The Journal of Bone and Joint Surgery 79:1529-38 (1997)
© 1997 The Journal of Bone and Joint Surgery, Inc.
Complications of Total Hip Arthroplasty Associated with the Use of an Acetabular Component with a Hylamer Liner*
B. J. LIVINGSTON, M.D. ,
M. J. CHMELL, M.D. ,
M. SPECTOR, M.D. and
R. POSS, M.D. , BOSTON, MASSACHUSETTS
Investigation performed at Brigham and Women's Hospital, Boston
We observed early failure and radiographic signs of accelerated wear as early as one to three years after insertion, without cement, of a metal-backed acetabular component with a liner made of Hylamer (DePuy). This finding prompted us to review a larger cohort of patients in whom that liner had been used. Two hundred and thirty-three components with a Hylamer liner, evaluated at a minimum of two years, had a mean rate of wear of 0.27 millimeter per year compared with 0.12 millimeter per year for a contemporaneous group of fifty acetabular cups with a conventional ultra-high molecular weight polyethylene liner made by another manufacturer.
At a mean of 3.2 years, we found a significant difference (p < 0.000000006) between the mean rate of wear (0.20 millimeter per year) when the Hylamer liner articulated with a DePuy modular cobalt-chromium femoral head and the mean rate (0.29 millimeter per year) when the liner articulated with an Osteonics modular cobalt-chromium femoral head. Radiographic evaluation revealed a significant correlation between the total linear wear and the prevalence of osteolytic lesions (r2 = 0.76, linear regression analysis). We found that wear of 1.5 millimeters or more could be detected by the unaided eye. Because of the positive correlation between osteolysis and wear of 1.5 millimeters or more, we defined a hip with a liner that had that amount of wear as a hip at risk.
We concluded that the wear characteristics of a Hylamer liner in vivo are inferior to those of a conventional ultra-high molecular weight polyethylene liner. Also, the rate of wear of the liner is greater when the femoral head is from a manufacturer other than DePuy. A patient who has a total hip replacement that includes a Hylamer liner should be monitored frequently for signs of wear and osteolytic changes. Additional investigations, with longer durations of follow-up and larger populations, are needed to understand fully the importance of our findings.

CiteULike Connotea Del.icio.us Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
A. C. Gordon, D. D. D'Lima, and C. W. Colwell Jr
Highly Cross-linked Polyethylene in Total Hip Arthroplasty.
J. Am. Acad. Ortho. Surg.,
September 1, 2006;
14(9):
511 - 523.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. von Schewelov, L. Sanzen, I. Onsten, A. Carlsson, and J. Besjakov
Total hip replacement with a zirconium oxide ceramic femoral head: A RANDOMISED ROENTGEN STEREOPHOTOGRAMMETRIC STUDY
J Bone Joint Surg Br,
December 1, 2005;
87-B(12):
1631 - 1635.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Visentin, S. Stea, S. Squarzoni, M. Reggiani, C. Fagnano, B. Antonietti, and A. Toni
Isolation and Characterization of Wear Debris Generated in Patients Wearing Polyethylene Hylamer Inserts, Gamma Irradiated in Air
J Biomater Appl,
October 1, 2005;
20(2):
103 - 121.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
C. J. Sychterz, K. F. Orishimo, and C. A. Engh
Sterilization and Polyethylene Wear: Clinical Studies to Support Laboratory Data
J. Bone Joint Surg. Am.,
May 1, 2004;
86(5):
1017 - 1022.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Heisel, M. Silva, M. A. dela Rosa, and T. P. Schmalzried
Short-Term in Vivo Wear of Cross-Linked Polyethylene
J. Bone Joint Surg. Am.,
April 1, 2004;
86(4):
748 - 751.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. F. Orishimo, A. M. Claus, C. J. Sychterz, and C. A. Engh
Relationship Between Polyethylene Wear and Osteolysis in Hips witha Second-Generation Porous-Coated Cementless Cup After Seven Years of Follow-up
J. Bone Joint Surg. Am.,
May 28, 2003;
85(6):
1095 - 1099.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. Illgen II and H. E. Rubash
The Optimal Fixation of the Cementless Acetabular Component in Primary Total Hip Arthroplasty
J. Am. Acad. Ortho. Surg.,
January 1, 2002;
10(1):
43 - 56.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. and
Data in Clinical Studies of Volumetric Wear
J. Bone Joint Surg. Am.,
April 1, 2001;
83(4):
622 - 622.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. McKellop, F.-w. Shen, B. Lu, P. Campbell, and R. Salovey
Effect of Sterilization Method and Other Modifications on the Wear Resistance of Acetabular Cups Made of Ultra-High Molecular Weight Polyethylene : A Hip-Simulator Study
J. Bone Joint Surg. Am.,
December 1, 2000;
82(12):
1708 - 1708.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
G. LEWIS
Commentary - Design Issues in Clinical Studies of the in Vivo Volumetric Wear Rate of Polyethylene Bearing Components
J. Bone Joint Surg. Am.,
February 1, 2000;
82(2):
281 - 87.
[Full Text]
|
 |
|

|
 |

|
 |
 
B. K. VAUGHN, T. B. DAMERON, T. W. BAUER, Y. MOCHIDA, T. AKISUE, and R. W. EBERLE
Early Osteolysis Following Total Hip Arthroplasty with Use of a Hylamer Liner in Combination with a Modular Ceramic Femoral Head. A Case Report
J. Bone Joint Surg. Am.,
October 1, 1999;
81(10):
1446 - 9.
[Full Text]
|
 |
|

|
 |

|
 |
 
T. P. SCHMALZRIED and J. J. CALLAGHAN
Current Concepts Review - Wear in Total Hip and Knee Replacements
J. Bone Joint Surg. Am.,
January 1, 1999;
81(1):
115 - 136.
[Full Text]
|
 |
|

|
 |

|
 |
 
T. P. SCHMALZRIED, F. J. DOREY, and H. MCKELLOP
The Multifactorial Nature of Polyethylene Wear in Vivo
J. Bone Joint Surg. Am.,
August 1, 1998;
80(8):
1234 - 42.
[Full Text]
|
 |
|

|
 |

|
 |
 
R. POSS and M. SPECTOR
Response to Commentary
J. Bone Joint Surg. Am.,
August 1, 1998;
80(8):
1242 - 43.
[Full Text]
|
 |
|
|