The Journal of Bone and Joint Surgery 79:682-89 (1997)
© 1997 The Journal of Bone and Joint Surgery, Inc.
Measurement of Polyethylene Wear in Acetabular Components Inserted with and without Cement. A Randomized Trial*
PETER A. DEVANE, M.B., CH.B. ,
ERIC J. ROBINSON, M.D. ,
ROBERT B. BOURNE, M.D. ,
CECIL H. RORABECK, M.D. ,
NARESH N. NAYAK, M.D. and
J. GEOFFREY HORNE, M.B., CH.B. , LONDON, ONTARIO, CANADA
Investigation performed at the Department of Orthopaedic Surgery, University Hospital, London
We measured the three-dimensional wear of polyethylene after total hip arthroplasty with a titanium metal-backed Mallory-Head prosthesis that was inserted with cement in sixty-nine patients (sixty-nine hips) and with a press-fit titanium metal-backed Mallory-Head prosthesis that was inserted without cement in seventy patients (seventy hips). A modular titanium femoral head was used in all of the hips. The patients in the present study were part of a larger double-blind randomized trial comparing the result of total hip arthroplasty performed with cement with that of the same procedure performed without cement in 250 patients. The criterion for inclusion in the study of polyethylene wear was a minimum duration of follow-up of four years, which was met by 148 patients. As adequate radiographs for digitization were not available for nine patients, 139 patients were included in the present study. The age of the patient, the postoperative Harris hip score, the diameter of the femoral head, the thickness of the liner in the polar region of the acetabular component, and the duration of follow-up were similar for the two groups. The mean rate of volumetric wear of the polyethylene was significantly greater in the prostheses that had been inserted without cement than in those that had been inserted with cement (155.1 cubic millimeters per year compared with 98.5 cubic millimeters per year; p = 0.000008). Thirty-four (49 per cent) of the seventy hips in which the prosthesis had been inserted without cement had evidence of osteolysis on radiographs, compared with twelve (17 per cent) of the sixty-nine hips in the other group (p = 0.0002). Osteolysis was associated with an increased rate of polyethylene wear only in the hips in which the prosthesis had been inserted without cement.

CiteULike Connotea Del.icio.us Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
A. Chougle, M.V. Hemmady, and J.P. Hodgkinson
Long-Term Survival of the Acetabular Component After Total Hip Arthroplasty with Cement in Patients with Developmental Dysplasia of the Hip
J. Bone Joint Surg. Am.,
January 1, 2006;
88(1):
71 - 79.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. W. McCalden, D. D. Naudie, X. Yuan, and R. B. Bourne
Radiographic Methods for the Assessment of Polyethylene Wear After Total Hip Arthroplasty
J. Bone Joint Surg. Am.,
October 1, 2005;
87(10):
2323 - 2334.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. A. Conditt, M. T. Thompson, M. M. Usrey, S. K. Ismaily, and P. C. Noble
Backside Wear of Polyethylene Tibial Inserts: Mechanism and Magnitude of Material Loss
J. Bone Joint Surg. Am.,
February 1, 2005;
87(2):
326 - 331.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. N. Charles, R. B. Bourne, J. R. Davey, A. S. Greenwald, B. F. Morrey, and C. H. Rorabeck
Soft-Tissue Balancing of the Hip. The Role of Femoral Offset Restoration
J. Bone Joint Surg. Am.,
May 1, 2004;
86(5):
1078 - 1088.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Patil, A. Bergula, P. C. Chen, C. W. Colwell Jr., and D. D. D'Lima
Polyethylene Wear and Acetabular Component Orientation
J. Bone Joint Surg. Am.,
November 1, 2003;
85(90004):
56 - 63.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. M. Martell, E. Berkson, R. Berger, and J. Jacobs
Comparison of Two and Three-Dimensional Computerized Polyethylene Wear Analysis After Total Hip Arthroplasty
J. Bone Joint Surg. Am.,
May 28, 2003;
85(6):
1111 - 1117.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. H. Hopper Jr., A. M. Young, K. F. Orishimo, and C. A. Engh Jr.
Effect of Terminal Sterilization with Gas Plasma or Gamma Radiation on Wear of Polyethylene Liners
J. Bone Joint Surg. Am.,
March 11, 2003;
85(3):
464 - 468.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. D. Crowther and P. F. Lachiewicz
Survival and Polyethylene Wear of Porous-Coated Acetabular Components in Patients Less than Fifty Years Old : Results at Nine to Fourteen Years
J. Bone Joint Surg. Am.,
May 1, 2002;
84(5):
729 - 735.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. McKellop, F.-w. Shen, B. Lu, P. Campbell, and R. Salovey
Effect of Sterilization Method and Other Modifications on the Wear Resistance of Acetabular Cups Made of Ultra-High Molecular Weight Polyethylene : A Hip-Simulator Study
J. Bone Joint Surg. Am.,
December 1, 2000;
82(12):
1708 - 1708.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
G. LEWIS
Commentary - Design Issues in Clinical Studies of the in Vivo Volumetric Wear Rate of Polyethylene Bearing Components
J. Bone Joint Surg. Am.,
February 1, 2000;
82(2):
281 - 87.
[Full Text]
|
 |
|

|
 |

|
 |
 
T. P. SCHMALZRIED and J. J. CALLAGHAN
Current Concepts Review - Wear in Total Hip and Knee Replacements
J. Bone Joint Surg. Am.,
January 1, 1999;
81(1):
115 - 136.
[Full Text]
|
 |
|

|
 |

|
 |
 
T. P. SCHMALZRIED, F. J. DOREY, and H. MCKELLOP
The Multifactorial Nature of Polyethylene Wear in Vivo
J. Bone Joint Surg. Am.,
August 1, 1998;
80(8):
1234 - 42.
[Full Text]
|
 |
|
|