as any differences in the populations of patients, durations of follow-up, measurement techniques, and statistical analyses all must be carefully considered when evaluating and comparing rates of polyethylene wear in vivo.

Clinical rates of wear traditionally have been expressed with use of a denominator of time because of convenience, not accuracy. More appropriately, investigators performing in vitro studies involving laboratory wear simulators have always used the number of loading cycles as a denominator. Similar to the use of a set of automobile tires, the wear of a prosthetic hip or knee is a function of use or the number of cycles and not a function of time in situ. The assumption made in clinical studies is that all patients with a joint replacement have about the same level of activity—that is, the actual use of or the number of cycles on the bearing is about the same—or that any differences average out over a large sample size. The limitations of this assumption must be recognized.

One of us (T. P. S.) and colleagues studied, with use of an electronic digital pedometer, the walking activity of 111 patients who had a total joint replacement210. We calculated an average of about 0.9 million cycles for each joint in the lower extremity per year. This number is close to the average of 1.0 million cycles per year proposed by Seedhom et al.213 on the basis of their study of the gait cycles of nine elderly individuals, on vacation, who did not have a joint prosthesis. The most important finding of our study, however, was not the average but the fact that there was a forty-fivefold difference in the range of gait cycles between the least active and the most active individual. The most active individual averaged 3.2 million cycles per year, about 3.6 times higher than the average. These data indicate that individual differences in activity are a substantial source of variability in rates of wear. A forty-fivefold difference in rates of wear as well as rates that are more than 3.5 times the average can be accounted for by differences in an individual's activity. Age was associated with daily walking activity but with a high degree of variability (the standard deviation was 3040 steps per day). Individuals who were less than sixty years of age walked about 30 percent more on the average than those who were sixty years of age or older (p = 0.023). The men walked 28 percent more on the average than the women (p = 0.037), and the men who were less than sixty years old walked 40 percent more on the average than the other individuals (p = 0.011). Thus, the variation in an individual's activity contributes to the variability in rates of wear that is consistently seen in in vivo studies. These results should be considered when analyzing studies of in vivo wear rates that have a denominator of time in situ.


    Periprosthetic Bone Loss
 Top
 Introduction
 Wear Modes
 Friction and Frictional Torque
 Polyethylene Wear
 Periprosthetic Bone Loss
 The Effective Joint Space
 Overview
 References
 
Periprosthetic bone loss can occur as a result of a reduction in the load transmitted to bone, so-called stress-shielding. Periprosthetic bone loss also occurs as a result of an inflammatory reaction to small particles, such as those produced by the various wear modes. To varying degrees, both processes occur simultaneously in complex mechanical-biological systems such as joint replacements, and the adverse effects can be additive. Bone with decreased density secondary to stress-shielding may be more susceptible to osteolysis64. Relative motion between an implant and bone can cause bone loss through both mechanical and biological mechanisms79.

Cellular Mechanisms
The tissue adjacent to total hip and knee prostheses consists of synovial tissue, variably organized and variably vascularized fibrous tissue, lymphocytes (occasionally), and foreign-body inflammatory cells (macrophages and giant cells) that are present roughly in proportion to the number of small particles79,105,173,198. Prosthetic particles elicit a cascade of responses at the cellular and tissue levels. The cell whose function is central to the biological reaction to prosthetic wear particles appears to be the macrophage1,196,197. The mononuclear stem cell, which originates in the bone marrow, is the progenitor for both mononuclear macrophages and osteoclasts. Macrophages phagocytose small wear particles and may fuse to form foreign-body multinucleated giant cells, usually in association with larger particles. Osteoblasts and fibroblasts also may be important in the response to wear particles resulting in altered formation of bone and connective tissue126. Although lymphocytes are occasionally present, their role in the inflammatory reaction is unclear.

Most periprosthetic bone resorption is effected by osteoclasts, but there is evidence that macrophages and foreign-body giant cells are capable of direct, low-grade bone resorption9,102,177. In vitro studies have indicated that activated macrophages release cytokines, including interleukins and prostaglandins, which play a role in the recruitment and differentiation of cells and stimulate bone resorption178, but the specificity of the cytokine response and the regulatory mechanisms have not been defined. Under certain conditions, macrophages appear to directly release interleukin-1 beta and tumor necrosis factor. Several cytokines, including interleukin-1 beta, stimulate osteoclast maturation. Although cytokines released by macrophages may directly stimulate bone resorption by osteoclasts, other effects may be mediated by intermediary cells such as fibroblasts or osteoblasts106,126,248. Matrix metalloproteinases (collagenase, gelatinase, and stromelysin), which are capable of effecting bone resorption, are also produced by interfacial membrane tissue around failed total hip and knee replacements44,59,79,105. On the basis of the knowledge of such cellular and biochemical mechanisms of bone resorption, there has been increasing interest in and investigation of pharmacological agents that may modify these cellular responses20,94,107,215. Other work indicates that, in addition to bone resorption, there is also a decrease in bone formation in association with periprosthetic osteolysis82,255.

It appears that all of the materials used in total joint replacement are capable of inducing an inflammatory foreign-body reaction if the particles are within a certain size-range and there are enough of them. The bone-resorbing ability of macrophages in vitro is a function of the size, shape, and composition of the particles, and it is dose-dependent214. It has been previously recognized that there is an upper size limit for particle reactivity106, but there may also be a lower size limit. For a given concentration of particles, the stimulatory effect of polyethylene particles in vitro decreases when the particles are larger than about seven micrometers or smaller than about 0.2 micrometer83. This information suggests that polyethylene wear particles generated by current designs of total hip and knee replacements are especially apt to cause problems because they are produced in very high numbers and are predominantly of a size within the range of peak biological activity. This also suggests that the biological effects of the very small wear particles (linear dimensions measured in nanometers) produced by metal-on-metal articulations may be different than those of polyethylene wear particles58. In aggregate, these very small particles have a large surface area and the systemic levels of metal ions are elevated in patients who have metal-on-metal articulations. The clinical importance of this is not known114.

Osteolysis
The classic descriptions of osteolysis adjacent to total joint replacements have been based on radiographs and have referred to a nonlinear, scalloped, or erosive form of femoral endosteal bone resorption in association with total hip replacements inserted with cement38,90. The term osteolysis has subsequently been used generically in reference to bone resorption that occurs in association with a foreign-body response to particles from a prosthetic joint3. Although there may be similarities at the cellular level, the pattern, location, extent, and rate of progression of inflammatory periprosthetic bone resorption are variable because of specific differences in the anatomy and physiology of the joint, the operative procedure, and the prosthetic implants.

The evolution of the understanding of the anatomical, mechanical, and biological components of osteolysis associated with total joint replacement has been one of the more fascinating sagas in the history of orthopaedic surgery and a topic of controversy and confusion. It should be acknowledged that the conclusions of retrieval analyses are a function of the material available for study. Furthermore, the questions to be answered determine the methods with which the specimens are processed, which influence the observations and conclusions that can be made. In addition, even the most complete retrieval analysis can describe only one moment in time of a complex mechanical and biological process for an individual specimen. An appreciation of these limitations may at least partially explain the apparent differences in observations that have been reported in various studies.

Osteolysis in Association with Total Hip Replacement
In 1968, Charnley et al. reported a pattern of nonlinear endosteal erosion in association with the Charnley total hip prosthesis inserted with cement and they suspected that this type of localized bone loss was due to infection38. Charnley also observed cystic erosion of bone in the femoral diaphysis in association with fracture of the stem and believed that a deficient cement mantle was probably the main cause of these fatigue fractures41. The tissue was reported to contain fragments of polymethylmethacrylate, but polyethylene particles were not identified. In 1976, Harris et al. reported a similar pattern of nonlinear localized bone resorption (osteolysis) in the proximal part of the femur around loose total hip replacements90. Lesional tissues were characterized by osteoclastic bone resorption, a high concentration of macrophages, and occasional foreign-body giant cells, with spherical spaces of variable size that were consistent with particles of polymethylmethacrylate. This appears to be the origin of the concept of so-called cement disease97,106,130. Jasty et al. subsequently reported osteolysis in association with polymethylmethacrylate particles adjacent to well fixed total hip replacements that had been inserted with cement115. Polyethylene wear particles were not identified. It was not until endosteal osteolysis was identified around femoral components inserted without cement163 that polyethylene, metals, and other materials in particulate form were seriously considered as potential contributors to the inflammatory reaction and associated bone resorption.

In 1990, Anthony et al.6 demonstrated a communication between the articulation and the endosteal surface of the femur, through a space between the stem and the cement (as described by Fornasier and Cameron73), and then through a defect in the cement mantle. Particles of metal, cement, and polyethylene were identified in macrophages in these osteolytic lesions. Arthrography of the hip, performed after the patients had exercised the hip joint, demonstrated transfer of contrast material from the femoral-acetabular articular space to the area of the osteolysis. Studies116,118,162 of well functioning total hip replacements retrieved at autopsy have indicated that some degree of separation of the femoral stem from the cement mantle, so-called debonding, occurs frequently and as early as two weeks after implantation. Fractures of the cement mantle were seen in association with debonding and had originated at the metal-cement interface, most often at the corners of the mantle or in association with pores in the cement. Cement fractures most frequently occurred in areas where the cement was thin or in association with frank defects in the mantle.

A common finding in patients who have femoral endosteal osteolysis around a cemented component appears to be a defect in the cement mantle. The risk of osteolysis can be decreased when technical errors are avoided92,176. Separation of the stem from the cement mantle does not lead to localized endosteal osteolysis if the cement mantle remains intact, and such separation appears to be compatible with satisfactory long-term function of certain types of femoral components77,212 but not others174. The relationships between osteolysis and the geometry of the stem, the surface finish of the stem, and the operative techniques have continued to be investigated91,149,174,243,244.

Compared with femoral components that have been inserted with cement, femoral components that have been inserted without cement and that have a limited proximal porous coating can be associated with earlier, more frequent localized endosteal bone resorption in the femoral diaphysis92,163,229. The size of the lesions can increase with time as can the prevalence, which has exceeded 30 percent in some reports92,163,229. This appears to be due to the fact that a limited proximal porous coating, especially if it is not circumferential, allows joint fluid and wear particles relatively easy access to the diaphyseal endosteum21,198,233. Diaphyseal endosteal osteolysis has not been reported in association with extensive, circumferential, porous coating even in the presence of substantial stress-shielding28,64.

Inflammatory bone resorption (induced by small particles) also occurs in a more linear pattern, which may progress along the cement-bone interface and cause or contribute to loosening of the implant, especially in the acetabulum101,203,256. There may be mechanical factors that contribute to interfacial particle transport. Once stability has been lost, additional motion can only be detrimental203. Even with the relatively low rate of volumetric wear of the Charnley total hip prosthesis, accumulated generation of wear particles eventually decreases the stability of the acetabular implant. A highly significant association (p < 0.01) has been demonstrated between the depth or rate of polyethylene wear and loosening of the acetabular component75,137,250-252. A high rate of polyethylene wear precludes a long service life for the implant112.

A similar process of linear osteolysis can occur in association with acetabular components inserted without cement188,198-200,202,204,206. The integrity of the peripheral implant-bone interface governs the ingress of joint fluid and wear particles. Progressive radiolucent lines are more likely to develop adjacent to components with initial peripheral interface gaps than they are likely to develop adjacent to components without such initial peripheral gaps199. Tissues from the peripheral acetabular implant-bone interface contain macrophages laden with small polyethylene wear particles198. Insertion of acetabular components with a tight peripheral press-fit has led to a reduced prevalence of initial peripheral interface gaps, a reduced prevalence of progressive peripheral interface radiolucency, and a higher prevalence of hip replacements without any interface radiolucency206. Regardless of other design features, a similar process of inflammatory bone resorption can occur in the proximal part of the femur after hip replacement with or without cement85,128,129,198,203. This process can cause or contribute to the loosening of femoral components inserted with cement79. Inflammatory bone resorption may play a role in loosening, after previous rigid fixation by bone ingrowth, of a component inserted without cement and with a limited proximal porous coating117. This same process has not resulted in loosening of a component with an extensive porous coating64.

