This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow [Supplementary Material]
Right arrow Letters to the Editor: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when Letters to the Editor are posted
Right arrow Alert me if a correction is posted
Services
Right arrow E-mail this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrowReprints and Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Martell, J. M.
Right arrow Articles by Jacobs, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Martell, J. M.
Right arrow Articles by Jacobs, J.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Technorati  
What's this?
The Journal of Bone and Joint Surgery (American) 85:1111-1117 (2003)
© 2003 The Journal of Bone and Joint Surgery, Inc.


Scientific Article

Comparison of Two and Three-Dimensional Computerized Polyethylene Wear Analysis After Total Hip Arthroplasty

John M. Martell, MD, Eric Berkson, MD, MA, Richard Berger, MD and Joshua Jacobs, MD

Investigation performed at The University of Chicago Hospitals and Clinics, Chicago, Illinois

John M. Martell, MD
Section of Orthopaedic Surgery and Rehabilitative Medicine, The University of Chicago, 5841 South Maryland Avenue, MC 3079, Chicago, IL 60637. E-mail address: jmartell{at}surgery.bsd.uchicago.edu

Eric Berkson, MD, MA
Richard Berger, MD
Joshua Jacobs, MD
Rush Arthritis and Orthopaedics Institute, 1653 W. Congress Parkway, Chicago, IL 60612

In support of their research or preparation of this manuscript, one or more of the authors received grants or outside funding from Stryker/Howmedica/Osteonics, the Department of Surgery of The University of Chicago; Zimmer; and National Institutes of Health/National Institute of Arthritis and Musculoskeletal and Skin Diseases Grant AR 39310. In addition, one or more of the authors received payments or other benefits or a commitment or agreement to provide such benefits from a commercial entity (Arch Development Corporation; Zimmer; The University of Chicago). Also, a commercial entity (Zimmer) paid or directed, or agreed to pay or direct, benefits to a research fund, foundation, educational institution, or other charitable or nonprofit organization with which the authors are affiliated or associated.

Background: The accurate determination of acetabular polyethylene wear in vivo is necessary to assess the clinical performance of the bearing surfaces of total hip replacements. Our objective in this study was to determine the clinical performance of two and three-dimensional computerized wear analysis and to assess the implications of this performance on requirements for patient enrollment in studies designed to detect wear of total hip prostheses.

Methods: Two and three-dimensional digital computerized analyses of acetabular polyethylene wear were performed on 153 hips in 140 patients. The acetabular components consisted of a polyethylene insert in a titanium shell, articulating with a 28-mm cobalt-chromium femoral component. The average duration of radiographic follow-up was 8.4 years. The correlation coefficient for two-dimensional versus three-dimensional analysis was calculated, as was the difference between the wear detected by the two techniques. The same observer analyzed each image twice, allowing an assessment of the repeatability of the two-dimensional and three-dimensional analyses. The impact of the clinical performance of each technique on the sample size needed for adequate power in prospective studies was evaluated.

Results: There was a high correlation between two-dimensional and three-dimensional wear analysis (r 2 = 0.933). In thirty-one (5.2%) of 595 observations, the wear values derived with the two-dimensional and three-dimensional techniques were not consistent. Logistic regression demonstrated that acetabular anteversion had a significant effect on the likelihood of such inconsistency occurring. The two-dimensional technique detected 90.1% of the total linear wear subsequently detected by the three-dimensional analysis. The average wear value was 1.09 mm as detected by two-dimensional analysis and 1.21 mm as detected by three-dimensional analysis. The two-dimensional technique was four times more repeatable than the three-dimensional technique. Power analysis indicated that up to 1.4 times more patients need to be enrolled if the three-dimensional technique is used for wear analysis.

Conclusions: While three-dimensional analysis detected 10% more wear, its repeatability was four times worse than that of the two-dimensional technique and, as a consequence, patient enrollment requirements for wear detection were higher. The poor quality of the lateral radiographs contributed to the decrease in the repeatability of the three-dimensional analysis. Three-dimensional analysis may be useful for highly anteverted cups, but the limited improvement in wear detection achieved with that technique, coupled with its inferior repeatability, limits its clinical value.

Level of Evidence: Diagnostic study, Level IV-2 (poor reference standard). See Instructions to Authors for a complete description of levels of evidence.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Technorati Technorati    What's this?


This article has been cited by other articles:


Home page
JBJSHome page
R. G. Bitsch, T. Loidolt, C. Heisel, S. Ball, and T. P. Schmalzried
Reduction of Osteolysis with Use of Marathon Cross-Linked Polyethylene. A Concise Follow-up, at a Minimum of Five Years, of a Previous Report
J. Bone Joint Surg. Am., July 1, 2008; 90(7): 1487 - 1491.
[Abstract] [Full Text] [PDF]


Home page
J Am Acad Orthop SurgHome page
A. V. Lombardi Jr and K. R. Berend
Isolated Acetabular Liner Exchange
J. Am. Acad. Ortho. Surg., May 1, 2008; 16(5): 243 - 246.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
C. A. Jacobs, C. P. Christensen, A. S. Greenwald, and H. McKellop
Clinical Performance of Highly Cross-Linked Polyethylenes in Total Hip Arthroplasty
J. Bone Joint Surg. Am., December 1, 2007; 89(12): 2779 - 2786.
[Abstract] [Full Text] [PDF]


Home page
JBJSHome page
R. W. McCalden, D. D. Naudie, X. Yuan, and R. B. Bourne
Radiographic Methods for the Assessment of Polyethylene Wear After Total Hip Arthroplasty
J. Bone Joint Surg. Am., October 1, 2005; 87(10): 2323 - 2334.
[Abstract] [Full Text] [PDF]