Shelf Age of the Polyethylene Tibial Component and Outcome of Unicondylar Knee Arthroplasty
Matthew B. Collier, MS1,
C. Anderson Engh, Jr., MD1 and
Gerard A. Engh, MD1
1 Anderson Orthopaedic Research Institute, P.O. Box 7088, Alexandria, VA 22307.
E-mail address for M.B. Collier:
collier{at}aori.org
Investigation performed at the Anderson Orthopaedic Research Institute,
Alexandria, Virginia
The authors did not receive grants or outside funding in support of their
research or preparation of this manuscript. They did not receive payments or
other benefits or a commitment or agreement to provide such benefits from a
commercial entity. A commercial entity (INOVA Health Care Services, Falls
Church, Virginia) paid or directed, or agreed to pay or direct, benefits to a
research fund, foundation, educational institution, or other charitable or
nonprofit organization with which the authors are affiliated or
associated.
Background: A recent report linked shelf-aging of unicondylar
polyethylene tibial components with accelerated fatigue wear and rapid
failure. We retrospectively reviewed our experiences with another unicondylar
knee system to investigate the relationship between the shelf age of the
polyethylene components and clinical outcome.
Methods: One hundred Single Compartment Replacement unicompartmental
knee arthroplasties were performed with cement between 1990 and 1996. The
median shelf age for the polyethylene inserts was 1.7 years. At the time of
the review, four knees had been lost to follow-up, sixteen were in patients
who had died, and nineteen had been revised. Sixty-one knee replacements
remained in situ at a mean and standard deviation of 8 ± 2 years
postoperatively.
Results: With revision as the end point, Kaplan-Meier survivorship
analysis demonstrated a six-year rate of survival of 96% when the shelf age of
the insert was less than the median shelf age but only 71% when the median
shelf age was exceeded. Fatigue wear was identified on all retrieved
components except for two that had shelf ages under one year and had been
revised within three years. At four to six years of clinical follow-up, the
group of knees in which the median shelf age had been exceeded also had poorer
Knee Society knee scores and function scores.
Conclusions: Aging on the shelf accelerated fatigue failure of
polyethylene inserts sterilized with gamma irradiation in air and compromised
our intermediate-term clinical outcomes with this unicondylar knee replacement
system. Attention to the relationship between the shelf age and the clinical
performance of the polyethylene component continues to be warranted with
unicondylar knee arthroplasty implants.
Level of Evidence: Therapeutic study, Level III-2
(retrospective cohort study). See Instructions to Authors for a complete
description of levels of evidence.

CiteULike Connotea Del.icio.us Facebook Technorati Twitter What's this?
This article has been cited by other articles:

|
 |

|
 |
 
R. D. Crowninshield and O. K. Muratoglu
How have new sterilization techniques and new forms of polyethylene influenced wear in total joint replacement?
J. Am. Acad. Ortho. Surg.,
July 1, 2008;
16(suppl_1):
S80 - S85.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Borus and T. Thornhill
Unicompartmental Knee Arthroplasty
J. Am. Acad. Ortho. Surg.,
January 1, 2008;
16(1):
9 - 18.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. B. Collier, C. A. Engh Jr., J. P. McAuley, and G. A. Engh
Factors Associated with the Loss of Thickness of Polyethylene Tibial Bearings After Knee Arthroplasty
J. Bone Joint Surg. Am.,
June 1, 2007;
89(6):
1306 - 1314.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. J. Archibeck and R. E. White Jr.
What's New in Adult Reconstructive Knee Surgery
J. Bone Joint Surg. Am.,
July 1, 2005;
87(7):
1656 - 1666.
[Full Text]
[PDF]
|
 |
|
|