Acetabular components inserted without cement have a lower prevalence of interface radiolucency than components inserted with cement199,206,242. However, the patterns of bone resorption around acetabular components inserted without cement are different than those seen around components inserted with cement. The bone loss associated with acetabular components inserted with cement typically occurs predominantly along the interface, following the contours of the cement mantle. The bone loss associated with acetabular components inserted without cement typically progresses away from the interface into the cancellous bone of the pelvis, resulting in localized bone resorption, the classic nonlinear or expansile form of osteolysis165,196,204,256. Although less common, this form of pelvic osteolysis can also occur in association with acetabular components inserted with cement43,189. Pelvic osteolysis is associated with a younger age256, vertical positioning of the acetabular component, and high volumetric wear of the polyethylene8,204. Substantial bone resorption around well fixed components may be asymptomatic until a pelvic fracture occurs. For this reason, annual evaluations with radiographs made in multiple planes are recommended167,204,256.

Other concerns associated with acetabular components inserted without cement are wear of the convex surface of the modular polyethylene liner against the metal shell (so-called backside wear) and fretting of the fixation screws placed through the shell8,110. On the basis of reports of several series of acetabular components inserted without cement, no consistent relationship has been demonstrated between the presence of screw-holes and the development of osteolysis behind the component165,256. Furthermore, particle access by way of screw-holes cannot be the cause of isolated osteolysis in the ischium and pubis. It appears that an interplay of specific design and implantation variables is involved in the development of pelvic osteolysis14,199,202,204,206.

Osteolysis in Association with Total Knee Replacement
Classic, expansile osteolysis associated with total knee replacements in which all components were inserted with cement has been relatively unrecognized in the literature53,69,193,194 despite the fact that retrieval studies have demonstrated substantial wear of the tibial polyethylene103,246. Although uncommon, when osteolysis has occurred in association with components inserted with cement, it has generally been associated with a flat tibial articulation. Long-term follow-up studies have not indicated that inflammatory bone resorption, either classic osteolysis or interfacial bone resorption, is a frequent cause of failure of cemented total knee replacements with a more conforming articulation45.

Osteolysis necessitating reoperation has developed in association with several designs of total knee replacements without cement15,31,135,136,142,182,187. Osteolysis has been reported in the proximal part of the tibia in as many as 16 percent (twenty-seven) of 174 patients142,187, in the distal part of the femur in 11 percent (thirty) of 271 patients31,135, and in the patella in 80 percent (twenty-four) of thirty patients who had a prosthesis implanted without cement136. Radiographs may underrepresent the true extent of the bone loss194. The development of osteolysis is not a direct function of the absence of cement but is related to other design, operative technique, and patient-related variables that are associated with so-called first-generation total knee replacements without cement. These variables include the mechanism of attachment of the modular polyethylene insert to the metal base-plate and the interface between the polyethylene and the metal base-plate as well as the design and location of holes in the base-plate and the presence of fixation screws142. With limited distribution of porous coating, wear particles can enter the implant-bone interface and metaphyseal bone by way of the unbonded interface between the smooth metal and bone. Fully porous-coated components are associated with a lower prevalence of osteolysis than those with a less extensive or discontinuous porous coating240. In addition, access to tibial metaphyseal bone may be gained by means of holes through the tibial base-plate or along the course of tibial fixation screws67,187,240. Increased rates of polyethylene wear have been associated with fixation without cement due to the use of thin polyethylene components with low tibiofemoral conformity and metal-backed patellar components as well as with a younger age, male gender, and greater body weight31,69,136,147,194,195,222. In aggregate, this information indicates that the fundamental variables for the long-term success of a total hip or knee arthroplasty include the attainment and maintenance of the integrity of the implant-bone interface, which provides not only mechanical stability but also a barrier to joint fluid and particulate debris, and a low rate of generation of biologically active particles from all sources. These findings illustrate the interplay between fixation and wear, and they show that many factors, including the materials and design of the implants, the operative techniques, and the amount and type of the patient's activity contribute to the long-term success (or failure) of a total joint replacement.


    The Effective Joint Space
 Top
 Introduction
 Wear Modes
 Friction and Frictional Torque
 Polyethylene Wear
 Periprosthetic Bone Loss
 The Effective Joint Space
 Overview
 References
 
In 1977, Willert reported his classic observations after reoperations for the removal of failed joint replacements241. He found that capsular tissue has some capacity to transport particles through the lymphatic system by means of perivascular lymph spaces, leading to regional and systemic distribution. If the capacity for elimination by this mechanism is exceeded, then particles accumulate in the periarticular tissues. The pseudocapsule appears to be the primary location for phagocytosis of particles. This process results in the development of foreign-body granulomas with areas of necrosis and fibrosis that are somewhat proportional to the amount of particles. In principle, the whole environment of the implant could become involved. Extension of this foreign-body response into the cement-bone interface could cause loosening of the implant. The findings from subsequent experimental models108 and analyses of specimens from well functioning implants retrieved at autopsy are in accord with this principle188,198,201,203,233. Aside from the observations that particles larger than about ten micrometers in linear dimension tend to remain localized and that rapid wear with the production of an extreme amount of particles is associated with systemic distribution, the variables that influence local accumulation compared with systemic distribution of wear particles have not been defined29,43,65,99,217.

Polyethylene wear particles are dispersed in the fluid around a prosthetic joint. Conceptually, the effective joint space includes all periprosthetic regions that are accessible to joint fluid and, thus, accessible to wear particles199. The limits of the effective joint space are determined by the intimacy of the contact between the prosthesis and the bone and the variability of this contact within a given reconstruction. This variability determines the access routes for joint fluid and particles, to and along various interfaces, into bone, and through soft tissues as well. Joint fluid flows according to pressure gradients, following the path of least resistance. Patterns of joint-fluid flow influence the shape and extent of osteolysis. The local concentration of particles is a factor in the local inflammatory reaction and, hence, the degree of bone resorption in that location. As bone is resorbed, a larger space is produced, encouraging preferential flow of joint fluid and wear particles into that location, which fuels additional bone resorption in that area, leading to an expansile lesion. If the rate of particle production is low, or if joint fluid is more evenly distributed, the rate of bone resorption is slower and is accompanied by a fibroblastic response, which results in a more linear pattern of bone loss that often follows the contours of the implant.

In addition to wear particles, soluble factors capable of directly or indirectly effecting bone resorption have been identified in joint fluid. These factors include prostaglandins, interleukins, and matrix metalloproteinases59,79,138,228. Elevated levels of these factors may play a central role in rapidly destructive arthropathy of the hip138. While proteolytic enzyme activity in interfacial tissues contributes to local bone resorption, pseudocapsular tissues also release bone-resorbing factors into joint fluid228. Cytokines and matrix metalloproteinases can be elevated in the fluid around total joint replacements59 and could effect bone resorption at distant sites from the articulating surfaces of the joint.

Variations in the pressure of the joint fluid play a role in transporting the joint fluid and wear particles around the effective joint space. Hendrix et al.96 reported large variations in the pressure of the intracapsular fluid around total hip replacements during activities of daily living, with peak pressures of more than 750 millimeters of mercury (100 kilopascals). During reoperations for osteolysis, intracapsular pressures of more than 500 millimeters of mercury (sixty-seven kilopascals) have been measured204. Fluid pressure of 198 millimeters of mercury (twenty-six kilopascals) has been measured in a femoral diaphyseal lesion6. In addition to moving joint fluid around the effective joint space, the fluid pressure may contribute to expansion of the effective joint space. The expansile and balloon-like nature of classic osteolytic lesions may be a consequence of fluid pressure. The recorded pressures are high enough to interfere with the normal perfusion and oxygenation of bone6, and a physical effect of pressure may account for cases of osteolysis in which no prosthetic particles are identified in the lytic lesions165,234.

In normal synovial joints, such as the hip or knee, bone is not exposed to joint fluid. The boundaries of the joint are defined by the capsule and, within the capsule, bone is covered by cartilage or synovial tissue150,219. In disease processes such as osteoarthritis, and in total joint replacement, the normal anatomical and physiological compartmentalization of a synovial joint can be disrupted. Osteoarthritic cysts, or so-called geodes, are a form of periarticular bone resorption that occurs in the absence of prosthetic implants. Elevated intracapsular fluid pressure can lead to intrusion of joint fluid into cancellous bone through gaps in degenerated articular cartilage. Numerous investigations have shown evidence of a role for intrusion of joint fluid and fluid pressure in the enlargement of these cysts74,139,192. Intracapsular fluid pressure is a function of many variables, including capsular compliance, fluid compartmentalization, joint position, and muscle action. The compliance of the osteoarthritic joint capsule is reduced. Fluid pressures are increased in osteoarthritis and vary with use of the joint. In addition, these pressures can be transmitted to periarticular cysts68,123-125,141,146,179. An important difference between the etiology of geodes and that of osteolysis associated with total joint arthroplasty is activated macrophages; geodes develop without this foreign-body response to prosthetic particles208 but demonstrate a cytokine profile similar to that of the osteolytic lesions around total joint implants127.

The effective joint space can expand into soft tissue as well as bone17,51,93,109,170,224. Increased intracapsular fluid pressures can be painful and may result in the formation of synovial cysts or rupture of the capsule78,122. The presence of an effusion or synovial cysts may be apparent on examination of the knee because of the relatively limited amount of overlying soft tissues. Such evidence of fluid accumulation is less often appreciated in the hip. Synovial cysts may have a protective function by accommodating fluid volume and limiting the increase in intra-articular pressure, thus protecting bone from pressure damage. An inverse relationship has been demonstrated between the development of periarticular bone cysts and the development of synovial cysts76. The natural hip joint can communicate with the iliopsoas bursa37. Pressure-driven synovial fluid may be pumped into the bursa, causing distension that may be symptomatic17. Similar events can occur after total hip and knee arthroplasty109,170,172. In the effective joint space, joint fluid seeks the path of least resistance; the path is variable depending on the anatomical and physiological specifics of the reconstruction and includes soft tissue as well as bone198.

Thus, after total hip or knee replacement, the fluid in the effective joint space has at least three components that can contribute to periprosthetic bone resorption: wear particles, soluble factors, and the physical effects of fluid pressure. Wear particles are released into joint fluid and are variably distributed throughout the effective joint space (Figs. 1-A and 1-B). These particles cause inflammation and contribute to the development of an effusion. Chronic inflammation causes fibrosis and decreased capsular compliance, which can result in elevations of intracapsular joint-fluid pressures. The flow of joint fluid is an active process driven by fluctuations in joint-fluid pressures that occur with activities of daily living and can contribute to progressive expansion of the effective joint space. This fluid also contains various soluble factors that are capable of effecting bone resorption at distant sites. Fluid pressure can cause bone erosion from a purely physical effect or necrosis due to circulatory compromise. In contrast to osteoarthritic joints, in which the progression of the osteoarthritis limits the use of the joint, a well fixed and otherwise well functioning prosthetic joint may be asymptomatic; the patient remains active and maintains the driving forces for osteolysis, which may account for the progression and extreme size of some osteolytic lesions associated with total joint replacement.



View larger version (35K):
[in this window]
[in a new window]
 
FIG1-A: Fig. 1-A Diagram showing the effective joint space, which includes all periprosthetic regions that are accessible to joint fluid. The operative implantation procedure alters the natural anatomy of a joint. In prosthetic total hip and knee replacement, some bone as well as the implant-bone interfaces are exposed within the new joint space. Contraction of muscles, such as the psoas and abductor muscles, and changes in joint position (flexion, extension, abduction, adduction, and rotation) can alter the volume of the effective joint space, resulting in changes in intracapsular joint-fluid pressure.

 


View larger version (41K):
[in this window]
[in a new window]
 
FIG1-B: Fig. 1-B Diagram showing wear and the release of wear particles into the effective joint space. Capsular tissue has some capacity to transport particles through the lymphatic system (large straight arrow). In the effective joint space, joint fluid and wear particles follow the path of least resistance, which is dependent on the prosthetic and anatomical details of each specific reconstruction. The pseudocapsule becomes thickened (small straight arrows) because of phagocytosis of wear particles and the development of foreign-body granulomas. The effective joint space can extend along interfacial planes, expand into bone, or expand into soft tissues or a variety of combinations is possible (curved arrows).

 


    Overview
 Top
 Introduction
 Wear Modes
 Friction and Frictional Torque
 Polyethylene Wear
 Periprosthetic Bone Loss
 The Effective Joint Space
 Overview
 References
 
In summary, improving the durability of total hip and knee replacements requires a reduction in the total production and release of small particles into the biological environment. There is a need not only for more wear-resistant bearing materials to decrease mode-1 wear but also for concomitant improvements in the design and manufacturing of the implant and the operative techniques to minimize the occurrence of mode-2, 3, and 4 wear. Another limitation of current prosthetic joint arthroplasty is the variable disruption of the anatomy and physiology of the joint that occurs as a result of the prosthetic implantation and that may allow variability in the access of joint fluid (and wear particles) to bone. Although a reduction in the production of particles is desirable, limiting access of joint fluid to bone is necessary to decrease periarticular inflammatory bone resorption.

