The Journal of Bone and Joint Surgery (American) 86:106-115 (2004)
© 2004 The Journal of Bone and Joint Surgery, Inc.
Initial Stability of Cemented Femoral Stems as a Function of Surface Finish, Collar, and Stem Size
Edward Ebramzadeh, PhD1,
Sophia N. Sangiorgio, MS1,
Donald B. Longjohn, MD2,
Cyrus F. Buhari, MS, DO2 and
Lawrence D. Dorr, MD3
1 Biomechanics Laboratory, Los Angeles Orthopaedic Hospital/University of
California at Los Angeles, 2400 South Flower Street, Los Angeles, CA
90007-2697. E-mail address for E. Ebramzadeh:
ebramzad{at}usc.edu
2 Departments of Orthopaedics (D.B.L.) and Biomedical Engineering (C.F.B.),
University of Southern California, University Park Campus, Los Angeles, CA
90089
3 Dorr Arthritis Institute of Centinela Hospital, 501 East Hardy Street, Suite
300, Inglewood, CA 90301
Investigation performed at the Biomechanics Laboratory of the J. Vernon
Luck, Sr., Orthopaedic Research Center, Los Angeles Orthopaedic
Hospital/University of California at Los Angeles, Los Angeles, and the Dorr
Arthritis Institute of Centinela Hospital, Inglewood, California
In support of their research or preparation of this manuscript, one or more
of the authors received grants or outside funding from Centerpulse
Orthopedics, Austin, Texas. They did not receive payments or other benefits or
a commitment or agreement to provide such benefits from a commercial entity.
No commercial entity paid or directed, or agreed to pay or direct, any
benefits to any research fund, foundation, educational institution, or other
charitable or nonprofit organization with which the authors are affiliated or
associated.
Background: The optimum surface roughness of cemented femoral stems
used for total hip replacement is a subject of controversy. While rougher
surfaces provide stronger cement adhesion, it has been hypothesized that
polished, tapered, noncollared stems settle into the cement mantle, providing
improved stability. However, the effects of surface finish on the stability of
straight, cemented stems tapered only in the coronal plane are not known.
Methods: Using composite model femora, we assessed the initial
stability of a straight, cemented femoral stem as a function of surface
roughness, the presence or absence of a collar, stem size, and the resultant
cement thickness under simulated walking and stair-climbing loads. Otherwise
identical stems were manufactured with polished or rough surfaces, with or
without a collar, in two different sizes. We isolated these three variables
and compared their relative contributions to the motion at the stem-cement
interface throughout cyclic loading. We defined three indicators of stability:
per-cycle motion, rate of migration, and final migration.
Results: Surface roughness had a greater influence on per-cycle
motions than did the presence or absence of a collar or cement thickness.
Specifically, in the medial-lateral direction, per-cycle motion of polished
stems was 43 µm greater than that of rough stems (p < 0.01). None of the
per-cycle motions decreased over the 77,000 load cycles. In contrast, with all
stems, the rate of migration decreased over the course of cyclic loading, but
the rate of migration of the polished stems was greater than that of the rough
stems. Final migrations of the stems over the course of loading were generally
distal, medial, and into retroversion. Compared with rough stems, polished
stems had 8 to 18 µm more axial migration (p < 0.001), 48 µm more
anterior-posterior migration (p < 0.001), and 0.4° more rotational
migration (p = 0.01).
Conclusions and Clinical Relevance: The results indicated that, for
cemented, straight femoral stems tapered only in the coronal plane, a rough
surface offers the advantage of less per-cycle motion. These results may apply
to widely used cemented stem designs based on the profile of the original
Charnley femoral component, which has approximately parallel anterior and
posterior aspects.