This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Letters to the Editor: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when Letters to the Editor are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrowReprints and Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by RUBMAN, M. H.
Right arrow Articles by LENHART, S. W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by RUBMAN, M. H.
Right arrow Articles by LENHART, S. W.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Technorati  
What's this?
The Journal of Bone and Joint Surgery 80:1264-1269 (1998)
© 1998 The Journal of Bone and Joint Surgery, Inc.

Levels of Carbon Dioxide in Helmet Systems Used during Orthopaedic Operations*

MARC H. RUBMAN, M.D.{dagger}, MARK G. SIEGEL, M.D.{ddagger}, ALAN S. ECHT, M.P.H.§, G. EDWARD BURROUGHS, § and STEVEN W. LENHART, §, CINCINNATI, OHIO

Investigation performed at Cincinnati Sportsmedicine Research and Education Foundation and Deaconess Hospital, Cincinnati

The use of isolation helmets has gained popularity as a method of possible protection of the operating-room personnel from diseases that can be transmitted during operative procedures. However, the use of these systems has been associated with a variety of symptoms, including fatigue, diaphoresis, nausea, headache, and irritability. These symptoms have often been attributed to the mental stress of the operative procedure or the physical discomfort of the helmet. As far as we know, no manufacturers include the measured levels of carbon dioxide or the rate of air exchange of their helmet system. A possible common cause of discomfort with helmet systems is the level of carbon dioxide to which the person wearing the device is exposed. We measured the levels of carbon dioxide in four helmet systems from three different manufacturers during light exercise designed to approximate the exertion during an orthopaedic operation. All but one unit failed to meet the exposure limits recommended by the National Institute for Occupational Safety and Health and the Occupational Safety and Health Administration regarding exposure to carbon dioxide. One unit, the Stackhouse Freedom Aire self-contained system, did meet these standards, but the levels of carbon dioxide in this helmet were more than 1000 per cent greater than the ambient levels in air (440 parts per million compared with 4939 parts per million). Isolation systems must be evaluated carefully not only for comfort but also for the physiological effects caused by exposure to elevated levels of carbon dioxide. Operating-room personnel who use such systems should be aware that many of the physical symptoms that they experience may be associated with elevated levels of carbon dioxide.