NOTE: The authors thank Christian D. McClung, M.Phil., for his assistance in obtaining and organizing the references and in the preparation of this manuscript.


    Footnotes
 
*One or more of the authors has received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this article. In addition, benefits have been or will be directed to a research fund, foundation, educational institution, or other nonprofit organization with which one or more of the authors is associated. Funds were received in total or partial support of the research or clinical study presented in this article. The funding source was National Institutes of Health Grant AR43314.

{dagger}Joint Replacement Institute, 2400 South Flower Street, Los Angeles, California 90007.

{ddagger}Department of Orthopaedic Surgery, University of Iowa College of Medicine, Iowa City, Iowa 52242-1008.


    References
 Top
 Introduction
 Wear Modes
 Friction and Frictional Torque
 Polyethylene Wear
 Periprosthetic Bone Loss
 The Effective Joint Space
 Overview
 References
 

  1. Adams, D. O., and and Hamilton, T. A.: The cell biology of macrophage activation. Ann. Rev. Immunol., 2: 283-318, 1984.[Medline]

  2. Agins, H. J.; Alcock, N. W.; Bansal, M.; Salvati, E. A.; Wilson, P. D., Jr.; Pellicci, P. M.; and and Bullough, P. G.: Metallic wear in failed titanium-alloy total hip replacements. A histological and quantitative analysis. J Bone and Joint Surg., 70-A: 347-356, March 1988.[Abstract/Free Full Text]

  3. Amstutz, H. C.; Campbell, P.; Kossovsky, N.; and and Clarke, I. C.: Mechanism and clinical significance of wear debris-induced osteolysis. Clin. Orthop., 276: 7-18, 1992.

  4. Amstutz, H. C.; Campbell, P.; McKellop, H.; Schmalzried, T. P.; Gillespie, W. J.; Howie, D.; Jacobs, J.; Medley, J.; and and Merritt, K.: Metal on metal total hip replacement workshop consensus document. Clin. Orthop., 329S: 297-S303, 1996.

  5. Andersson, G. B. J.; Freeman, M. A. R.; and and Swanson, S. A. V.: Loosening of the cemented acetabular cup in total hip replacement. J. Bone and Joint Surg., 54-B(4): 590-599, 1972.

  6. Anthony, P. P.; Gie, G. A.; Howie, C. R.; and and Ling, R. S. M.: Localised endosteal bone lysis in relation to the femoral components of cemented total hip arthroplasties. J. Bone and Joint Surg., 72-B(6): 971-979, 1990.

  7. Argenson, J.-N., and and O'Connor, J. J.: Polyethylene wear in meniscal knee replacement. A one to nine-year retrieval analysis of the Oxford knee. J. Bone and Joint Surg., 74-B(2): 228-232, 1992.

  8. Astion, D. J.; Saluan, P.; Stulberg, B. N.; Rimnac, C. M.; and and Li, S.: The porous-coated anatomic total hip prosthesis: failure of the metal-backed acetabular component. J. Bone and Joint Surg., 78-A: 755-766, May 1996.[Abstract/Free Full Text]

  9. Athanasou, N. A.; Quinn, J.; and and Bulstrode, C. J. K.: Resorption of bone by inflammatory cells derived from the joint capsule of hip arthroplasties. J. Bone and Joint Surg., 74-B(1): 57-62, 1992.

  10. Bankston, A. B.; Cates, H.; Ritter, M. A.; Keating, E. M.; and and Faris, P. M.: Polyethylene wear in total hip arthroplasty. Clin. Orthop., 317: 7-13, 1995.

  11. Bankston, A. B.; Keating, E. M.; Ranawat, C.; Faris, P. M.; and and Ritter, M. A.: Comparison of polyethylene wear in machined versus molded polyethylene. Clin. Orthop., 317: 37-43, 1995.

  12. Barbour, P. S. M.; Barton, D. C.; and and Fisher, J.: The influence of contact stress on the wear of UHMWPE for total replacement hip prostheses. Wear, 181: 250-257, 1995.

  13. Bartel, D. L.; Bicknell, V. L.; and and Wright, T. M.: The effect of conformity, thickness, and material on stresses in ultra-high molecular weight components for total joint replacement. J. Bone and Joint Surg., 68-A: 1041-1051, Sept. 1986.[Abstract/Free Full Text]

  14. Beauchesne, R. P.; Engh, C. A.; and and Schmalzried, T. P.: Roentgenographic evaluation of the AML porous-coated acetabular component: a six-year minimum follow-up study. Orthop. Trans., 16: 318, 1992.

  15. Berry, D. J.; Wold, L. E.; and and Rand, J. A.: Extensive osteolysis around an aseptic, stable, uncemented total knee replacement. Clin. Orthop., 293: 204-207, 1993.

  16. Besong, A. A.; Hailey, J. L.; Ingham, E.; Stone, M.; Wroblewski, B. M.; and and Fisher, J.: Synergistic effects of femoral head roughness and ageing after irradiation in air on the wear of ultra high molecular weight polyethylene in artificial hip joints. Trans. Orthop. Res. Soc., 22: 777, 1997.

  17. Binek, R., and and Levinsohn, E. M.: Enlarged iliopsoas bursa. An unusual cause of thigh mass and hip pain. Clin. Orthop., 224: 158-163, 1987.

  18. Black, J.: The future of polyethylene. J. Bone and Joint Surg., 60-B(3): 303-306, 1978.

  19. Black, J.: Orthopaedic Biomaterials in Research and Practice, pp. 214-215. New York, Churchill Livingstone, 1988.

  20. Blaine, T. A.; Pollice, P. F.; Rosier, R. N.; Reynolds, P. R.; Puzas, J. E.; and and O'Keefe, R. J.: Modulation of the production of cytokines in titanium-stimulated human peripheral blood monocytes by pharmacological agents. The role of cAMP-mediated signaling mechanisms. J. Bone and Joint Surg., 79-A: 1519-1528, Oct. 1997.[Abstract/Free Full Text]

  21. Bobyn, J. D.; Jacobs, J. J.; Tanzer, M.; Urban, R. M.; Aribindi, R.; Sumner, D. R.; Turner, T. M.; and and Brooks, C. E.: The susceptibility of smooth implant surfaces to periimplant fibrosis and migration of polyethylene wear debris. Clin. Orthop., 311: 21-39, 1995.

  22. Bono, J. V.; Sanford, L.; and and Toussaint, J. T.: Severe polyethylene wear in total hip arthroplasty. Observations from retrieved AML PLUS hip implants with an ACS polyethylene liner. J. Arthroplasty, 9: 119-125, 1994.[Medline]

  23. Bosco, J.; Benjamin, J.; and and Wallace, D.: Quantitative and qualitative analysis of polyethylene wear particles in synovial fluid of patients with total knee arthroplasty. A preliminary report. Clin. Orthop., 309: 11-19, 1994.

  24. Buchert, P. K.; Vaughn, B. K.; Mallory, T. H.; Engh, C. A.; and and Bobyn, J. D.: Excessive metal release due to loosening and fretting of sintered particles on porous-coated hip prostheses. Report of two cases. J. Bone and Joint Surg., 68-A: 606-609, April 1986.[Free Full Text]

  25. Buechel, F. F., and and Pappas, M. J.: The New Jersey low-contact-stress knee replacement system: biomechanical rationale and review of the first 123 cemented cases. Arch. Orthop. and Traumatic Surg., 105: 197-204, 1986.

  26. Buechel, F. F., and and Pappas, M. J.: New Jersey low contact stress knee replacement system. Ten-year evaluation of meniscal bearings. Orthop. Clin. North America, 20: erratum, 20: 147-177, 519, 1989.[Medline]

  27. Buechel, F. F.: New Jersey low-contact-stress (LCS) knee replacement system. Clin. Orthop., 264: 309-310, 1991.

  28. Bugbee, W. D.; Culpepper, W. J., II; Engh, C. A., Jr.; and and Engh, C. A., Sr.: Long-term clinical consequences of stress-shielding after total hip arthroplasty without cement. J. Bone and Joint Surg., 79-A: 1007-1012, July 1997.[Abstract/Free Full Text]

  29. Bullough, P. G.: Tissue reaction to wear debris generated from total hip replacements. In The Hip. Proceedings of the First Open Scientific Meeting of the Hip Society, p. 80. St. Louis, C. V. Mosby, 1973.

  30. Burke, D. W.; O'Connor, D. O.; Zalenski, E. B.; Jasty, M.; and and Harris, W. H.: Micromotion of cemented and uncemented femoral components. J. Bone and Joint Surg., 73-B(1): 33-37, 1991.

  31. Cadambi, A.; Engh, G. A.; Dwyer, K. A.; and and Vinh, T. N.: Osteolysis of the distal femur after total knee arthroplasty. J. Arthroplasty, 9: 579-594, 1994.[Medline]

  32. Callaghan, J. J.; Pedersen, D. R.; Olejniczak, J. P.; Goetz, D. D.; and and Johnston, R. C.: Radiographic measurement of wear in 5 cohorts of patients observed for 5 to 22 years. Clin. Orthop., 317: 14-18, 1995.

  33. Cameron, H. U.: Tibial component wear in total knee replacement. Clin. Orthop., 309: 29-32, 1994.

  34. Campbell, P.; Ma, S.; Yeom, B.; McKellop, H.; Schmalzried, T. P.; and and Amstutz, H. C.: Isolation of predominantly submicron-sized UHMWPE wear particles from periprosthetic tissues. J. Biomed. Mater. Res., 29: 127-131, 1995.[Medline]

  35. Capello, W. N.; D'Antonio, J. A.; Feinberg, J. R.; and and Manley, M. T.: Hydroxyapatite-coated total hip femoral components in patients less than fifty years old. Clinical and radiographic results after five to eight years of follow-up. J. Bone and Joint Surg., 79-A: 1023-1029, July 1997.[Abstract/Free Full Text]

  36. Cates, H. E.; Faris, P. M.; Keating, E. M.; and and Ritter, M. A.: Polyethylene wear in cemented metal-backed acetabular cups. J. Bone and Joint Surg., 75-B(2): 249-253, 1993.[Abstract/Free Full Text]

  37. Chandler, S. B.: The iliopsoas bursa in man. Anat. Rec., 58: 235-240, 1934.

  38. Charnley, J.; Follacci, F. M.; and and Hammond, B. T.: The long-term reaction of bone to self-curing acrylic cement. J. Bone and Joint Surg., 50-B(4): 822-829, 1968.

  39. Charnley, J.; Kamangar, A.; and and Longfield, M. D.: The optimum size of prosthetic heads in relation to the wear of plastic sockets in total replacement of the hip. Med. and Biol. Eng., 7: 31-39, 1969.

  40. Charnley, J., and and Cupic, Z.: The nine and ten year results of the low-friction arthroplasty of the hip. Clin. Orthop., 95: 9-25, 1973.

  41. Charnley, J.: Fracture of femoral prostheses in total hip replacement. A clinical study. Clin. Orthop., 111: 105-120, 1975.

  42. Charnley, J., and and Halley, D. K.: Rate of wear in total hip replacement. Clin. Orthop., 112: 170-179, 1975.

  43. Charnley, J.: Low Friction Arthroplasty of the Hip. Theory and Practice. New York, Springer, 1979.

  44. Chiba, J., and and Rubash, H. E.: A biochemical, histologic, and immunohistologic analysis of membranes obtained from failed cemented and cementless total knee arthroplasty. Clin. Orthop., 299: 114-124, 1994.

  45. Colizza, W. A.; Insall, J. N.; and and Scuderi, G. R.: The posterior stabilized total knee prosthesis. Assessment of polyethylene damage and osteolysis after a ten-year-minimum follow-up. J. Bone and Joint Surg., 77-A: 1713-1720, Nov. 1995.[Abstract/Free Full Text]

  46. Collier, J. P.; Mayor, M. B.; McNamara, J. L.; Surprenant, V. A.; and and Jensen, R. E.: Analysis of the failure of 122 polyethylene inserts from uncemented tibial knee components. Clin. Orthop., 273: 232-242, 1991.

