Accuracy in the Measurement of Compartment Pressures: A Comparison of Three Commonly Used Devices
Antony R. Boody, MD1 and
Montri D. Wongworawat, MD1
1 Department of Orthopaedic Surgery, Loma Linda University, 11406 Loma Linda
Drive, Suite 218, Loma Linda, CA 92354. E-mail address for M.D. Wongworawat:
wongwora{at}usc.edu
Investigation performed at the Department of Orthopaedic Surgery, Loma
Linda University, Loma Linda, California
The authors did not receive grants or outside funding in support of their
research or preparation of this manuscript. They did not receive payments or
other benefits or a commitment or agreement to provide such benefits from a
commercial entity. No commercial entity paid or directed, or agreed to pay or
direct, any benefits to any research fund, foundation, educational
institution, or other charitable or nonprofit organization with which the
authors are affiliated or associated.
Background: In situations in which accurate physical diagnosis is
inconclusive, an objective method for measuring compartment pressure can aid
in the diagnosis of compartment syndrome. Previous studies have compared
measurement devices with each other but not with an accurately determined gold
standard. The purpose of the present study was to devise a reproducible in
vitro model of compartment pressure and to compare commonly used measurement
devices in order to determine their accuracy.
Methods: With a graduated cylinder being used to generate a known
pressure, freshly harvested ovine muscle was placed into a chamber for
testing. The cylinder was incrementally filled with saline solution (in
fifty-five steps), and measurements of tissue pressure were obtained with use
of the Stryker Intracompartmental Pressure Monitor System, an arterial line
manometer, and the Whitesides apparatus. Each device was tested with a
straight needle, a side-port needle, and a slit catheter, for a total of nine
setups in all. Five trials were done with each setup. Control pressures were
calculated on the basis of the height of the saline solution column (test
range, 0.13 to 10.80 kPa). Multiple regression analysis was used to compare
measured tissue pressures with calculated control pressures.
Results: Most methods demonstrated excellent correlation
(R2 > 0.95) between calculated and measured pressures. The
arterial line manometer with the slit catheter showed the best correlation
(R2 = 0.9978), and the Whitesides apparatus with the side-port
needle showed the worst (R2 = 0.9115). Furthermore, the Stryker
system with the side-port needle demonstrated the least constant bias (+0.06
kPa). Straight needles tended to overestimate pressure. Two of the three
needle configurations involving the Whitesides apparatus overestimated
pressure. The data for the Whitesides methods had the highest standard errors,
showing clinically unacceptable scatter.
Conclusion: Side-port needles and slit catheters are more accurate
than straight needles are. The arterial line manometer is the most accurate
device. The Stryker device is also very accurate. The Whitesides manometer
apparatus lacks the precision needed for clinical use.
Clinical Relevance: When physical examination findings are
inconclusive, accurate measurement of compartment pressures can aid in timely
management and can minimize patient morbidity. Measurement should be done with
use of the most accurate technique available.

CiteULike Connotea Del.icio.us Technorati What's this?
Related articles in JBJS:
- Accuracy in the Measurement of Compartment Pressures: A Comparison of Three Commonly Used Devices
- Michael R. Barnes
JBJS 2006 88: 1165.
[Extract]
[Full Text]
Letters to the Editor:
Read all Letters to the Editor
- Measurement of Compartment Pressures
- Michael R Barnes
- JBJS Online, 12 Dec 2005
[Full text]
- Drs. Wongworawat and Boody respond to Dr. Barnes
- Montri D. Wongworawat, et al.
- JBJS Online, 12 Dec 2005
[Full text]
|