Ceramic Failure After Total Hip Arthroplasty with an Alumina-on-Alumina Bearing
Youn-Soo Park, MD1,
Sung-Kwan Hwang, MD2,
Won-Sik Choy, MD3,
Yong-Sik Kim, MD4,
Young-Wan Moon, MD1 and
Seung-Jae Lim, MD1
1 Department of Orthopedic Surgery, Sungkyunkwan University School of Medicine,
Samsung Medical Center, 50 Ilwon-Dong, Kangnam-Ku, Seoul 135-710, South Korea.
E-mail address for Y.-S. Park:
yspark{at}smc.samsung.co.kr
2 Department of Orthopedic Surgery, Yonsei University Wonju College of Medicine,
Wonju Christian Hospital, 162 Ilsan-Dong, Wonju, Gangwon-Do 220-701, South
Korea
3 Department of Orthopedic Surgery, Eulji Medical College, Eulji University
Hospital, 1306 Dunsan-Dong, Seo-Ku, Daejon 302-799, South Korea
4 Department of Orthopedic Surgery, KangNam St. Mary's Hospital, The Catholic
University of Korea, 505 Banpo-Dong, Seocho-Ku, Seoul 137-040, South
Korea
Investigation performed at the Department of Orthopedic Surgery,
Sungkyunkwan University School of Medicine, Samsung Medical Center, Seoul,
South Korea
The authors did not receive grants or outside funding in support of their
research for or preparation of this manuscript. They did not receive payments
or other benefits or a commitment or agreement to provide such benefits from a
commercial entity. No commercial entity paid or directed, or agreed to pay or
direct, any benefits to any research fund, foundation, educational
institution, or other charitable or nonprofit organization with which the
authors are affiliated or associated.
Background: The mechanical properties of alumina ceramic, now in its
third generation, have been markedly improved through the evolution of design
features and manufacturing processes and the introduction of proof-testing.
Nonetheless, because of the lack of ductility of alumina ceramic, there is
concern regarding the risk of fracture during insertion or in vivo use. The
purpose of the present study was to present a multicenter review of primary
total hip arthroplasties performed with use of a polyethylene-ceramic
composite liner combined with a ceramic femoral head, with particular
attention to failure of the ceramic bearing.
Methods: We evaluated 357 primary total hip arthroplasties that had
been performed in 319 patients with use of a contemporary alumina-on-alumina
bearing design incorporating a polyethylene-ceramic composite liner within a
titanium-alloy shell coupled with a 28-mm-diameter ceramic femoral head. The
procedures were performed at four participating centers from 1998 to 2001.
Ceramic failure without trauma occurred in six hips (1.7%). All of these hips
were revised, and the retrieved alumina implants were examined by means of
visual inspection and scanning electron microscopy equipped with
energy-dispersive x-ray spectrometry.
Results: Two femoral heads fractured during the first postoperative
year, and four alumina liners fractured after an average of 36.8 months in
vivo. All four of the explanted alumina liners revealed evidence of rim
contact with the metal neck of the femoral component. Composition analysis
confirmed that surface-stain materials were titanium particles transferred
from the femoral component.
Conclusions: Despite the theoretical improvement in the fracture
toughness of a polyethylene-alumina composite liner, a relatively high rate of
catastrophic ceramic bearing surface failure was still observed at the time of
short-term follow-up. This finding prompted us to discontinue the use of this
type of alumina bearing design.
Level of Evidence: Therapeutic Level IV. See Instructions
to Authors for a complete description of levels of evidence.

CiteULike Connotea Del.icio.us Facebook Technorati Twitter What's this?
This article has been cited by other articles:

|
 |

|
 |
 
J. J. Callaghan, J. M. Cuckler, J. I. Huddleston, and J. O. Galante
How have alternative bearings (such as metal-on-metal, highly cross-linked polyethylene, and ceramic-on-ceramic) affected the prevention and treatment of osteolysis?
J. Am. Acad. Ortho. Surg.,
July 1, 2008;
16(suppl_1):
S33 - S38.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K.-H. Koo, Y.-C. Ha, W. H. Jung, S.-R. Kim, J. J. Yoo, and H. J. Kim
Isolated Fracture of the Ceramic Head After Third-Generation Alumina-on-Alumina Total Hip Arthroplasty
J. Bone Joint Surg. Am.,
February 1, 2008;
90(2):
329 - 336.
[Abstract]
[Full Text]
[PDF]
|
 |
|
Letters to the Editor:
Read all Letters to the Editor
- Lateralization of hip centre and low offset can lead to ceramic failure
- Jibanananda Satpathy, et al.
- JBJS Online, 15 May 2006
[Full text]
- Good Results of Ceramic-Ceramic THA
- Daniel Reis, FRCS
- JBJS Online, 5 Jun 2006
[Full text]
- Dr. Park et al. respond to Dr. Satpathy
- Youn-Soo Park, M.D., et al.
- JBJS Online, 6 Jul 2006
[Full text]
- Dr. Park et al. respond to Dr. Reis
- Youn-Soo Park, M.D., et al.
- JBJS Online, 6 Jul 2006
[Full text]
- Failure Mechanisms of Ceramic Total Hip Arthroplasty
- Francesco M. Benazzo
- JBJS Online, 25 Sep 2006
[Full text]
- Dr. Park and Colleagues Respond to Dr. Benazzo
- Youn-Soo Park, M.D., et al.
- JBJS Online, 30 Oct 2006
[Full text]
|