  47. Collier, J. P.; Sperling, D. K.; Currier, J. H.; Sutula, L. C.; Saum, K. A.; and and Mayor, M. B.: Impact of gamma sterilization on clinical performance of polyethylene in the knee. J. Arthroplasty, 11: 377-389, 1996.[Medline]

  48. Collier, J. P.; Sutula, L. C.; Currier, B. H.; Currier, J. H.; Wooding, R. E.; Williams, I. R.; Farber, K. B.; and and Mayor, M. B.: Overview of polyethylene as a bearing material. Comparison of sterilization methods. Clin. Orthop., 333: 76-86, 1996.

  49. Cooper, J. R.; Dowson, D.; Fisher, J.; and and Jobbins, B.: Ceramic bearing surfaces in artificial joints: resistance to third body damage. J. Med. Eng. and Technol., 15: 63-67, 1991.

  50. Cooper, J.; Dowson, D.; and and Fisher, J.: The effect of transfer film and surface roughness on the wear of lubricated ultrahigh molecular weight polyethylene. Clin. Mater., 14: 295-302, 1993.

  51. Coventry, M. B.; Polley, H. F.; and and Weiner, A. D.: Rheumatoid synovial cyst of the hip. Report of three cases. J. Bone and Joint Surg., 41-A: 721-730, June 1959.[Abstract/Free Full Text]

  52. Currier, B. H.; Currier, J. H.; Collier, J. P.; Mayor, M. B.; and and Scott, R. D.: Shelf life and in vivo duration. Impacts on performance of tibial bearings. Clin. Orthop., 342: 111-122, 1997.

  53. Dannenmaier, W. C.; Haynes, D. W.; and and Nelson, C. L.: Granulomatous reaction and cystic bony destruction associated with high wear rate in a total knee prosthesis. Clin. Orthop., 198: 224-230, 1985.

  54. Devane, P. A.; Bourne, R. B.; Rorabeck, C. H.; Hardie, R. M.; and and Horne, J. G.: Measurement of polyethylene wear in metal-backed acetabular cups. I. Three-dimensional technique. Clin. Orthop., 319: 303-316, 1995.

  55. Devane, P. A.; Bourne, R. B.; Rorabeck, C. H.; MacDonald, S.; and and Robinson, E. J.: Measurement of polyethylene wear in metal-backed acetabular cups. II. Clinical application. Clin. Orthop., 319: 317-326, 1995.

  56. Devane, P. A.; Robinson, E. J.; Bourne, R. B.; Rorabeck, C. H.; Nayak, N. N.; and and Horne, I. G.: Measurement of polyethylene wear in acetabular components inserted with and without cement. A randomized trial. J. Bone and Joint Surg., 79-A: 682-689, May 1997.[Abstract/Free Full Text]

  57. Diduch, D. R.; Insall, J. N.; Scott, W. N.; Scuderi, G. R.; and and Font-Rodriguez, D.: Total knee replacement in young, active patients. Long-term follow-up and functional outcome. J. Bone and Joint Surg., 79-A: 575-582, April 1997.[Abstract/Free Full Text]

  58. Doorn, P. F.; Mirra, J. M.; Campbell, P. A.; and and Amstutz, H. C.: Tissue reaction to metal on metal total hip prostheses. Clin. Orthop., 329S: 187-S205, 1996.

  59. Dorr, L. D.; Bloebaum, R.; Emmanual, J.; and and Meldrum, R.: Histologic, biochemical, and ion analysis of tissue and fluids retrieved during total hip arthroplasty. Clin. Orthop., 261: 82-95, 1990.

  60. Dowling, J. W.; Atkinson, J. R.; Dowson, D.; and and Charnley, J.: The characteristics of acetabular cups worn in the human body. J. Bone and Joint Surg., 60-B(3): 375-382, 1978.

  61. Dowson, D.; Taheri, S.; and and Wallbridge, N.: The role of counterface imperfections in the wear of polyethylene. Wear, 119: 277, 1987.

  62. Dowson, D.: A comparative study of the performance of metallic and ceramic femoral head components in total replacement hip joints. Wear, 190: 171-183, 1995.

  63. Eftekhar, N. S., and Pawluk, R. J.: Role of surgical preparation in acetabular cup fixation. In The Hip. Proceedings of the Eighth Open Scientific Meeting of the Hip Society, pp. 308-328. St. Louis, C. V. Mosby, 1980.

  64. Engh, C. A., Jr.; Culpepper, W. J., II; and and Engh, C. A.: Long-term results of use of the anatomic medullary locking prosthesis in total hip arthroplasty. J. Bone and Joint Surg., 79-A: 177-184, Feb. 1997.[Abstract/Free Full Text]

  65. Engh, C. A., Jr.; Moore, K. D.; Vinh, T. N.; and and Engh, G. A.: Titanium prosthetic wear debris in remote bone marrow. A report of two cases. J. Bone and Joint Surg., 79-A: 1721-1725, Nov. 1997.[Free Full Text]

  66. Engh, G. A.: Failure of the polyethylene bearing surface of a total knee replacement within four years. A case report. J. Bone and Joint Surg., 70-A: 1093-1096, Aug. 1988.[Free Full Text]

  67. Engh, G. A.; Dwyer, K. A.; and and Hanes, C. K.: Polyethylene wear of metal-backed tibial components in total and unicompartmental knee prostheses. J. Bone and Joint Surg., 74-B(1): 9-17, 1992.

  68. Eyring, E. J., and and Murray, W. R.: The effect of joint position on the pressure of intra-articular effusion. J. Bone and Joint Surg., 46-A: 1235-1241, Sept. 1964.[Free Full Text]

  69. Ezzet, K. A.; Garcia, R.; and and Barrack, R. L.: Effect of component fixation method on osteolysis in total knee arthroplasty. Clin. Orthop., 321: 86-91, 1995.

  70. Fisher, J.: Surface damage to femoral head prostheses [letter]. J. Bone and Joint Surg., 76-B(5): 852, 1994.

  71. Fisher, J.; Chan, K. L.; Hailey, J. L.; Shaw, D.; and and Stone, M.: Preliminary study of the effect of aging following irradiation on the wear of ultrahigh-molecular-weight polyethylene. J. Arthroplasty, 10: 689-692, 1995.[Medline]

  72. Fisher, J.; Firkins, P.; Reeves, E. A.; Hailey, J. L.; and and Isaac, G. H.: The influence of scratches to metallic counterfaces on the wear of ultra-high molecular weight polyethylene. Proc. Inst. Mech. Eng., Part H—J. Eng. Med., 209: 263-264, 1995.

  73. Fornasier, V. L., and and Cameron, H. U.: The femoral stem/cement interface in total hip replacement. Clin. Orthop., 116: 248-252, 1976.

  74. Freund, E.: The pathologic significance of intra-articular pressure. Edinburgh Med. J., 47: 192-203, 1940.

  75. García-Cimbrelo, E.; Diez-Vazquez, V.; Madero, R.; and and Munuera, L.: Progression of radiolucent lines adjacent to the acetabular component and factors influencing migration after Charnley low-friction total hip arthroplasty. J. Bone and Joint Surg., 79-A: 1373-1380, Sept. 1997.[Abstract/Free Full Text]

  76. Genovese, G. R.; Jayson, M. I. V.; and and Dixon, A. St. J.: Protective value of synovial cysts in rheumatoid knees. Ann. Rheumat. Dis., 31: 179-182, 1972.[Free Full Text]

  77. Gie, G. A.; Fowler, J. L.; Lee, A. J. C.; and and Ling, R. S. M.: The long-term behaviour of a totally collarless polished femoral component in cemented total hip arthroplasty. In Proceedings of the British Orthopaedic Association. J. Bone and Joint Surg., 72-B(5): 935, 1990.[Free Full Text]

  78. Goddard, N. J., and and Gosling, P. T.: Intra-articular fluid pressure and pain in osteoarthritis of the hip. J. Bone and Joint Surg., 70-B(1): 52-55, 1988.

  79. Goldring, S. R.; Schiller, A. L.; Roelke, M.; Rourke, C. M.; O'Neill, D. A.; and and Harris, W. H.: The synovial-like membrane at the bone-cement interface in loose total hip replacements and its proposed role in bone lysis. J. Bone and Joint Surg., 65-A: 575-584, June 1983.[Abstract/Free Full Text]

  80. Goodfellow, J., and and O'Connor, J.: The mechanics of the knee and prosthesis design. J. Bone and Joint Surg., 60-B(3): 358-369, 1978.

  81. Goodfellow, J.: Knee prostheses—one step forward, two steps back. J. Bone and Joint Surg., 74-B(1): 1-2, 1992.[Free Full Text]

  82. Goodman, S.; Aspenberg, P.; Song, Y.; Knoblich, G.; Huie, P.; Regula, D.; and and Lidgren, L.: Tissue ingrowth and differentiation in the bone-harvest chamber in the presence of cobalt-chromium-alloy and high-density-polyethylene particles. J. Bone and Joint Surg., 77-A: 1025-1035, July 1995.[Abstract/Free Full Text]

  83. Green, T.; Fisher, J.; and and Ingham, E.: Polyethylene particles of a "critical" size are necessary for the induction of IL-6 by macrophages in vitro. Trans. Orthop. Res. Soc., 22: 733, 1997.

  84. Griffith, M. J.; Seidenstein, M. K.; Williams, D.; and and Charnley, J.: Socket wear in Charnley low friction arthroplasty of the hip. Clin. Orthop., 137: 37-47, 1978.

  85. Gruen, T. A.; McNeice, G. M.; and and Amstutz, H. C.: "Modes of failure" of cemented stem-type femoral components. A radiographic analysis of loosening. Clin. Orthop., 141: 17-27, 1979.

  86. Guttmann, D.; Schmalzried, T. P.; Jasty, M.; and and Harris, W. H.: Light microscopic identification of submicron polyethylene wear debris. J. Appl. Biomater., 4: 303-307, 1993.

  87. Guttmann, D.; Schmalzried, T. P.; Kabo, J. M.; and and Amstutz, H. C.: Characterization of back-side wear in modular polyethylene liners. Orthop. Trans., 18: 418-419, 1994.

  88. Hahn, D. W.; Wolfarth, D. L.; and and Parks, N. L.: Characterization of submicron polyethylene wear debris from synovial-fluid samples of revised knee replacements using a light-scattering technique. J. Biomed. Mater. Res., 31: 355-363, 1996.[Medline]

  89. Hailey, J. L.; Ingham, E.; Stone, M.; Wroblewski, B. M.; and and Fisher, J.: Quantitative comparison of acetabular cup wear rates and femoral head damage in explanted Charnley hip prostheses. Trans. Orthop. Res. Soc., 22: 853, 1997.

  90. Harris, W. H.; Schiller, A. L.; Scholler, J.-M.; Freiberg, R. A.; and and Scott, R.: Extensive localized bone resorption in the femur following total hip replacement. J. Bone and Joint Surg., 58-A: 612-618, July 1976.[Abstract/Free Full Text]

  91. Harris, W. H.: Is it advantageous to strengthen the cement-metal interface and use a collar for cemented femoral components of total hip replacements?. Clin. Orthop., 285: 67-72, 1992.

  92. Harris, W. H.: The problem is osteolysis. Clin. Orthop., 311: 46-53, 1995.

  93. Hattrup, S. J.; Bryan, R. S.; Gaffey, T. A.; and and Stanhope, C. R.: Pelvic mass causing vesical compression after total hip arthroplasty. Case report. Clin. Orthop., 227: 184-189, 1988.[Medline]

  94. Haynes, D. R.; Rogers, S. D.; Howie, D. W.; Pearcy, M. J.; and and Vernon-Roberts, B.: Drug inhibition of the macrophage response to metal wear particles in vitro. Clin. Orthop., 323: 316-326, 1996.

  95. Heim, C. S.; Postak, P. D.; and Greenwald, A. S.: Factors influencing the longevity of UHMWPE tibial components. In Instructional Course Lectures, American Academy of Orthopaedic Surgeons. Vol. 45, pp. 303-312. Rosemont, Illinois, American Academy of Orthopaedic Surgeons, 1996.

  96. Hendrix, R. W.; Wixon, R. L.; Rana, N. A.; and and Rogers, L. F.: Arthrography after total hip arthroplasty: a modified technique used in the diagnosis of pain. Radiology, 148: 647-652, 1983.[Abstract/Free Full Text]

  97. Herman, J. H.; Sowder, W. G.; Anderson, D.; Appel, A. M.; and and Hopson, C. N.: Polymethylmethacrylate-induced release of bone-resorbing factors. J. Bone and Joint Surg., 71-A: 1530-1541, Dec. 1989.[Abstract/Free Full Text]

  98. Hernandez, J. R.; Keating, E. M.; Faris, P. M.; Meding, J. B.; and and Ritter, M. A.: Polyethylene wear in uncemented acetabular components. J. Bone and Joint Surg., 76-B(2): 263-266, 1994.

