The Journal of Bone and Joint Surgery (American). 2007;89:2779-2786.
doi:10.2106/JBJS.G.00043
© 2007 The Journal of Bone and Joint Surgery, Inc.
This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow CME: Take the exams for this article:
Adult Hip Reconstruction Test 27: Winter 2008 (publication date February 15...
CME 4: October, November, December 2007 (publication date January 4, 2008; ...
Right arrow [FREE Spanish Translation]
Right arrow Letters to the Editor: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when Letters to the Editor are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrowReprints and Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Jacobs, C. A.
Right arrow Articles by McKellop, H.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Jacobs, C. A.
Right arrow Articles by McKellop, H.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Technorati  
What's this?

Current Concepts Review

Clinical Performance of Highly Cross-Linked Polyethylenes in Total Hip Arthroplasty

Cale A. Jacobs, PhD1, Christian P. Christensen, MD1, A. Seth Greenwald, DPhil(Oxon)2 and Harry McKellop, PhD3

1 Lexington Clinic, 1221 South Broadway, Lexington, KY 40504. E-mail address for C.A. Jacobs: cjaco{at}lexclin.com
2 Orthopaedic Research Laboratories, Lutheran Hospital, 1730 West 25th Street, Cleveland, OH 44113
3 Department of Orthopaedic Surgery, University of California at Los Angeles Orthopaedic Hospital, 2400 South Flower Street, Los Angeles, CA 90007

Disclosure: The authors did not receive any outside funding or grants in support of their research for or preparation of this work. One or more of the authors, or a member of his or her immediate family, received, in any one year, payments or other benefits in excess of $10,000 or a commitment or agreement to provide such benefits from commercial entities (DePuy Orthopaedics and Biomet Orthopedics). Commercial entities (DePuy Orthopaedics, Biomet Orthopedics, and Smith and Nephew) paid or directed in any one year, or agreed to pay or direct, benefits in excess of $10,000 to a research fund, foundation, division, center, clinical practice, or other charitable or nonprofit organization with which one or more of the authors, or a member of his or her immediate family, is affiliated or associated.


Aseptic loosening secondary to wear-debris-induced osteolysis has been identified as the leading cause of late failure of total hip arthroplasty.

Highly cross-linked polyethylene acetabular liners were developed as one approach to reducing this wear.

Preclinical laboratory wear testing showed a number of cross-linked polyethylenes to have dramatically less wear than the polyethylene that had been in use for several decades.

After the initial bedding-in phase (one to two years), the percent reductions in the wear rate, as indicated by the amount of penetration of the head into the socket evident on serial radiographs, have been comparable with what was predicted from preclinical hip-simulator testing of the highly cross-linked polyethylenes.

To our knowledge, there have been no reports of clinically relevant osteolysis that was clearly attributable to wear of a highly cross-linked polyethylene acetabular liner. However, the clinical performance of these materials should be closely monitored with long-term follow-up.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Technorati Technorati    What's this?