This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow CME: Take the exams for this article:
Adult Hip Reconstruction Test 6: Total Hip Arthroplasty Techniques
CME 1: January, February, March 2004
Right arrow Letters to the Editor: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when Letters to the Editor are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrowReprints and Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Ebramzadeh, E.
Right arrow Articles by Dorr, L. D.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Ebramzadeh, E.
Right arrow Articles by Dorr, L. D.
Related Collections
Right arrow Adult Hip
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Technorati  
What's this?
The Journal of Bone and Joint Surgery (American) 86:106-115 (2004)
© 2004 The Journal of Bone and Joint Surgery, Inc.

Initial Stability of Cemented Femoral Stems as a Function of Surface Finish, Collar, and Stem Size

Edward Ebramzadeh, PhD1, Sophia N. Sangiorgio, MS1, Donald B. Longjohn, MD2, Cyrus F. Buhari, MS, DO2 and Lawrence D. Dorr, MD3

1 Biomechanics Laboratory, Los Angeles Orthopaedic Hospital/University of California at Los Angeles, 2400 South Flower Street, Los Angeles, CA 90007-2697. E-mail address for E. Ebramzadeh: ebramzad{at}usc.edu
2 Departments of Orthopaedics (D.B.L.) and Biomedical Engineering (C.F.B.), University of Southern California, University Park Campus, Los Angeles, CA 90089
3 Dorr Arthritis Institute of Centinela Hospital, 501 East Hardy Street, Suite 300, Inglewood, CA 90301

Investigation performed at the Biomechanics Laboratory of the J. Vernon Luck, Sr., Orthopaedic Research Center, Los Angeles Orthopaedic Hospital/University of California at Los Angeles, Los Angeles, and the Dorr Arthritis Institute of Centinela Hospital, Inglewood, California

In support of their research or preparation of this manuscript, one or more of the authors received grants or outside funding from Centerpulse Orthopedics, Austin, Texas. They did not receive payments or other benefits or a commitment or agreement to provide such benefits from a commercial entity. No commercial entity paid or directed, or agreed to pay or direct, any benefits to any research fund, foundation, educational institution, or other charitable or nonprofit organization with which the authors are affiliated or associated.


Background: The optimum surface roughness of cemented femoral stems used for total hip replacement is a subject of controversy. While rougher surfaces provide stronger cement adhesion, it has been hypothesized that polished, tapered, noncollared stems settle into the cement mantle, providing improved stability. However, the effects of surface finish on the stability of straight, cemented stems tapered only in the coronal plane are not known.

Methods: Using composite model femora, we assessed the initial stability of a straight, cemented femoral stem as a function of surface roughness, the presence or absence of a collar, stem size, and the resultant cement thickness under simulated walking and stair-climbing loads. Otherwise identical stems were manufactured with polished or rough surfaces, with or without a collar, in two different sizes. We isolated these three variables and compared their relative contributions to the motion at the stem-cement interface throughout cyclic loading. We defined three indicators of stability: per-cycle motion, rate of migration, and final migration.

Results: Surface roughness had a greater influence on per-cycle motions than did the presence or absence of a collar or cement thickness. Specifically, in the medial-lateral direction, per-cycle motion of polished stems was 43 µm greater than that of rough stems (p < 0.01). None of the per-cycle motions decreased over the 77,000 load cycles. In contrast, with all stems, the rate of migration decreased over the course of cyclic loading, but the rate of migration of the polished stems was greater than that of the rough stems. Final migrations of the stems over the course of loading were generally distal, medial, and into retroversion. Compared with rough stems, polished stems had 8 to 18 µm more axial migration (p < 0.001), 48 µm more anterior-posterior migration (p < 0.001), and 0.4° more rotational migration (p = 0.01).

Conclusions and Clinical Relevance: The results indicated that, for cemented, straight femoral stems tapered only in the coronal plane, a rough surface offers the advantage of less per-cycle motion. These results may apply to widely used cemented stem designs based on the profile of the original Charnley femoral component, which has approximately parallel anterior and posterior aspects.