This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow CME: Take the exams for this article:
Adult Hip Reconstruction Test 10: Topics of Interest in Hip Arthroplasty
CME 4: October, November, December 2004
Right arrow [Supplementary Material]
Right arrow Letters to the Editor: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when Letters to the Editor are posted
Right arrow Alert me if a correction is posted
Services
Right arrow E-mail this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrowReprints and Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Walde, T. A.
Right arrow Articles by Engh, C. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Walde, T. A.
Right arrow Articles by Engh, C. A., Sr.
Related Collections
Right arrow Adult Hip
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Technorati  
What's this?
The Journal of Bone and Joint Surgery (American) 86:2251-2256 (2004)
© 2004 The Journal of Bone and Joint Surgery, Inc.

Advancement and Separation of Apex Hole Eliminators with Cementless Duraloc 100 Cups

A Report of Nineteen Cases

Tim A. Walde, MD1, Alexandra M. Claus, MD, PhD2, Vivek Mohan, MD3, Robert H. Hopper, Jr., PhD3, C. Anderson Engh, Jr., MD3 and Charles A. Engh, Sr., MD3

1 Abteilung fuer Unfallchirurgie, Plastische und Wiederherstellungschirurgie der Georg-August-Universitaet, Goettingen, Germany
2 Fakultaet fuer Klinische Medizin der Universitaet Heidelberg, Klinikum Mannheim gGmbH, Orthopaedische Universitaetsklinik, Mannheim, Germany
3 Anderson Orthopaedic Research Institute, 2501 Parker's Lane, Suite 200, Alexandria, VA 22306. E-mail address: research{at}aori.org

Investigation performed at the Anderson Orthopaedic Research Institute, Alexandria, Virginia

General research funding for this study was provided by Inova Health System, Falls Church, Virginia. Two of the authors receive royalties and serve as consultants for DePuy, a Johnson and Johnson company. One author owns Johnson and Johnson stock.


Background: From 1992 to 1996, we performed 541 cementless total hip arthroplasties using a Duraloc 100 acetabular component with an apex hole eliminator that did not have a positive stop design. During routine follow-up radiographic evaluation, we noted several hips with advanced or separated hole eliminators. This study presents the clinical consequences of this occurrence and its relationship to polyethylene wear and osteolysis.

Methods: A total of 541 hips (358 with a minimum duration of follow-up of five years) were evaluated clinically with use of the method of Merle D'Aubigné and Postel. Anteroposterior pelvic and iliac oblique radiographs were used to evaluate polyethylene wear, osteolysis, hole eliminator position, and component stability.

Results: Nineteen hips that had been followed for an average of eighty-nine months showed advancement or complete separation of the hole eliminator on radiographic evaluation. In most hips, advancement of the hole eliminator was visible on only the iliac oblique radiograph. In all but one hip with separation of the hole eliminator, complete plug separation was detectable on both the iliac oblique and anteroposterior pelvic radiographs. No association was found between polyethylene wear and either advancement or separation of the hole eliminator. No osteolysis was visible radiographically in the hips with hole eliminator advancement, but osteolysis was always visible in the hips that had complete separation of the hole eliminator.

Conclusions: The hole eliminator has been redesigned since November 1995 to prevent possible advancement. However, patients with earlier designs should be followed carefully because advancement or separation can be expected in about 5% of the patients who have been followed for a minimum of five years. Despite this complication, which is seen radiographically, all patients were satisfied with the function of the hip. We do not recommend surgery solely to remove or replace the hole eliminator.

Level of Evidence: Therapeutic study, Level IV (case series [no, or historical, control group]). See Instructions to Authors for a complete description of levels of evidence.