  99. Hicks, D. G.; Judkins, A. R.; Sickel, J. Z.; Rosier, R. N.; Puzas, J. E.; and and O'Keefe, R. J.: Granular histiocytosis of pelvic lymph nodes following total hip arthroplasty. The presence of wear debris, cytokine production, and immunologically activated macrophages. J. Bone and Joint Surg., 78-A: 482-496, April 1996.[Abstract/Free Full Text]

  100. Hirakawa, K.; Bauer, T. W.; Stulberg, B. N.; and and Wilde, A. H.: Comparison and quantitation of wear debris of failed total hip and total knee arthroplasty. J. Biomed. Mater. Res., 31: 257-263, 1996.[Medline]

  101. Hodgkinson, J. P.; Shelley, P.; and and Wroblewski, B. M.: The correlation between the roentgenographic appearance and operating findings at the bone-cement junction of the socket in Charnley low friction arthroplasties. Clin. Orthop., 228: 105-109, 1998.

  102. Holtrop, M. E.; Cox, K. A.; and and Glowacki, J.: Cells of the mononuclear phagocytic system resorb implanted bone matrix: a histologic and ultrastructural study. Calcif. Tissue Internat., 34: 488-494, 1982.[Medline]

  103. Hood, R. W.; Wright, T. M.; and and Burstein, A. H.: Retrieval analysis of total knee prostheses: a method and its application to 48 total condylar prostheses. J. Biomed. Mater. Res., 17: 829-842, 1983.[Medline]

  104. Hop, J. D.; Callaghan, J. J.; Olejniczak, J. P.; Pedersen, D. R.; Brown, T. D.; and and Johnston, R. C.: Contribution of cable debris generation to accelerated polyethylene wear. Clin. Orthop., 344: 20-32, 1997.

  105. Horikoshi, M.; Macaulay, W.; Booth, R. E.; Crossett, L. S.; and and Rubash, H. E.: Comparison of interface membranes obtained from failed cemented and cementless hip and knee prostheses. Clin. Orthop., 309: 69-87, 1994.

  106. Horowitz, S. M.; Doty, S. B.; Lane, J. M.; and and Burstein, A. H.: Studies of the mechanism by which the mechanical failure of polymethylmethacrylate leads to bone resorption. J. Bone and Joint Surg., 75-A: 802-813, June 1993.[Abstract/Free Full Text]

  107. Horowitz, S. M.; Algan, S. A.; and and Purdon, M. A.: Pharmacologic inhibition of particulate-induced bone resorption. J. Biomed. Mater. Res., 31: 91-96, 1996.[Medline]

  108. Howie, D. W.; Vernon-Roberts, B.; Oakeshott, R.; and and Manthey, B.: A rat model of resorption of bone at the cement-bone interface in the presence of polyethylene wear particles. J. Bone and Joint Surg., 70-A: 257-263, Feb. 1988.[Abstract/Free Full Text]

  109. Howie, D. W.; Cain, C. M. J.; and and Cornish, B. L.: Pseudo-abscess of the psoas bursa in failed double-cup arthroplasty of the hip. J. Bone and Joint Surg., 73-B(1): 29-32, 1991.

  110. Huk, O. L.; Bansal, M.; Betts, F.; Rimnac, C. M.; Lieberman, J. R.; Huo, M. H.; and and Salvati, E. A.: Polyethylene and metal debris generated by non-articulating surfaces of modular acetabular components. J. Bone and Joint Surg., 76-B(4): 568-574, 1994.

  111. Insall, J. N.; Hood, R. W.; Flawn, L. B.; and and Sullivan, D. J.: The total condylar knee prosthesis in gonarthrosis. A five to nine-year follow-up of the first one hundred consecutive replacements. J. Bone and Joint Surg., 65-A: 619-628, June 1983.[Abstract/Free Full Text]

  112. Isaac, G. H.; Wroblewski, B. M.; Atkinson, J. R.; and and Dowson, D.: A tribological study of retrieved hip prostheses. Clin. Orthop., 276: 115-125, 1992.

  113. Isaac, G. H.; Dowson, D.; and and Wroblewski, B. M.: An investigation into the origins of time-dependent variation in penetration rates with Charnley acetabular cups—wear, creep or degradation?. Proc. Inst. Mech. Eng., Part H—J. Eng. Med., 210: 209-216, 1996.

  114. Jacobs, J. J.; Skipor, A. K.; Doorn, P. F.; Campbell, P.; Schmalzried, T. P.; Black, J.; and and Amstutz, H. C.: Cobalt and chromium concentrations in patients with metal on metal total hip replacements. Clin. Orthop., 329S: 256-S263, 1996.

  115. Jasty, M. J.; Floyd, W. E., III; Schiller, A. L.; Goldring, S. R.; and and Harris, W. H.: Localized osteolysis in stable, non-septic total hip replacement. J. Bone and Joint Surg., 68-A: 912-919, July 1986.[Abstract/Free Full Text]

  116. Jasty, M.; Maloney, W. J.; Bragdon, C. R.; Haire, T.; and and Harris, W. H.: Histomorphological studies of the long-term skeletal responses to well fixed cemented femoral components. J. Bone and Joint Surg., 72-A: 1220-1229, Sept. 1990.[Abstract/Free Full Text]

  117. Jasty, M.; Bragdon, C. R.; Maloney, W. J.; Haire, T.; and and Harris, W. H.: Ingrowth of bone in failed fixation of porous-coated femoral components. J. Bone and Joint Surg., 73-A: 1331-1337, Oct. 1991.[Abstract/Free Full Text]

  118. Jasty, M.; Maloney, W. J.; Bragdon, C. R.; O'Connor, D. O.; Haire, T.; and and Harris, W. H.: The initiation of failure in cemented femoral components of hip arthroplasties. J. Bone and Joint Surg., 73-B(4): 551-558, 1991.[Abstract/Free Full Text]

  119. Jasty, M.; Bragdon, C. R.; Lee, K.; Hanson, A.; and and Harris, W. H.: Surface damage to cobalt-chrome femoral head prostheses. J. Bone and Joint Surg., 76-B(1): 73-77, 1994.

  120. Jasty, M.; Bragdon, C.; O'Connor, D.; Muratoglu, O.; Permnath, V.; and and Merrill, E.: Marker improvement in the wear resistance of a new form of UHMWPE in a physiologic hip simulator. Trans. Soc. Biomater., 20: 157, 1997.

  121. Jasty, M.; Goetz, D. D.; Bragdon, C. R.; Lee, K. R.; Hanson, A. E.; Elder, J. R.; and and Harris, W. H.: Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations. J. Bone and Joint Surg., 79-A: 349-358, March 1997.[Abstract/Free Full Text]

  122. Jayson, M. I. V.; Swannell, A. J.; Kirk, J. A.; and and Dixon, A. St. J.: Acute joint rupture. Ann. Phys. Med., 10: 175-179, 1969.[Medline]

  123. Jayson, M. I. V., and and Dixon, A. St. J.: Intra-articular pressure in rheumatoid arthritis of the knee. II. Effect of intra-articular pressure on blood circulation to the synovium. Ann. Rheumat. Dis., 29: 266-268, 1970.[Free Full Text]

  124. Jayson, M. I. V., and and Dixon, A. St. J.: Intra-articular pressure in rheumatoid arthritis of the knee. III. Pressure changes during joint use. Ann. Rheumat. Dis., 29: 401-408, 1970.[Free Full Text]

  125. Jayson, M. I. V.; Rubenstein, D.; and and Dixon, A. St. J.: Intra-articular pressure and rheumatoid geodes (bone cysts). Ann. Rheumat. Dis., 29: 496-502, 1970.[Free Full Text]

  126. Jiranek, W. A.; Machado, M.; Jasty, M.; Jevsevar, D.; Wolfe, H. J.; Goldring, S. R.; Goldberg, M. J.; and and Harris, W. H.: Production of cytokines around loosened cemented acetabular components. Analysis with immunohistochemical techniques and in situ hybridization. J. Bone and Joint Surg., 75-A: 863-879, June 1993.[Abstract/Free Full Text]

  127. Jiranek, W. A.; Dalury, D. F.; Qui, J.-Y.; and and Cardea, J.: Bone resorption in osteoarthritic cysts demonstrates a similar cytokine profile as osteolytic lesions around joint implants. Trans. Orthop. Res. Soc., 22: 159, 1997.

  128. Johnston, R. C., and and Crowninshield, R. D.: Roentgenologic results of total hip arthroplasty. A ten-year follow-up study. Clin. Orthop., 181: 92-98, 1983.

  129. Johnston, R. C.; Fitzgerald, R. H., Jr.; Harris, W. H.; Poss, R.; Müller, M. E.; and and Sledge, C. B.: Clinical and radiographic evaluation of total hip replacement. A standard system of terminology for reporting results. J. Bone and Joint Surg., 72-A: 161-168, Feb. 1990.[Free Full Text]

  130. Jones, L. C., and and Hungerford, D. S.: Cement disease. Clin. Orthop., 225: 192-206, 1987.

  131. Jones, S. M. G.; Pinder, I. M.; Moran, C. G.; and and Malcolm, A. J.: Polyethylene wear in uncemented knee replacements. J. Bone and Joint Surg., 74-B(1): 18-22, 1992.

  132. Kabo, J. M.; Gebhard, J. S.; Loren, G.; and and Amstutz, H. C.: In vivo wear of polyethylene acetabular components. J. Bone and Joint Surg., 75-B(2): 254-258, 1993.

  133. Kelley, S. S., and and Johnston, R. C.: Debris from cobalt-chrome cable may cause acetabular loosening. Clin. Orthop., 285: 140-146, 1992.

  134. Kilgus, D. J.; Moreland, J.; Finerman, G. A.; Funahashi, T. T.; and and Tipton, J. S.: Catastrophic wear of tibial polyethylene inserts. Clin. Orthop., 273: 223-231, 1991.

  135. Kilgus, D. J.; Funahashi, T. T.; and and Campbell, P. A.: Massive femoral osteolysis and early disintegration of a polyethylene-bearing surface of a total knee replacement. A case report. J. Bone and Joint Surg., 74-A: 770-774, June 1992.[Free Full Text]

  136. Kim, Y.-H.; Oh, J.-H.; and and Oh, S.-H.: Osteolysis around cementless porous-coated anatomic knee prostheses. J. Bone and Joint Surg., 77-B(2): 236-241, 1995.

  137. Kobayashi, S.; Takaoka, K.; Saito, N.; and and Hisa, K.: Factors affecting aseptic failure of fixation after primary Charnley total hip arthroplasty. Multivariate survival analysis. J. Bone and Joint Surg., 79-A: 1618-1627, Nov. 1997.[Abstract/Free Full Text]

  138. Komiya, S.; Inoue, A.; Sasaguri, Y.; Minamitani, K.; and and Morimatsu, M.: Rapidly destructive arthropathy of the hip. Studies on bone resorptive factors in joint fluid with a theory of pathogenesis. Clin. Orthop., 284: 273-282, 1992.

  139. Landells, J. W.: The bone cysts of osteoarthritis. J. Bone and Joint Surg., 35-B(4): 643-649, 1953.[Abstract/Free Full Text]

  140. Landy, M. M., and and Walker, P. S.: Wear of ultra-high-molecular-weight polyethylene components of 90 retrieved knee prostheses. J. Arthroplasty, (3 Supplement): 73-S85, 1988.[Medline]

  141. Levick, J. R.: Joint pressure-volume studies: their importance, design and interpretation. J. Rheumatol., 10: 353-357, 1983.[Medline]

  142. Lewis, P. L.; Rorabeck, C. H.; and and Bourne, R. B.: Screw osteolysis after cementless total knee replacement. Clin. Orthop., 321: 173-177, 1995.

  143. Li, S., and and Burstein, A. H.: Current concepts review. Ultra-high molecular weight polyethylene. The material and its use in total joint implants. J. Bone and Joint Surg., 76-A: 1080-1090, July 1994.[Free Full Text]

  144. Livermore, J.; Ilstrup, D.; and and Morrey, B.: Effect of femoral head size on wear of the polyethylene acetabular component. J. Bone and Joint Surg., 72-A: 518-528, April 1990.[Abstract/Free Full Text]

  145. Livingston, B. J.; Chmell, M. J.; Spector, M.; and and Poss, R.: Complications of total hip arthroplasty associated with the use of an acetabular component with a Hylamer liner. J. Bone and Joint Surg., 79-A: 1529-1538, Oct. 1997.[Abstract/Free Full Text]

  146. Lloyd-Roberts, G. C.: The role of capsular changes in osteoarthritis of the hip joint. J. Bone and Joint Surg., 35-B(4): 627-642, 1953.

