The Journal of Bone and Joint Surgery 79:1211-22 (1997)
© 1997 The Journal of Bone and Joint Surgery, Inc.
Semiconstrained Total Elbow Replacement for the Treatment of Post-Traumatic Osteoarthrosis*
ALBERTO G. SCHNEEBERGER, M.D. ,
ROBERT ADAMS, R.P.A. and
BERNARD F. MORREY, M.D. , ROCHESTER, MINNESOTA
Investigation performed at the Mayo Clinic, Rochester
Forty-one consecutive patients were managed for post-traumatic osteoarthrosis or dysfunction of the elbow with use of a non-customized semiconstrained Coonrad-Morrey total elbow prosthesis. The average age at the time of the operation was fifty-seven years (range, thirty-two to eighty-two years). The patients were followed for an average of five years and eight months (range, two to twelve years). Radiographs were made at least two years postoperatively (average, five years and one month; range, two to twelve years) for thirty-nine of the forty-one patients. According to the Mayo elbow performance score, sixteen patients (39 per cent) had an excellent result, eighteen (44 per cent) had a good result, five (12 per cent) had a fair result, and two (5 per cent) had a poor result. Thirty-six (95 per cent) of the thirty-eight patients who had a functioning implant at the time of follow-up considered the outcome to be satisfactory. Preoperatively, thirty-seven patients (90 per cent) had moderate or severe pain; postoperatively, thirty (73 per cent) had no or only mild discomfort. Motion improved from an average arc of flexion of 40 to 118 degrees preoperatively to an average arc of flexion of 27 to 131 degrees postoperatively. All thirty-eight functioning implants rendered the elbow stable.
Eleven patients (27 per cent) had a major complication. Nine of them (22 per cent of the series) needed an additional operation. There was no aseptic loosening, and most of the complications were primarily due to so-called mechanical failure. The ulnar component fractured in five patients (12 per cent), and the polyethylene bushings wore out in two (5 per cent). These complications were attributed principally to the performance of strenuous physical labor, such as lifting more than ten kilograms on a regular basis, against the advice of the surgeon; excessive preoperative deformity of the joint; or an unstable traumatic injury. Two patients (5 per cent) had an infection.
Semiconstrained joint replacement of the elbow can be a reliable form of treatment, and frequently is the only viable option, for the difficult problems encountered with post-traumatic destruction of a joint. Restoration of function, relief of pain, and patient satisfaction can be achieved even when a patient is less than sixty years old if that patient has low demands and a low level of activity. However, the mechanical failures underscore the fact that this procedure is relatively contraindicated in patients who anticipate strenuous physical activity or who are not expected to comply with the postoperative protocol. This observation reflects the tendency for increased and excessive use of a previously functionless joint, after it has been rendered stable and pain-free, to lead to mechanical failure.

CiteULike Connotea Del.icio.us Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
S. H. Goldberg, R. M. Urban, J. J. Jacobs, G. J.W. King, S. W. O'Driscoll, and M. S. Cohen
Modes of Wear After Semiconstrained Total Elbow Arthroplasty
J. Bone Joint Surg. Am.,
March 1, 2008;
90(3):
609 - 619.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. L. Shi, D. Zurakowski, D. G. Jones, M. J. Koris, and T. S. Thornhill
Semiconstrained Primary and Revision Total Elbow Arthroplasty with Use of the Coonrad-Morrey Prosthesis
J. Bone Joint Surg. Am.,
July 1, 2007;
89(7):
1467 - 1475.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. S. Athwal and B. F. Morrey
Revision Total Elbow Arthroplasty for Prosthetic Fractures
J. Bone Joint Surg. Am.,
September 1, 2006;
88(9):
2017 - 2026.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. B. Sneftrup, S. L. Jensen, H. V. Johannsen, and J. O. Sojbjerg
Revision of failed total elbow arthroplasty with use of a linked implant
J Bone Joint Surg Br,
January 1, 2006;
88-B(1):
78 - 83.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. P. Little, A. J. Graham, G. Karatzas, D. A. Woods, and A. J. Carr
Outcomes of Total Elbow Arthroplasty for Rheumatoid Arthritis: Comparative Study of Three Implants
J. Bone Joint Surg. Am.,
November 1, 2005;
87(11):
2439 - 2448.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. P. Lee, R. A. Adams, and B. F. Morrey
Polyethylene Wear After Total Elbow Arthroplasty
J. Bone Joint Surg. Am.,
May 1, 2005;
87(5):
1080 - 1087.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. P. Little, A. J. Graham, and A. J. Carr
Total elbow arthroplasty: A SYSTEMATIC REVIEW OF THE LITERATURE IN THE ENGLISH LANGUAGE UNTIL THE END OF 2003
J Bone Joint Surg Br,
April 1, 2005;
87-B(4):
437 - 444.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. A. Blaine, R. Adams, and B. F. Morrey
Total Elbow Arthroplasty After Interposition Arthroplasty for Elbow Arthritis
J. Bone Joint Surg. Am.,
February 1, 2005;
87(2):
286 - 292.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. A. Mighell, R. C. Dunham, E. A. Rommel, and M. A. Frankle
Primary semi-constrained arthroplasty for chronic fracture-dislocations of the elbow
J Bone Joint Surg Br,
February 1, 2005;
87-B(2):
191 - 195.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. I. Loebenberg, R. Adams, S. W. O'Driscoll, and B. F. Morrey
Impaction Grafting in Revision Total Elbow Arthroplasty
J. Bone Joint Surg. Am.,
January 1, 2005;
87(1):
99 - 106.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Kamineni and B.F. Morrey
Proximal Ulnar Reconstruction with Strut Allograft in Revision Total Elbow Arthroplasty
J. Bone Joint Surg. Am.,
June 1, 2004;
86(6):
1223 - 1229.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Kamineni and B. F. Morrey
Distal Humeral Fractures Treated with Noncustom Total Elbow Replacement
J. Bone Joint Surg. Am.,
May 1, 2004;
86(5):
940 - 947.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. Mansat, R. A. Adams, and B. F. Morrey
Allograft-Prosthesis Composite for Revision of Catastrophic Failure of Total Elbow Arthroplasty
J. Bone Joint Surg. Am.,
April 1, 2004;
86(4):
724 - 735.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. D. McKee, D. M.W. Pugh, R. R. Richards, E. Pedersen, C. Jones, and E. H. Schemitsch
Effect of Humeral Condylar Resection on Strength and Functional Outcome After Semiconstrained Total Elbow Arthroplasty
J. Bone Joint Surg. Am.,
May 1, 2003;
85(5):
802 - 807.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. Mansat and B. F. Morrey
Semiconstrained Total Elbow Arthroplasty for Ankylosed and Stiff Elbows
J. Bone Joint Surg. Am.,
September 1, 2000;
82(9):
1260 - 1260.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
D. R. J. GILL and B. F. MORREY
The Coonrad-Morrey Total Elbow Arthroplasty in Patients Who Have Rheumatoid Arthritis. A Ten to Fifteen-Year Follow-up Study
J. Bone Joint Surg. Am.,
September 1, 1998;
80(9):
1327 - 1335.
[Abstract]
[Full Text]
|
 |
|
|