The Journal of Bone and Joint Surgery 78:975-81 (1996)
© 1996 The Journal of Bone and Joint Surgery, Inc.
Porous-Coated Acetabular Components with Screw Fixation. Five to Ten-Year Results*
HARRISON A. LATIMER, M.D. and
PAUL F. LACHIEWICZ, M.D. , CHAPEL HILL, NORTH CAROLINA
Investigation performed at the Department of Orthopaedics, University of North Carolina at Chapel Hill, Chapel Hill
The results of 136 consecutive primary total hip arthroplasties performed by one surgeon with the Harris-Galante-I porous-coated acetabular component were reviewed at a mean of seven years (range, five to ten years). In all hips, the outer diameter of the acetabular component was the same as the diameter of the final reamer used in the preparation of the acetabulum. However, this reamer was used only briefly at the rim of the acetabulum, and therefore the components had so-called press-fit stability. A mean of four screws (range, three to six screws) were used for additional fixation of the component. The clinical evaluation was performed with use of the Harris hip score. Standardized anteroposterior radiographs of the pelvis were assessed for migration of the component, radiolucent and radiodense lines, linear wear of the polyethylene, and osteolysis.
No acetabular component had been revised for loosening and none were radiographically loose at the time of the most recent follow-up evaluation. There were no complications related to the use of the screws, and no screw had bent or broken. A non-progressive radiolucent line was seen in one acetabular zone in thirty-four hips (25 per cent) and in two acetabular zones in six hips (4 per cent). No hip had a radiolucent line in all three acetabular zones. The mean rate of linear wear of the polyethylene was 0.1 millimeter per year. There was no dissociation of the acetabular liner from the metal shell. Two hips (1 per cent) had asymptomatic osteolysis in the ischium and adjacent to the rim of the acetabular component; this was treated with grafting at the site of the lesion and exchange of the femoral head and the worn polyethylene liner. Five femoral components inserted without cement and one inserted with cement were revised because of loosening.
The data suggest that, at a mean of seven years, fixation of this porous-coated component was uniformly excellent. The low prevalence of radiolucent lines and the absence of loosening are probably related, in part, to the technique of implantation. The low prevalence of polyethylene wear and pelvic osteolysis is a notable improvement compared with the results of arthroplasty with other porous-coated acetabular components. Although the results of the present study are encouraging, longer follow-up is necessary to determine the prevalence of late loosening and osteolysis.

CiteULike Connotea Del.icio.us Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
K.-H. Koo, Y.-C. Ha, W. H. Jung, S.-R. Kim, J. J. Yoo, and H. J. Kim
Isolated Fracture of the Ceramic Head After Third-Generation Alumina-on-Alumina Total Hip Arthroplasty
J. Bone Joint Surg. Am.,
February 1, 2008;
90(2):
329 - 336.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. H. Weeden and W. G. Paprosky
Porous-Ingrowth Revision Acetabular Implants Secured with Peripheral Screws. A Minimum Twelve-Year Follow-up
J. Bone Joint Surg. Am.,
June 1, 2006;
88(6):
1266 - 1271.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Y.-G. Kim, S.-Y. Kim, S.-J. Kim, B.-C. Park, P.-T. Kim, and J.-C. Ihn
The use of cementless expansion acetabular component and an alumina-polyethylene bearing in total hip arthroplasty for osteonecrosis
J Bone Joint Surg Br,
June 1, 2005;
87-B(6):
776 - 780.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S.-Y. Kim, H.-S. Kyung, J.-C. Ihn, M.-R. Cho, K.-H. Koo, and C.-Y. Kim
Cementless Metasul Metal-on-Metal Total Hip Arthroplasty in Patients Less Than Fifty Years Old
J. Bone Joint Surg. Am.,
November 1, 2004;
86(11):
2475 - 2481.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. L. Gaffey, J. J. Callaghan, D. R. Pedersen, D. D. Goetz, P. M. Sullivan, and R. C. Johnston
Cementless Acetabular Fixation at Fifteen Years. A Comparison with the Same Surgeon's Results Following Acetabular Fixation with Cement
J. Bone Joint Surg. Am.,
February 1, 2004;
86(2):
257 - 261.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. A. Bojescul, J. S. Xenos, J. J. Callaghan, and C. G. Savory
Results of Porous-Coated Anatomic Total Hip Arthroplasty without Cement at Fifteen Years: A Concise Follow-up of a Previous Report
J. Bone Joint Surg. Am.,
May 28, 2003;
85(6):
1079 - 1083.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. F. Haft, A. D. Heiner, L. D. Dorr, T. D. Brown, and J. J. Callaghan
A Biomechanical Analysis of Polyethylene Liner Cementation into a Fixed Metal Acetabular Shell
J. Bone Joint Surg. Am.,
May 28, 2003;
85(6):
1100 - 1110.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. D. Crowther and P. F. Lachiewicz
Survival and Polyethylene Wear of Porous-Coated Acetabular Components in Patients Less than Fifty Years Old : Results at Nine to Fourteen Years
J. Bone Joint Surg. Am.,
May 1, 2002;
84(5):
729 - 735.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. A. RAGAB, M. J. KRAAY, and V. M. GOLDBERG
Clinical and Radiographic Outcomes of Total Hip Arthroplasty with Insertion of an Anatomically Designed Femoral Component without Cement for the Treatment of Primary Osteoarthritis. A Study with a Minimum of Six Years of Follow-up
J. Bone Joint Surg. Am.,
February 1, 1999;
81(2):
210 - 8.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
M. B. PETERSEN, I. H. POULSEN, J. THOMSEN, and S. SOLGAARD
The Hemispherical Harris-Galante Acetabular Cup, Inserted without Cement. The Results of an Eight to Eleven-Year Follow-up of One Hundred and Sixty-Eight Hips
J. Bone Joint Surg. Am.,
February 1, 1999;
81(2):
219 - 24.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
J. C. CLOHISY and W. H. HARRIS
Primary Hybrid Total Hip Replacement, Performed with Insertion of the Acetabular Component without Cement and a Precoat Femoral Component with Cement. An Average Ten-Year Follow-up Study
J. Bone Joint Surg. Am.,
February 1, 1999;
81(2):
247 - 55.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
J. C. CLOHISY and W. H. HARRIS
The Harris-Galante Porous-Coated Acetabular Component with Screw Fixation. An Average Ten-Year Follow-up Study
J. Bone Joint Surg. Am.,
January 1, 1999;
81(1):
66 - 73.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
A. G. URQUHART, D. D. D'LIMA, E. VENN-WATSON, C. W. COLWELL, and R. H. WALKER
Polyethylene Wear After Total Hip Arthroplasty: The Effect of a Modular Femoral Head with an Extended Flange-Reinforced Neck
J. Bone Joint Surg. Am.,
November 1, 1998;
80(11):
1641 - 7.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
J. J. CALLAGHAN
Instructional Course Lectures, The American Academy of Orthopaedic Surgeons - Periprosthetic Fractures of the Acetabulum during and following Total Hip Arthroplasty*{{dagger}}
J. Bone Joint Surg. Am.,
September 1, 1997;
79(9):
1416 - 21.
[Full Text]
|
 |
|
|