  147. Lombardi, A. V., Jr.; Engh, G. A.; Volz, R. G.; Albrigo, J. L.; and and Brainard, B. J.: Fracture/dissociation of the polyethylene in metal-backed patellar components in total knee arthroplasty. J. Bone and Joint Surg., 70-A: 675-679, June 1998.[Abstract/Free Full Text]

  148. Lombardi, A. V., Jr.; Mallory, T. H.; Vaughn, B. K.; and and Drouillard, P.: Aseptic loosening in total hip arthroplasty secondary to osteolysis induced by wear debris from titanium-alloy modular femoral heads. J. Bone and Joint Surg., 71-A: 1337-1342, Oct. 1989.[Abstract/Free Full Text]

  149. Lu, Z.; Ebramzadeh, E.; McKellop, H.; and and Sarmiento, A.: Stable partial debonding of the cement interfaces indicated by a finite element model of a total hip prosthesis. J. Orthop. Res., 14: 238-244, 1996.[Medline]

  150. McCarty, D. J.: The physiology of the normal synovium. In Joints and Synovial Fluid, edited by L. Sokoloff. Vol. 2, pp. 293-314. New York, Academic Press, 1980.

  151. McKellop, H.; Clarke, L.; Markolf, K.; and and Amstutz, H. A.: Friction and wear properties of polymer, metal, and ceramic prosthetic joint materials evaluated on a multichannel screening device. J. Biomed. Mater. Res., 15: 619-653, 1981.[Medline]

  152. McKellop, H.; Hosseinian, A.; Burgoyne, K.; and Clarke, I.: Polyethylene wear against titanium alloy compared to stainless steel and cobalt-chromium alloys. In Proceedings of the Second World Congress on Biomaterials, Tenth Annual Meeting of the Society for Biomaterials, Washington, D.C., p. 313, 1984.

  153. McKellop, H. A.; Sarmiento, A.; Schwinn, C. P.; and and Ebramzadeh, E.: In vivo wear of titanium-alloy hip prostheses. J. Bone and Joint Surg., 72-A: 512-517, April 1990.[Abstract/Free Full Text]

  154. McKellop, H., and Lu, B.: Friction and wear of polyethylene/metal and polyethylene/ceramic hip prostheses on a joint simulator. In Transactions of the Fourth World Biomaterials Congress, p. 118. Berlin, European Society for Biomaterials, April 1992.

  155. McKellop, H.; Lu, B.; and and Benya, P.: Friction, lubrication and wear of cobalt-chromium, alumina and zirconia hip prostheses compared on a joint simulator. Trans. Orthop. Res. Soc., 17: 402, 1992.

  156. McKellop, H. A.; Campbell, P.; Park, S.-H.; Schmalzried, T. P.; Grigoris, P.; Amstutz, H. C.; and and Sarmiento, A.: The Origin of submicron polyethylene wear debris in total hip arthroplasty. Clin. Orthop., 311: 3-20, 1995.

  157. McKellop, H.; Park, S.-H.; Chiesa, R.; Doorn, P.; Lu, B.; Normand, P.; Grigoris, P.; and and Amstutz, H.: In vivo wear of 3 types of metal on metal hip prostheses during 2 decades of use. Clin. Orthop., 329S: 128-S140, 1996.

  158. McKellop, H.; Röstlund, T.; Ebramzadeh, E.; and Sarmiento, A.: Wear of titanium 6-4 alloy in laboratory tests and in retrieved human joint replacements. In Medical Applications of Titanium and Its Alloys: The Material and Biological Issues, ASTM STP 1272, pp. 266-293. Edited by S. Brown and J. Lemons. West Conshohocken, Pennsylvania, American Society for Testing and Materials, 1996.

  159. McKellop, H.; Shen, F.; Yu, Y.; Lu, B.; Salovey, R.; and Campbell, P.: Effect of sterilization method and other modifications on the wear resistance of UHMWPE acetabular cups. In Polyethylene Wear in Orthopaedic Implants Workshop, pp. 20-31. Minneapolis, Society for Biomaterials, 1997.

  160. Ma, S. M.; Kabo, J. M.; and and Amstutz, H. C.: Frictional torque in surface and conventional hip replacement. J. Bone and Joint Surg., 65-A: 366-370, March 1983.[Abstract/Free Full Text]

  161. Mai, M. T.; Schmalzried, T. P.; Dorey, F. J.; Campbell, P. A.; and and Amstutz, H. C.: The contribution of frictional torque to loosening at the cement-bone interface in Tharies hip replacements. J. Bone and Joint Surg., 78-A: 505-511, April 1996.[Abstract/Free Full Text]

  162. Maloney, W. J.; Jasty, M.; Burke, D. W.; O'Connor, D. O.; Zalenski, E. B.; Bragdon, C.; and and Harris, W. H.: Biomechanical and histologic investigation of cemented total hip arthroplasties, A study of autopsy-retrieved femurs after in vivo cycling. Clin. Orthop., 249: 129-140, 1989.

  163. Maloney, W. J.; Jasty, M.; Harris, W. H.; Galante, J. O.; and and Callaghan, J. J.: Endosteal erosion in association with stable uncemented femoral components. J. Bone and Joint Surg., 72-A: 1025-1034, Aug. 1990.[Abstract/Free Full Text]

  164. Maloney, W. J.; Jasty, M.; Rosenberg, A.; and and Harris, W. H.: Bone lysis in well-fixed cemented femoral components. J. Bone and Joint Surg., 72-B(6): 966-970, 1990.

  165. Maloney, W. J.; Peters, P.; Engh, C. A.; and and Chandler, H.: Severe osteolysis of the pelvis in association with acetabular replacement without cement. J. Bone and Joint Surg., 75-A: 1627-1635, Nov. 1993.[Abstract/Free Full Text]

  166. Maloney, W. J.; Smith, R. L.; Schmalzried, T. P.; Chiba, J.; Huene, D.; and and Rubash, H.: Isolation and characterization of wear particles generated in patients who have had failure of a hip arthroplasty without cement. J. Bone and Joint Surg., 77-A: 1301-1310, Sept. 1995.[Abstract/Free Full Text]

  167. Maloney, W. J.; Herzwurm, P.; Paprosky, W.; Rubash, H. E.; and and Engh, C. A.: Treatment of pelvic osteolysis associated with a stable acetabular component inserted without cement as part of a total hip replacement. J. Bone and Joint Surg., 79-A: 1628-1634, Nov. 1997.[Abstract/Free Full Text]

  168. Margevicius, K. J.; Bauer, T. W.; McMahon, J. T.; Brown, S. A.; and and Merritt, K.: Isolation and characterization of debris in membranes around total joint prostheses. J. Bone and Joint Surg., 76-A: 1664-1675, Nov. 1994.[Abstract/Free Full Text]

  169. Martell, J. M., and and Berdia, S.: Determination of polyethylene wear in total hip replacements with use of digital radiographs. J. Bone and Joint Surg., 79-A: 1635-1641, Nov. 1997.[Abstract/Free Full Text]

  170. Matsumoto, K.; Hukuda, S.; Nishioka, J.; and and Fujita, T.: Iliopsoas bursal distension caused by acetabular loosening after total hip arthroplasty. A rare complication of total hip arthroplasty. Clin. Orthop., 279: 144-148, 1992.

  171. Maxian, T. A.; Brown, T. D.; Pedersen, D. R.; and and Callaghan, J. J.: 3-dimensional sliding/contact computational simulation of total hip wear. Clin. Orthop., 333: 41-50, 1996.

  172. Mintz, L.; Tsao, A.; McCrae, C. R.; Stulberg, S. D.; and and Wright, T.: The arthroscopic evaluation and characteristics of severe polyethylene wear in total knee arthroplasty. Clin. Orthop., 273: 215-222, 1991.

  173. Mirra, J. M.; Amstutz, H. C.; Matos, M.; and and Gold, R.: The pathology of the joint tissues and its clinical relevance in prosthesis failure. Clin. Orthop., 117: 221-240, 1976.

  174. Mohler, C. G.; Callaghan, J. J.; Collis, D. K.; and and Johnston, R. C.: Early loosening of the femoral component at the cement-prosthesis interface after total hip replacement. J. Bone and Joint Surg., 77-A: 1315-1322, Sept. 1995.[Abstract/Free Full Text]

  175. Morrey, B. F., and and Ilstrup, D.: Size of the femoral head and acetabular revision in total hip-replacement arthroplasty. J. Bone and Joint Surg., 71-A: 50-55, Jan. 1989.[Abstract/Free Full Text]

  176. Mulroy, W. F.; Estok, D. M.; and and Harris, W. H.: Total hip arthroplasty with use of so-called second-generation cementing techniques. A fifteen-year-average follow-up study. J. Bone and Joint Surg., 77-A: 1845-1852, Dec. 1995.[Abstract/Free Full Text]

  177. Mundy, C. R.; Altman, A. J.; Gondek, M. D.; and and Bandelin, J. G.: Direct resorption of bone by human monocytes. Science, 196: 1109-1111, 1977.[Abstract/Free Full Text]

  178. Murray, D. W., and and Rushton, N.: Macrophages stimulate bone resorption when they phagocytose particles. J. Bone and Joint Surg., 72-B(6): 988-992, 1990.[Abstract/Free Full Text]

  179. Myers, D. B., and and Palmer, D. G.: Capsular compliance and pressure-volume relationships in normal and arthritic knees. J. Bone and Joint Surg., 54-B(4): 710-716, 1972.

  180. Nashed, R. S.; Becker, D. A.; and and Gustilo, R. B.: Are cementless acetabular components the cause of excess wear and osteolysis in total hip arthroplasty?. Clin. Orthop., 317: 19-28, 1995.

  181. Nasser, S.; Campbell, P. A.; Kilgus, D.; Kossovsky, N.; and and Amstutz, H. C.: Cementless total joint arthroplasty prostheses with titanium-alloy articular surfaces. A human retrieval analysis. Clin. Orthop., 261: 171-185, 1990.

  182. Nolan, J. F., and and Bucknill, T. M.: Aggressive granulomatosis from polyethylene failure in an uncemented knee replacement. J. Bone and Joint Surg., 74-B(1): 23-24, 1992.

  183. Ohashi, T.; Inoue, S.; and Kajikawa, K.: The clinical wear rate of acetabular component accompanied with alumina ceramic head. In Bioceramics, p. 278. Edited by H. Oonishi, H. Aoli, and K. Sawai. St. Louis, Ishiyaku EuroAmerica, 1989.

  184. Okumura, H.; Yamamuro, T.; and Kumar, T.: Socket wear in total hip prosthesis with alumina ceramic head. In Biomedics: Proceedings of the First International Symposium on Bioceramics, p. 258. Edited by H. Oonishi, H. Aoki, and K. Sawai. St. Louis, Ishiyaku EuroAmerica, 1989.

  185. Oonishi, H.; Takayama, Y.; and and Tsuji, E.: Improvement of polyethylene by irradiation in artificial joints. Radiat. Phys. and Chem., 39: 495-504, 1992.

  186. Pedersen, D. R.; Brown, T. D.; Hillis, S. L.; and and Callaghan, J. J.: Prediction of long-term polyethylene wear in total hip arthroplasty, based on early wear measurements made using digital image analysis. J. Orthop. Res., 16: 557-563, 1998.[Medline]

  187. Peters, P. C., Jr.; Engh, G. A.; Dwyer, K. A.; and and Vinh, T. N.: Osteolysis after total knee arthroplasty without cement. J. Bone and Joint Surg., 74-A: 864-876, July 1992.[Abstract/Free Full Text]

  188. Pidhorz, L. E.; Urban, R. M.; Jacobs, J. J.; Sumner, D. R.; and and Galante, J. O.: A quantitative study of bone and soft tissues in cementless porous-coated acetabular components retrieved at autopsy. J. Arthroplasty, 8: 213-225, 1993.[Medline]

  189. Pierson, J. L., and and Harris, W. H.: Extensive osteolysis behind an acetabular component that was well fixed with cement. A case report. J. Bone and Joint Surg., 75-A: 268-271, Feb. 1993.[Free Full Text]

  190. Plante-Bordeneuve, P., and and Freeman, M. A. R.: Tibial high-density polyethylene wear in conforming tibiofemoral prostheses. J. Bone and Joint Surg., 75-B(4): 630-636, 1993.

