The Journal of Bone and Joint Surgery (American) 84:1836-1841 (2002)
© 2002 The Journal of Bone and Joint Surgery, Inc.
In Vivo Comparison of Hip Separation After Metal-on-Metal or Metal-on-Polyethylene Total Hip Arthroplasty
Richard D. Komistek, PhD,
Douglas A. Dennis, MD,
Jorge A. Ochoa, PhD,
Brian D. Haas, MD and
Curt Hammill, BS
Investigation performed at Rocky Mountain Musculoskeletal Research Laboratory, Denver, Colorado
Richard D. Komistek, PhD
Douglas A. Dennis, MD
Brian D. Haas, MD
Curt Hammill, BS
Rocky Mountain Musculoskeletal Research Laboratory, 2425 South Colorado Boulevard, Suite 280, Denver, CO 80222
Jorge A. Ochoa, PhD
DePuy Orthopaedics, 700 Orthopaedic Drive, Warsaw, IN 46581
In support of their research or preparation of this manuscript, one or more of the authors received grants or outside funding from DePuy, a Johnson and Johnson Company, Warsaw, Indiana, and Radiographic and Data Solutions, Minneapolis, Minnesota. None of the authors received payments or other benefits or a commitment or agreement to provide such benefits from a commercial entity. No commercial entity paid or directed, or agreed to pay or direct, any benefits to any research fund, foundation, educational institution, or other charitable or nonprofit organization with which the authors are affiliated or associated.
A commentary is available with the electronic versions of this article, on our web site (www.jbjs.org) and on our quarterly CD-ROM (call our subscription department, at 781-449-9780, to order the CD-ROM).
Background: Twenty subjects were analyzed in vivo with use of video fluoroscopy to determine if the femoral head separates from the acetabular component during normal gait and to determine if the amount of separation differs between metal-on-metal and metal-on-polyethylene total hip prostheses.
Methods: Ten subjects had been treated with a metal-on-metal total hip arthroplasty and ten, with a metal-on-polyethylene total hip arthroplasty. All of the prostheses were implanted by the same surgeon utilizing the same surgical technique, and all were judged to be clinically successful (a Harris hip score of >90 points). Each subject walked with a normal gait on a level treadmill while under fluoroscopic surveillance. The two-dimensional fluoroscopic videotapes were then converted into three-dimensional images with use of a computer-automated model-fitting technique. Each implant was analyzed at various flexion angles to assess the amount of femoral head sliding.
Results: No femoral head sliding was observed in the subjects with a metal-on-metal implant, whereas all ten subjects with a metal-on-polyethylene implant had sliding that was greater than our threshold value of 0.75 mm. The average amount of femoral head sliding in these subjects was 2.0 mm, and the sliding was observed during the swing phase of gait. The sliding was typically seen medially while the femoral head remained in contact with the acetabular component superolaterally.
Conclusions: Femoral head sliding commonly occurred following traditional metal-on-polyethylene total hip arthroplasty but not after metal-on-metal arthroplasty. These kinematic data may be of value in future hip-simulation studies to better duplicate wear patterns observed in retrieval analyses, assist in the understanding of the lubrication and wear rates of metal-on-metal designs, and facilitate designing of prosthetic components that minimize wear and optimize hip kinematics.

CiteULike Connotea Del.icio.us Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
H. A. McKellop and D. D'Lima
How have wear testing and joint simulator studies helped to discriminate among materials and designs?
J. Am. Acad. Ortho. Surg.,
July 1, 2008;
16(suppl_1):
S111 - S119.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Mahfouz, G. Nicholson, R. Komistek, D. Hovis, and M. Kubo
In Vivo Determination of the Dynamics of Normal, Rotator Cuff-Deficient, Total, and Reverse Replacement Shoulders
J. Bone Joint Surg. Am.,
December 1, 2005;
87(suppl_2):
107 - 113.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. C. Hermida, A. Bergula, P. Chen, C. W. Colwell Jr., and D. D. D'Lima
Comparison of the Wear Rates of Twenty-eight and Thirty-two-Millimeter Femoral Heads on Cross-Linked Polyethylene Acetabular Cups in a Wear Simulator
J. Bone Joint Surg. Am.,
December 1, 2003;
85(12):
2325 - 2331.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|