Three-Dimensional Tibiofemoral Articular Contact Kinematics of a Cruciate-Retaining Total Knee Arthroplasty
Guoan Li, PhD1,
Jeremy Suggs, MS1,
George Hanson, BS1,
Sridhar Durbhakula, MD2,
Todd Johnson, PhD3 and
Andrew Freiberg, MD4
1 Massachusetts General Hospital, 55 Fruit Street, GRJ 1215, Boston, MA 02114.
E-mail address for G. Li:
gli1{at}partners.org
2 6080 Falls Road, Suite 203, Baltimore, MD 21209
3 P.O. Box 708, 1800 West Center Street, Warsaw, IN 46581
4 Massachusetts General Hospital, Yawkey Center for Outpatient Care, 32 Fruit
Street, YAW-3-3B, Boston, MA 02114
Investigation performed at the Bioengineering Laboratory, Department of
Orthopaedic Surgery, Massachusetts General Hospital/Harvard Medical School,
Boston, Massachusetts
NOTE: The authors thank Dr. Louis DeFrate, Ramprasad
Papannagari, and Elizabeth DeSouza for their technical assistance.
In support of their research for or preparation of this manuscript, one or
more of the authors received grants or outside funding from Zimmer. In
addition, one or more of the authors received payments or other benefits or a
commitment or agreement to provide such benefits from a commercial entity
(Zimmer). Also, a commercial entity (Zimmer) paid or directed, or agreed to
pay or direct, benefits to a research fund, foundation, educational
institution, or other charitable or nonprofit organization with which the
authors are affiliated or associated.
Background: Accurate knowledge of the location of tibiofemoral
articular contact following total knee arthroplasty is important in order to
understand polyethylene wear and the mechanisms of component failure. The
present study was performed to determine the three-dimensional tibiofemoral
articular contact patterns of a posterior cruciate ligament-retaining total
knee replacement during in vivo weight-bearing flexion.
Methods: Nine osteoarthritic patients who were managed with a single
design of a posterior cruciate ligament-retaining total knee implant were
investigated with the use of an innovative dual orthogonal fluoroscopic
imaging system. The position of the components during in vivo weight-bearing
flexion was measured from full extension to maximum flexion in 15°
intervals. Tibiofemoral articular contact was determined by the overlap of the
tibiofemoral articular surfaces. The centroid of the surface intersection was
used to report the point of contact location. The average tibiofemoral contact
points on both the medial and lateral tibial component surfaces were reported
as a function of flexion.
Results: The average maximum weight-bearing flexion angle was
113.3° ± 13.1° (range, 96° to 138°). In the
anteroposterior direction, the contact location was relatively constant in the
medial compartment and moved posteriorly by 5.6 mm in the lateral compartment
as the knee flexed from full extension to 90° of flexion. The range of the
contact location in the mediolateral direction was 3.7 mm in the medial
compartment and 4.8 mm in the lateral compartment. For both compartments,
posterior translation of the contact point was significant from 90° to
maximum flexion, but the contact point at maximum flexion was not observed to
reach the posterior edge of the polyethylene tibial insert articular
surface.
Conclusions: While the minimum anteroposterior translation of the
contact point on the medial side might be interpreted as a medial pivot
rotation during knee flexion, the contact point did move in the mediolateral
direction with flexion. Beyond 90°, both medial and lateral contact points
were shown to move posteriorly but stopped before reaching the posterior edge
of the polyethylene tibial insert articular surface. It seemed that the
current component design did not allow the femoral condyle to roll off the
polyethylene edge at high degrees of flexion because of the geometry at the
posterior lip.
Clinical Relevance: These three-dimensional tibiofemoral contact
data may provide new insight for determining polyethylene tibial insert wear
patterns in vivo and for designing the articulating surfaces by accounting for
contact location in both the anteroposterior and mediolateral directions.

CiteULike Connotea Del.icio.us Technorati What's this?
Letters to the Editor:
Read all Letters to the Editor
- Three-Dimensional Tibiofemoral Articular Contact Kinematics of a Cruciate-Retaining Total Knee Arthr
- Joby John
- JBJS Online, 1 Mar 2006
[Full text]
- Dr. Li responds to Dr. John
- Guoan Li, Ph.D.
- JBJS Online, 28 Mar 2006
[Full text]
|