The Journal of Bone and Joint Surgery 80:547-54 (1998)
© 1998 The Journal of Bone and Joint Surgery, Inc.
A Comparison of Hydroxyapatite-Coated, Titanium-Coated, and Uncoated Tapered External-Fixation Pins. An in Vivo Study in Sheep*
ANTONIO MORONI, M.D. ,
SOREN TOKSVIG-LARSEN, M.D. ,
MARIA CRISTINA MALTARELLO, B.SC. ,
LUCA ORIENTI, M.D. ,
SUSANNA STEA, B.SC. and
SANDRO GIANNINI, M.D. , BOLOGNA, ITALY
Investigation performed at the Department of Orthopaedic Surgery, Bologna University, Rizzoli Orthopaedic Institute, Bologna
Three types of surfaces for external fixation pins were compared. One hundred and eight stainless-steel tapered 5/6-millimeter pins were divided into three groups: thirty-six pins remained uncoated (Group A), thirty-six were plasma-sprayed with hydroxyapatite (Group B), and thirty-six were plasma-sprayed with titanium (Group C). The pins were implanted in the left tibia of eighteen sheep, with each sheep receiving six pins from the same group. A unilateral fixator then was assembled on the pins. The medial aspect of the mid-part of the tibial diaphysis was exposed, and a five-millimeter-long cylinder of bone was removed so that load would be borne by the bone-pin interfaces. Six weeks after the procedure, radiographs demonstrated rarefaction of twenty-nine pin tracks in Group A, fifteen in Group B, and thirty in Group C (p = 0.021 for Group A compared with Group B and p = 0.016 for Group B compared with Group C). The mean final insertion torque (and standard deviation) was 4360 ± 1050 newton-millimeters in Group A, 3420 ± 676 newton-millimeters in Group B, and 3740 ± 643 newton-millimeters in Group C. With the numbers available, no significant differences could be detected among these values. The mean extraction torque was 253 ± 175 newton-millimeters in Group A, 3360 ± 1260 newton-millimeters in Group B, and 1720 ± 1030 newton-millimeters in Group C (p = 0.002 for Group A compared with Group B, p = 0.017 for Group A compared with Group C, and p = 0.03 for Group B compared with Group C). The extraction torque was significantly lower than the corresponding insertion torque in both Group A (p < 0.001) and Group C (p = 0.003); no significant difference could be found, with the numbers available, in Group B (hydroxyapatite-coated pins). At sixty times magnification, direct contact was seen along a mean of 16 ± 9 per cent of the bone-pin interface in Group A, 30 ± 12 per cent of the interface in Group B, and 28 ± 15 per cent of the interface in Group C (p = 0.042 for Group A compared with Group C). However, at 10,000 times magnification, direct bone-pin contact was found only in Group B.
CLINICAL RELEVANCE: The better fixation and greater interface strength of the hydroxyapatite-coated tapered pins should lead to a substantial decrease in the frequency of loosening and infection and consequently to a remarkable improvement in the clinical results of external fixation.

CiteULike Connotea Del.icio.us Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
J. J. Keeling, D. E. Gwinn, S. M. Tintle, R. C. Andersen, and F. X. McGuigan
Short-Term Outcomes of Severe Open Wartime Tibial Fractures Treated with Ring External Fixation
J. Bone Joint Surg. Am.,
December 1, 2008;
90(12):
2643 - 2651.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Moroni, C. Faldini, F. Pegreffi, A. Hoang-Kim, F. Vannini, and S. Giannini
Dynamic Hip Screw Compared with External Fixation for Treatment of Osteoporotic Pertrochanteric Fractures. A Prospective, Randomized Study
J. Bone Joint Surg. Am.,
April 1, 2005;
87(4):
753 - 759.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Moroni, C. Faldini, S. Marchetti, M. Manca, V. Consoli, and S. Giannini
Improvement of the Bone-Pin Interface Strength in Osteoporotic Bone with Use of Hydroxyapatite-Coated Tapered External-Fixation Pins : A Prospective, Randomized Clinical Study of Wrist Fractures
J. Bone Joint Surg. Am.,
May 1, 2001;
83(5):
717 - 721.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|