The Journal of Bone and Joint Surgery 79:1190-7 (1997)
© 1997 The Journal of Bone and Joint Surgery, Inc.
A Comparison of Fixation Screws for the Scaphoid during Application of Cyclical Bending Loads*
E. B. TOBY, M.D. ,
T. E. BUTLER, M.D. ,
T. J. McCORMACK, M.D. and
G. JAYARAMAN, PH.D. , KANSAS CITY, KANSAS
Investigation performed at The University of Kansas Medical Center, Kansas City
Matched pairs of scaphoids from cadavera were stressed with ramped intensity cyclical bending loads after osteotomy and fixation of one scaphoid with a Herbert screw and fixation of the other with an AO 3.5-millimeter cannulated screw, a Herbert-Whipple screw, an Acutrak cannulated screw, or a Universal Compression screw. The AO screw, Acutrak screw, and Herbert-Whipple screw demonstrated superior resistance to cyclical bending loads compared with the Herbert screw. The Universal Compression screw did not provide better fixation than the Herbert screw because of fractures that occurred at the time of insertion. The AO screw and the Herbert screw were then tested in a separate setup in which a segment of volar cortex had been removed in addition to the simple osteotomy. The loss of volar cortex greatly diminished the quality of the fixation provided by both of the screws during application of ramped intensity cyclical bending loads.
CLINICAL RELEVANCE: A fixation device in the scaphoid must be able to withstand the stresses that are placed on the scaphoid as a result of its position spanning the proximal and distal carpal rows. Also, because of the prolonged time required for healing of fractures or non-unions of the scaphoid, the device must be able to withstand many such cycles of stress. The present study demonstrates that commonly used screws for fixation of the scaphoid vary significantly (p < 0.005) in their ability to resist cyclical bending loads.

CiteULike Connotea Del.icio.us Facebook Technorati Twitter What's this?
This article has been cited by other articles:

|
 |

|
 |
 
A. P. Gutow
Percutaneous Fixation of Scaphoid Fractures
J. Am. Acad. Ortho. Surg.,
August 1, 2007;
15(8):
474 - 485.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Waitayawinyu, W. V. McCallister, N. M. Nemechek, and T. E. Trumble
Scaphoid Nonunion
J. Am. Acad. Ortho. Surg.,
May 1, 2007;
15(5):
308 - 320.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. M. Haisman, R. S. Rohde, and A. J. Weiland
Acute Fractures of the Scaphoid
J. Bone Joint Surg. Am.,
December 1, 2006;
88(12):
2750 - 2758.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. E. Trumble, P. Salas, T. Barthel, and K. Q. Robert III
Management of Scaphoid Nonunions
J. Am. Acad. Ortho. Surg.,
November 1, 2003;
11(6):
380 - 391.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. F. Slade III, W. B. Geissler, A. P. Gutow, and G. A. Merrell
Percutaneous Internal Fixation of Selected Scaphoid Nonunions with an Arthroscopically Assisted Dorsal Approach
J. Bone Joint Surg. Am.,
November 1, 2003;
85(90004):
20 - 32.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
W. V. McCallister, J. Knight, R. Kaliappan, and T. E. Trumble
Central Placement of the Screw in Simulated Fractures of the Scaphoid Waist: A Biomechanical Study
J. Bone Joint Surg. Am.,
January 17, 2003;
85(1):
72 - 77.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. F. Slade III, A. P. Gutow, and W. B. Geissler
Percutaneous Internal Fixation of Scaphoid Fracturesvia an Arthroscopically Assisted Dorsal Approach
J. Bone Joint Surg. Am.,
November 25, 2002;
84
(suppl_2):
S21 - S36.
[Abstract]
[Full Text]
[PDF]
|
 |
|
|