The Journal of Bone and Joint Surgery (American) 82:315-21 (2000)
© 2000 The Journal of Bone and Joint Surgery, Inc.
The Effect of the Orientation of the Acetabular and Femoral Components on the Range of Motion of the Hip at Different Head-Neck Ratios*
DARRYL D. D'LIMA, M.D. ,
ANDREW G. URQUHART, M.D. ,
KNUTE O. BUEHLER, M.D. ,
RICHARD H. WALKER, M.D. and
CLIFFORD W. COLWELL, JR. , M.D., LA JOLLA, CALIFORNIA
Investigation performed at the Division of Orthopaedic Surgery, Scripps Clinic, La Jolla
Background: Prosthetic impingement due to poor positioning can limit the range of motion of the hip after total hip arthroplasty. In this study, a computer model was used to determine the effects of the positions of the acetabular and femoral components and of varying head-neck ratios on impingement and range of motion.
Methods: A three-dimensional generic hip prosthesis with a hemispherical cup, a neck diameter of 12.25 millimeters, and a head size ranging from twenty-two to thirty-two millimeters was simulated on a computer. The maximum range of motion of the hip was measured, before the neck impinged on the liner of the cup, for acetabular abduction angles ranging from 35 to 55 degrees and acetabular and femoral anteversion ranging from 0 to 30 degrees. Stability of the hip was estimated as the maximum possible flexion coupled with 10 degrees of adduction and 10 degrees of internal rotation and also as the maximum possible extension coupled with 10 degrees of external rotation. The effects of prosthetic orientation on activities of daily living were analyzed as well.
Results: Acetabular abduction angles of less than 45 degrees decreased flexion and abduction of the hip, whereas higher angles decreased adduction and rotation. Femoral and acetabular anteversion increased flexion but decreased extension. Acetabular abduction angles of between 45 and 55 degrees permitted a better overall range of motion and stability when combined with appropriate acetabular and femoral anteversion. Lower head-neck ratios decreased the range of motion that was possible without prosthetic impingement. The addition of a modular sleeve that increased the diameter of the femoral neck by two millimeters decreased the range of motion by 1.5 to 8.5 degrees, depending on the direction of motion that was studied.
Conclusions: There is a complex interplay between the angles of orientation of the femoral and acetabular components. Acetabular abduction angles between 45 and 55 degrees, when combined with appropriate acetabular and femoral anteversion, resulted in a maximum overall range of motion and stability with respect to prosthetic impingement.
Clinical Relevance: During total hip arthroplasty, acetabular abduction is often constrained by available bone coverage, while femoral anteversion may be dictated by the geometry of the femoral shaft. For each combination of acetabular abduction and femoral anteversion, there is an optimum range of acetabular anteversion that allows the potential for a maximum range of motion without prosthetic impingement after total hip arthroplasty. These data can be used intraoperatively to determine optimum position.

CiteULike Connotea Del.icio.us Facebook Technorati Twitter What's this?
This article has been cited by other articles:

|
 |

|
 |
 
S. A. Stuchin
Anatomic Diameter Femoral Heads in Total Hip Arthroplasty: A Preliminary Report
J. Bone Joint Surg. Am.,
August 1, 2008;
90(Supplement_3):
52 - 56.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Shimmin, P. E. Beaule, and P. Campbell
Metal-on-Metal Hip Resurfacing Arthroplasty
J. Bone Joint Surg. Am.,
March 1, 2008;
90(3):
637 - 654.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. W. Babisch, F. Layher, and L.-P. Amiot
The Rationale for Tilt-Adjusted Acetabular Cup Navigation
J. Bone Joint Surg. Am.,
February 1, 2008;
90(2):
357 - 365.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Malik, A. Maheshwari, and L. D. Dorr
Impingement with Total Hip Replacement
J. Bone Joint Surg. Am.,
August 1, 2007;
89(8):
1832 - 1842.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Parratte and J.-N. A. Argenson
Validation and Usefulness of a Computer-Assisted Cup-Positioning System in Total Hip Arthroplasty. A Prospective, Randomized, Controlled Study
J. Bone Joint Surg. Am.,
March 1, 2007;
89(3):
494 - 499.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Kalteis, M. Handel, H. Bathis, L. Perlick, M. Tingart, and J. Grifka
Imageless navigation for insertion of the acetabular component in total hip arthroplasty: IS IT AS ACCURATE AS CT-BASED NAVIGATION?
J Bone Joint Surg Br,
February 1, 2006;
88-B(2):
163 - 167.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. C. Hermida, A. Bergula, P. Chen, C. W. Colwell Jr., and D. D. D'Lima
Comparison of the Wear Rates of Twenty-eight and Thirty-two-Millimeter Femoral Heads on Cross-Linked Polyethylene Acetabular Cups in a Wear Simulator
J. Bone Joint Surg. Am.,
December 1, 2003;
85(12):
2325 - 2331.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Patil, A. Bergula, P. C. Chen, C. W. Colwell Jr., and D. D. D'Lima
Polyethylene Wear and Acetabular Component Orientation
J. Bone Joint Surg. Am.,
November 1, 2003;
85(90004):
56 - 63.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. L. Barrack
Dislocation After Total Hip Arthroplasty: Implant Design and Orientation
J. Am. Acad. Ortho. Surg.,
March 1, 2003;
11(2):
89 - 99.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. J. Botte, K. A. Ezzet, L. L. Pacelli, M. J. Guzman, R. S. Meyer, M. J. Meunier, D. D. D'Lima, and C. W. Colwell
What's New in Orthopaedic Rehabilitation
J. Bone Joint Surg. Am.,
December 9, 2002;
84(12):
2312 - 2320.
[Full Text]
[PDF]
|
 |
|
|