  191. Ramamurti, B. S.; Bragdon, C. R.; O'Connor, D. O.; Lowenstein, J. P.; Jasty, M.; Estok, D. M.; and and Harris, W. H.: Loci of movement of selected points on the femoral head during normal gait. Three-dimensional computer simulation. J. Arthroplasty, 11: 845-852, 1996.[Medline]

  192. Resnick, D.; Niwayama, G.; and and Coutts, R. D.: Subchondral cysts (geodes) in arthritic disorders: pathologic and radiographic appearance of the hip joint. Am. J. Roentgenol., 128: 799-806, 1977.[Abstract]

  193. Ries, M. D.; Guiney, W., Jr.; and and Lynch, F.: Osteolysis associated with cemented total knee arthroplasty. A case report. J. Arthroplasty, 9: 555-558, 1994.[Medline]

  194. Robinson, E. J.; Mulliken, B. D.; Bourne, R. B.; Rorabeck, C. H.; and and Alvarez, C.: Catastrophic osteolysis in total knee replacement. A report of 17 cases. Clin. Orthop., 321: 98-105, 1995.

  195. Rosenberg, A. G.; Andriacchi, T. P.; Barden, R.; and and Galante, J. O.: Patellar component failure in cementless total knee arthroplasty. Clin. Orthop., 236: 106-114, 1988.

  196. Santavirta, S.; Konttinen, Y. T.; Bergroth, V.; Eskola, A.; Tallroth, K.; and and Lindholm, T. S.: Aggressive granulomatous lesions associated with hip arthroplasty. J. Bone and Joint Surg., 72-A: 252-258, Feb. 1990.[Abstract/Free Full Text]

  197. Santavirta, S.; Konttinen, Y. T.; Hoikka, V.; and and Eskola, A.: Immunopathological response to loose cementless acetabular components. J. Bone and Joint Surg., 73-B(1): 38-42, 1991.

  198. Schmalzried, T. P.; Jasty, M.; and and Harris, W. H.: Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space. J. Bone and Joint Surg., 74-A: 849-863, July 1992.[Abstract/Free Full Text]

  199. Schmalzried, T. P., and and Harris, W. H.: The Harris-Galante porous-coated acetabular component with screw fixation. Radiographic analysis of eighty-three primary hip replacements at a minimum of five years. J. Bone and Joint Surg., 74-A: 1130-1139, Sept. 1992.[Abstract/Free Full Text]

  200. Schmalzried, T. P.; Jasty, M.; and and Harris, W. H.: The role of polyethylene (PE) wear debris in acetabular component loosening. Orthop. Trans., 16: 492-493, 1992.

  201. Schmalzried, T. P.; Kwong, L. M.; Jasty, M.; Sedlacek, R. C.; Haire, T. C.; O'Connor, D. O.; Bragdon, C. R.; Kabo, J. M.; Malcolm, A. J.; and and Harris, W. H.: The mechanism of loosening of cemented acetabular components in total hip arthroplasty. Analysis of specimens retrieved at autopsy. Clin. Orthop., 274: 60-78, 1992.

  202. Schmalzried, T. P., and and Harris, W. H.: Hybrid total hip replacement. A 6.5-year follow-up study. J. Bone and Joint Surg., 75-B(4): 608-615, 1993.

  203. Schmalzried, T. P.; Maloney, W. J.; Jasty, M.; Kwong, L. M.; and and Harris, W. H.: Autopsy studies of the bone-cement interface in well-fixed cemented total hip arthroplasties. J. Arthroplasty, 8: 179-188, 1993.[Medline]

  204. Schmalzried, T. P.; Guttmann, D.; Grecula, M.; and and Amstutz, H. C.: The relationship between the design, position, and articular wear of acetabular components inserted without cement and the development of pelvic osteolysis. J. Bone and Joint Surg., 76-A: 677-688, May 1994.[Abstract/Free Full Text]

  205. Schmalzried, T. P.; Jasty, M.; Rosenberg, A.; and and Harris, W. H.: Polyethylene wear debris and tissue-reactions in knee as compared to hip-replacement prostheses. J. Appl. Biomater., 5: 185-190, 1994.[Medline]

  206. Schmalzried, T. P.; Wessinger, S. J.; Hill, G. E.; and and Harris, W. H.: The Harris-Galante porous acetabular component press-fit without screw fixation. Five-year radiographic analysis of primary cases. J. Arthroplasty, 9: 235-242, 1994.[Medline]

  207. Schmalzried, T. P.; Peters, P. C.; Maurer, B. T.; Bragdon, C. R.; and and Harris, W. H.: Long-duration metal-on-metal total hip arthroplasties with low wear of the articulating surfaces. J. Arthroplasty, 11: 322-331, 1996.[Medline]

  208. Schmalzried, T. P.; Akizuki, K. H.; Fedenko, A. N.; and and Mirra, J.: The role of access of joint fluid to bone in periarticular osteolysis. A report of four cases. J. Bone and Joint Surg., 79-A: 447-452, March 1997.[Free Full Text]

  209. Schmalzried, T. P.; Campbell, P.; Schmitt, A. K.; Brown, I. C.; and and Amstutz, H. C.: Shapes and dimensional characteristics of polyethylene wear particles generated in vivo by total knee replacements compared to total hip replacements. J. Biomed. Mater. Res., 38: 203-210, 1997.[Medline]

  210. Schmalzried, T. P.; Szuszczewicz, E. S.; Northfield, M. R.; Akizuki, K. H.; Frankel, R. E.; Belcher, G.; and and Amstutz, H. C.: Quantitative assessment of walking activity after total hip or knee replacement. J. Bone and Joint Surg., 80-A: 54-59, Jan. 1998.[Abstract/Free Full Text]

  211. Schmalzried, T. P.; Dorey, F. J.; and and McKellop, H.: Commentary. The multifactorial nature of polyethylene wear in vivo. J. Bone and Joint Surg., 80-A: 1234-1242, Aug. 1998.[Free Full Text]

  212. Schulte, K. R.; Callaghan, J. J.; Kelley, S. S.; and and Johnston, R. C.: The outcome of Charnley total hip arthroplasty with cement after a minimum twenty-year follow-up. The results of one surgeon. J. Bone and Joint Surg., 75-A: 961-975, July 1993.[Abstract/Free Full Text]

  213. Seedhom, B.; Dowson, D.; and and Wright, V.: Wear of solid phase formed high density polyethylene in relation to the life of artificial hips and knees. Wear, 24: 35-51, 1973.

  214. Shanbhag, A. S.; Jacobs, J. J.; Glant, T. T.; Gilbert, J. L.; Black, J.; and and Galante, J. O.: Composition and morphology of wear debris in failed uncemented total hip replacement. J. Bone and Joint Surg., 76-B(1): 60-67, 1994.[Abstract/Free Full Text]

  215. Shanbhag, A. S.; Hasselman, C. T.; and and Rubash, H. E.: Inhibition of wear debris mediated osteolysis in a canine total hip arthroplasty model. Clin. Orthop., 344: 33-43, 1997.

  216. Shaver, S. M.; Brown, T. D.; Hillis, S. L.; and and Callaghan, J. J.: Digital edge-detection measurement of polyethylene wear after total hip arthroplasty. J. Bone and Joint Surg., 79-A: 690-700, May 1997.

  217. Shinto, Y.; Uchida, A.; Yoshikawa, H.; Araki, N.; Kato, T.; and and Ono, K.: Inguinal lymphadenopathy due to metal release from a prosthesis. A case report. J. Bone and Joint Surg., 75-B(2): 266-269, 1993.

  218. Silverton, C. D.; Jacobs, J. J.; Rosenberg, A. G.; Kull, L.; Conley, A.; and and Galante, J. O.: Complications of a cable grip system. J. Arthroplasty, 11: 400-404, 1996.[Medline]

  219. Simkin, P. A.: Physiology of normal and abnormal synovium. Sem. Arthrit. and Rheumat., 21: 179-183, 1991.

  220. Smith, J., and and Liu, C.: Stresses due to tangential and normal loads on an elastic solid application to some contact stress problems. J. Appl. Mech., 20: 157-166, 1953.

  221. Smith, S. E., and and Harris, W. H.: Total hip arthroplasty performed with insertion of the femoral component with cement and the acetabular component without cement. Ten to thirteen-year results. J. Bone and Joint Surg., 79-A: 1827-1833, Dec. 1997.[Abstract/Free Full Text]

  222. Stulberg, S. D.; Stulberg, B. N.; Hamati, Y.; and and Tsao, A.: Failure mechanisms of metal-backed patellar components. Clin. Orthop., 236: 88-105, 1988.

  223. Sutula, L. C.; Collier, J. P.; Saum, K. A.; Currier, B. H.; Currier, J. H.; Sanford, W. M.; Mayor, M. B.; Wooding, R. E.; Sperling, D. K.; Williams, I. R.; Kasprzak, D. J.; and and Surprenant, V. A.: Impact of gamma sterilization on clinical performance of polyethylene in the hip. Clin. Orthop., 319: 28-40, 1995.

  224. Svensson, O.; Mathiesen, E. B.; Reinholt, F. P.; and and Blomgren, G.: Formation of a fulminant soft-tissue pseudotumor after uncemented hip arthroplasty. A case report. J. Bone and Joint Surg., 70-A: 1238-1242, Sept. 1988.[Free Full Text]

  225. Sychterz, C. J.; Moon, K. H.; Hashimoto, Y.; Terefenko, K. M.; Engh, C. A., Jr.; and and Bauer, T. W.: Wear of polyethylene cups in total hip arthroplasty. A study of specimens retrieved post mortem. J. Bone and Joint Surg., 78-A: 1193-1200, Aug. 1996.[Abstract/Free Full Text]

  226. Sychterz, C. J.; Engh, C. A., Jr.; Shah, N.; and and Engh, C. A., Sr.: Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J. Bone and Joint Surg., 79-A: 1040-1046, July 1997.[Abstract/Free Full Text]

  227. Sychterz, C. J.; Shan, N.; and and Engh, C. A.: Radiographic evaluation of femoral head penetration into polyethylene liners: Hylamer versus Enduron. Trans. Orthop. Res. Soc., 22: 140, 1997.

  228. Takagi, M.; Konttinen, Y. T.; Lindy, O.; Sorsa, T.; Kurvinen, H.; Suda, A.; and and Santavirta, S.: Gelatinase/type IV collagenases in the loosening of total hip replacement endoprostheses. Clin. Orthop., 306: 136-144, 1994.

  229. Tanzer, M.; Maloney, W. J.; Jasty, M.; and and Harris, W. H.: The progression of femoral cortical osteolysis in association with total hip arthroplasty without cement. J. Bone and Joint Surg., 74-A: 404-410, March 1992.[Abstract/Free Full Text]

  230. Tipper, J. L.; Ingham, E.; Hailey, J. L.; Besong, A. A.; Stone, M.; Wroblewski, B. M.; and and Fisher, J.: Quantitative comparison of polyethylene wear debris, wear rate and head damage in retrieved hip prostheses. Trans. Orthop. Res. Soc., 22: 355, 1997.

  231. Tsao, A.; Mintz, L.; McRae, C. R.; Stulberg, S. D.; and and Wright, T.: Failure of the porous-coated anatomic prosthesis in total knee arthroplasty due to severe polyethylene wear. J. Bone and Joint Surg., 75-A: 19-26, Jan. 1993.[Abstract/Free Full Text]

  232. Urban, R. M.; Jacobs, J. J.; Gilbert, J. L.; and and Galante, J. O.: Migration of corrosion products from modular hip prostheses. Particle microanalysis and histopathological findings. J. Bone and Joint Surg., 76-A: 1345-1359, Sept. 1994.[Abstract/Free Full Text]

  233. Urban, R. M.; Jacobs, J. J.; Sumner, D. R.; Peters, C. L.; Voss, F. R.; and and Galante, J. O.: The bone-implant interface of femoral stems with non-circumferential porous coating. A study of specimens retrieved at autopsy. J. Bone and Joint Surg., 78-A: 1068-1081, July 1996.[Abstract/Free Full Text]

  234. Van Der Vis, H. M.; Aspenberg, P.; Marti, R. K.; Tigchelaar, W.; and and Van Noorden, C. J. F.: Fluid pressure causes bone resorption in a rabbit model of prosthetic loosening. Clin. Orthop., 350: 201-208, 1998.

  235. Volz, R. G., and and Wilson, R. J.: Factors affecting the mechanical stability of the cemented acetabular component in total hip replacement. J. Bone and Joint Surg., 59-A: 501-504, June 1977.[Abstract/Free Full Text]

  236. Wang, A.; Essner, A.; Polineni, V.; Sun, D.; Stark, C.; and Dumbleton, J.: Wear mechanisms and wear testing of ultra-high molecular weight polyethylene in total joint replacements. In Polyethylene Wear in Orthopaedic Implants Workshop, pp. 4-18. Minneapolis, Society for Biomaterials, 1997.

  237. Wasielewski, R. C.; Galante, J. O.; Leighty, R. M.; Natarajan, R. N.; and and Rosenberg, A. G.: Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty. Clin. Orthop., 299: 31-43, 1994.

  238. Waugh, W.: John Charnley, the Man and the Hip. New York, Springer, 1990.

  239. Weightman, B.; Swanson, S. A. V.; Isaac, G.; and and Wroblewski, B. M.: Polyethylene wear from retrieved acetabular cups. J. Bone and Joint Surg., 73-B(5): 806-810, 1991.

  240. Whiteside, L. A.: Effect of porous-coating configuration on tibial osteolysis after total knee arthroplasty. Clin. Orthop., 321: 92-97, 1995.

  241. Willert, H.-G.: Reactions of the articular capsule to wear products of artificial joint prostheses. J. Biomed. Mater. Res., 11: 157-164, 1977.[Medline]

  242. Wixon, R. L.; Stulberg, S. D.; and and Mehlhoff, M.: Total hip replacement with cemented, uncemented, and hybrid prostheses. A comparison of clinical and radiographic results at two to four years. J. Bone and Joint Surg., 73-A: 257-270, Feb. 1991.[Abstract/Free Full Text]

  243. Woolson, S. T., and and Murphy, M. G.: Wear of the polyethylene of Harris-Galante acetabular components inserted without cement. J. Bone and Joint Surg., 77-A: 1311-1314, 1995.[Abstract/Free Full Text]

  244. Woolson, S. T.; Milbauer, J. P.; Bobyn, J. D.; Yue, S.; and and Maloney, W. J.: Fatigue fracture of a forged cobalt-chromium-molybdenum femoral component inserted with cement. A report of ten cases. J. Bone and Joint Surg., 79-A: 1842-1848, Dec. 1997.[Abstract/Free Full Text]

  245. Wright, T. M., and and Bartel, D. L.: The problem of surface damage in polyethylene total knee components. Clin. Orthop., 205: 67-74, 1986.

  246. Wright, T. M.; Rimnac, C. M.; Faris. P. M.; and and Bansal, M.: Analysis of surface damage in retrieved carbon fiber-reinforced and plain polyethylene tibial components from posterior stabilized total knee replacements. J. Bone and Joint Surg., 70-A: 1312-1319, Oct. 1988.[Abstract/Free Full Text]

  247. Wright, T. M.; Rimnac, C. M.; Stulberg, S. D.; Mintz, L.; Tsao, A. K.; Klein, R. W.; and and McCrae, C.: Wear of polyethylene in total joint replacements. Observations from retrieved PCA knee implants. Clin. Orthop., 276: 126-134, 1992.

  248. Wright, T., and Goodman, S.: Implant Wear: The Future of Total Joint Replacement. Rosemont, Illinois, American Academy of Orthopaedic Surgeons, 1995.

  249. Wroblewski, B. M.: Direction and rate of socket wear in Charnley low-friction arthroplasty. J. Bone and Joint Surg., 67-B(5): 757-761, 1985.

  250. Wroblewski, B. M.: 15–21-year results of the Charnley low-friction arthroplasty. Clin. Orthop., 211: 30-35, 1986.

  251. Wroblewski, B. M., and and Siney, P. D.: Charnley low-friction arthroplasty in the young patient. Clin. Orthop., 285: 45-47, 1992.

  252. Wroblewski, B. M.; Taylor, G. W.; and and Siney, P.: Charnley low-friction arthroplasty: 19- to 25-year results. Orthopedics, 15: 421-424, 1992.[Medline]

  253. Wroblewski, B. M.; Siney, P. D.; Dowson, D.; and and Collins, S. N.: Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups. J. Bone and Joint Surg., 78-B(2): 280-285, 1996.

  254. Yamaguchi, M.; Bauer, T. W.; and and Hashimoto, Y.: Three-dimensional analysis of multiple wear vectors in retrieved acetabular cups. J. Bone and Joint Surg., 79-A: 1539-1544, Oct. 1997.[Abstract/Free Full Text]

  255. Yao, J.; Cs-Szabó, G.; Jacobs, J. J.; Kuettner, K. E.; and and Glant, T. T.: Suppression of osteoblast function by titanium particles. J. Bone and Joint Surg., 79-A: 107-112, Jan. 1997.[Abstract/Free Full Text]

  256. Zicat, B.; Engh, C. A.; and and Gokcen, E.: Patterns of osteolysis around total hip components inserted with and without cement. J. Bone and Joint Surg., 77-A: 432-439, March 1995.[Abstract/Free Full Text]


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Facebook Facebook   Add to Technorati Technorati   Add to Twitter Twitter    What's this?


This article has been cited by other articles:


Home page
JBJSHome page
R. G. Bitsch, T. Loidolt, C. Heisel, S. Ball, and T. P. Schmalzried
Reduction of Osteolysis with Use of Marathon Cross-Linked Polyethylene. A Concise Follow-up, at a Minimum of Five Years, of a Previous Report
J. Bone Joint Surg. Am., July 1, 2008; 90(7): 1487 - 1491.
[Abstract] [Full Text] [PDF]


Home page
Ann Rheum DisHome page
M H A Malik, F Jury, A Bayat, W E R Ollier, and P R Kay
Genetic susceptibility to total hip arthroplasty failure: a preliminary study on the influence of matrix metalloproteinase 1, interleukin 6 polymorphisms and vitamin D receptor
Ann Rheum Dis, August 1, 2007; 66(8): 1116 - 1120.
[Abstract] [Full Text] [PDF]


Home page
J Am Acad Orthop SurgHome page
D. D.R. Naudie, D. J. Ammeen, G. A. Engh, and C. H. Rorabeck
Wear and Osteolysis Around Total Knee Arthroplasty
J. Am. Acad. Ortho. Surg., January 1, 2007; 15(1): 53 - 64.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
M. W. Hanna and T. S. Thornhill
Thigh Mass and Lytic Diaphyseal Femoral Lesion Associated with Polyethylene Wear After Hybrid Total Knee Arthroplasty. A Case Report
J. Bone Joint Surg. Am., November 1, 2006; 88(11): 2473 - 2478.
[Full Text] [PDF]


Home page
JBJSHome page
J. J. Jacobs, N. J. Hallab, R. M. Urban, and M. A. Wimmer
Wear Particles
J. Bone Joint Surg. Am., April 1, 2006; 88(suppl_2): 99 - 102.
[Abstract] [Full Text] [PDF]


Home page
J Bone Joint Surg BrHome page
A. J. Price, A. Short, C. Kellett, D. Beard, H. Gill, H. Pandit, C. A. F. Dodd, and D. W. Murray
Ten-year in vivo wear measurement of a fully congruent mobile bearing unicompartmental knee arthroplasty
J Bone Joint Surg Br, November 1, 2005; 87-B(11): 1493 - 1497.
[Abstract] [Full Text] [PDF]


Home page
J Bone Joint Surg BrHome page
J. Arora and A. C. Ogden
Osteolysis in a surface-cemented, primary, modular Freeman-Samuelson total knee replacement
J Bone Joint Surg Br, November 1, 2005; 87-B(11): 1502 - 1506.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
F.H.R. De Man, W. Tigchelaar, R.K. Marti, C.J.F. Van Noorden, and H.M. Van der Vis
Effects of Mechanical Compression of a Fibrous Tissue Interface on Bone with or without High-Density Polyethylene Particles in a Rabbit Model of Prosthetic Loosening
J. Bone Joint Surg. Am., July 1, 2005; 87(7): 1522 - 1533.
[Abstract] [Full Text] [PDF]


Home page
J Bone Joint Surg BrHome page
A. Lakdawala, S. Todo, and G. Scott
The significance of surface changes on retrieved femoral components after total knee replacement
J Bone Joint Surg Br, June 1, 2005; 87-B(6): 796 - 799.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
A. Alhadlaq and J. J. Mao
Tissue-Engineered Osteochondral Constructs in the Shape of an Articular Condyle
J. Bone Joint Surg. Am., May 1, 2005; 87(5): 936 - 944.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
L. Bradford, D. A. Baker, J. Graham, A. Chawan, M. D. Ries, and L. A. Pruitt
Wear and Surface Cracking in Early Retrieved Highly Cross-linked Polyethylene Acetabular Liners
J. Bone Joint Surg. Am., June 1, 2004; 86(6): 1271 - 1282.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
C. Heisel, M. Silva, M. A. dela Rosa, and T. P. Schmalzried
Short-Term in Vivo Wear of Cross-Linked Polyethylene
J. Bone Joint Surg. Am., April 1, 2004; 86(4): 748 - 751.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
M. Silva, K. H. Lee, C. Heisel, M. A. dela Rosa, and T. P. Schmalzried
The Biomechanical Results of Total Hip Resurfacing Arthroplasty
J. Bone Joint Surg. Am., January 1, 2004; 86(1): 40 - 46.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
H. Sugimoto, I. Hirose, E. Miyaoka, A. Fujita, Y. Kinebuchi, W. Yamamoto, and Y.-i. Itoh
Low-Field-Strength MR Imaging of Failed Hip Arthroplasty: Association of Femoral Periprosthetic Signal Intensity with Radiographic, Surgical, and Pathologic Findings
Radiology, December 1, 2003; 229(3): 718 - 723.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
C. Heisel, M. Silva, and T. P. Schmalzried
Bearing Surface Options for Total Hip Replacement in Young Patients
J. Bone Joint Surg. Am., July 3, 2003; 85(7): 1366 - 1379.
[Full Text] [PDF]


Home page
JBJSHome page
R. L. Barrack and T. P. Schmalzried
Impingement and Rim Wear Associated with Early Osteolysis After a Total Hip Replacement : A Case Report
J. Bone Joint Surg. Am., July 9, 2002; 84(7): 1218 - 1220.
[Full Text] [PDF]


Home page
J Am Acad Orthop SurgHome page
P. E. Beaule, J. M. Matta, and J. W. Mast
Hip Arthrodesis: Current Indications and Techniques
J. Am. Acad. Ortho. Surg., July 1, 2002; 10(4): 249 - 258.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
S. K.T. Puloski, R. W. McCalden, S. J. MacDonald, C. H. Rorabeck, and R. B. Bourne
Tibial Post Wear in Posterior {-}Stabilized Total Knee Arthroplasty : An Unrecognized Source of Polyethylene Debris
J. Bone Joint Surg. Am., March 1, 2001; 83(3): 390 - 390.
[Abstract] [Full Text]


Home page
JBJSHome page
S. A. Mikulak, O. M. Mahoney, M. A. delaRosa, and T. P. Schmalzried
Loosening and Osteolysis with the Press-Fit Condylar Posterior-Cruciate-Substituting Total Knee Replacement
J. Bone Joint Surg. Am., March 1, 2001; 83(3): 398 - 398.
[Abstract] [Full Text]


Home page
ANN INTERN MEDHome page
D. T. Felson, R. C. Lawrence, M. C. Hochberg, T. McAlindon, P. A. Dieppe, M. A. Minor, S. N. Blair, B. M. Berman, J. F. Fries, M. Weinberger, et al.
Osteoarthritis: New Insights. Part 2: Treatment Approaches
Ann Intern Med, November 7, 2000; 133(9): 726 - 737.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
G. LEWIS
Commentary - Design Issues in Clinical Studies of the in Vivo Volumetric Wear Rate of Polyethylene Bearing Components
J. Bone Joint Surg. Am., February 1, 2000; 82(2): 281 - 87.
[Full Text]


This Article
Right arrow Extract Freely available
Right arrow Full Text (PDF) Free
Right arrow Letters to the Editor: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when Letters to the Editor are posted
Right arrow Alert me if a correction is posted
Services
Right arrow E-mail this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrow Rights and Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by SCHMALZRIED, T. P.
Right arrow Articles by CALLAGHAN, J. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by SCHMALZRIED, T. P.
Right arrow Articles by CALLAGHAN, J. J.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Facebook   Add to Technorati   Add to Twitter  
What's this?