Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space
TP Schmalzried, M Jasty and WH Harris
Orthopaedic Biomechanics Laboratory, Massachusetts General Hospital, Boston 02114.
Thirty-four hips in which there had been prosthetic replacement were
selected for study because of the presence of linear (diffuse) or lytic
(localized) areas of periprosthetic bone loss. In all hips, there was
careful documentation of the anatomical location of the material that had
been obtained for histological analysis, and the specific purpose of the
removal of the tissue was for examination to determine the cause of the
resorption of bone. Specimens from twenty-three hips were retrieved during
an operation and from eleven hips, at autopsy. The area of bone loss was
linear only in sixteen hips, lytic only in thirteen, and both linear and
lytic in five. In all thirty-four hips, intracellular particulate debris
was found in the macrophages that were present in the area of bone
resorption. All thirty-four had intracellular particles of polyethylene,
many of which were less than one micrometer in size. Thirty-one hips had
extracellular particles of polyethylene as well. Twenty-two of the
thirty-four hips had intracellular metallic debris; in ten, metallic debris
was found extracellularly as well. Ten of the sixteen cemented specimens
had intracellular and extracellular polymethylmethacrylate debris. In the
mechanically stable prostheses--cemented and uncemented--polyethylene wear
debris was identified in areas of bone resorption far from the articular
surfaces. The number of macrophages in a microscopic field was directly
related to the amount of particulate polyethylene debris that was visible
by light microscopy. Although the gross radiographic appearances of linear
bone loss and lytic bone loss were different, the histological appearance
of the regions in which there was active bone resorption was similar.
Regardless of the radiographic appearance and anatomical origin of the
specimen, bone resorption was found to occur in association with
macrophages that were laden with polyethylene debris. In general, the
number of macrophages present had a direct relationship to the degree of
bone resorption that was seen. We believe that these findings indicate that
joint fluid penetrates far more extensively than previously thought, even
in a well fixed component, along the interface between the prosthesis and
bone and in the periprosthetic tissues; it is often more extensive than is
shown by arthrography. We therefore suggest the concept of the effective
joint space to include all periprosthetic regions that are accessible to
joint fluid and thus accessible to particulate debris.(ABSTRACT TRUNCATED
AT 400 WORDS)

CiteULike Connotea Del.icio.us Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
A. Marshall, M. D. Ries, and W. Paprosky
How prevalent are implant wear and osteolysis, and how has the scope of osteolysis changed since 2000?
J. Am. Acad. Ortho. Surg.,
July 1, 2008;
16(suppl_1):
S1 - S6.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. W. Nam, J. J. Yoo, Y. Lae Kim, Y.-M. Kim, M.-H. Lee, and H. J. Kim
Alumina-Debris-Induced Osteolysis in Contemporary Alumina-on-Alumina Total Hip Arthroplasty. A Case Report
J. Bone Joint Surg. Am.,
November 1, 2007;
89(11):
2499 - 2503.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R Dattani
Femoral osteolysis following total hip replacement
Postgrad. Med. J.,
May 1, 2007;
83(979):
312 - 316.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. C. Gordon, D. D. D'Lima, and C. W. Colwell Jr
Highly Cross-linked Polyethylene in Total Hip Arthroplasty.
J. Am. Acad. Ortho. Surg.,
September 1, 2006;
14(9):
511 - 523.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. S. Shanbhag
Use of bisphosphonates to improve the durability of total joint replacements.
J. Am. Acad. Ortho. Surg.,
April 1, 2006;
14(4):
215 - 225.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. A. Macheras, P. J. Papagelopoulos, K. Kateros, A. T. Kostakos, D. Baltas, and T. S. Karachalios
Radiological evaluation of the metal-bone interface of a porous tantalum monoblock acetabular component
J Bone Joint Surg Br,
March 1, 2006;
88-B(3):
304 - 309.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. W. McCalden, D. D. Naudie, X. Yuan, and R. B. Bourne
Radiographic Methods for the Assessment of Polyethylene Wear After Total Hip Arthroplasty
J. Bone Joint Surg. Am.,
October 1, 2005;
87(10):
2323 - 2334.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. Iesaka, W. L. Jaffe, C. M. Jones, and F. J. Kummer
The effects of fluid penetration and interfacial porosity on the fixation of cemented femoral components
J Bone Joint Surg Br,
September 1, 2005;
87-B(9):
1298 - 1302.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
H. M. Mayer
Total lumbar disc replacement
J Bone Joint Surg Br,
August 1, 2005;
87-B(8):
1029 - 1037.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F.H.R. De Man, W. Tigchelaar, R.K. Marti, C.J.F. Van Noorden, and H.M. Van der Vis
Effects of Mechanical Compression of a Fibrous Tissue Interface on Bone with or without High-Density Polyethylene Particles in a Rabbit Model of Prosthetic Loosening
J. Bone Joint Surg. Am.,
July 1, 2005;
87(7):
1522 - 1533.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N. Kitamura, D. D.R. Naudie, S. B. Leung, R. H. Hopper Jr., and C. A. Engh Sr.
Diagnostic Features of Pelvic Osteolysis on Computed Tomography: The Importance of Communication Pathways
J. Bone Joint Surg. Am.,
July 1, 2005;
87(7):
1542 - 1550.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Y.-H. Kim
Long-term results of the cementless porous-coated anatomic total hip prosthesis
J Bone Joint Surg Br,
May 1, 2005;
87-B(5):
623 - 627.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. J. Yoo, Y.-M. Kim, K. S. Yoon, K.-H. Koo, W. S. Song, and H. J. Kim
Alumina-on-Alumina Total Hip Arthroplasty. A Five-Year Minimum Follow-up Study
J. Bone Joint Surg. Am.,
March 1, 2005;
87(3):
530 - 535.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. A. Walde, A. M. Claus, V. Mohan, R. H. Hopper Jr., C. A. Engh Jr., and C. A. Engh Sr.
Advancement and Separation of Apex Hole Eliminators with Cementless Duraloc 100 Cups. A Report of Nineteen Cases
J. Bone Joint Surg. Am.,
October 1, 2004;
86(10):
2251 - 2256.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Y.-H. Kim and J.-S. Kim
Comparison of Anterior-Posterior-Glide and Rotating-Platform Low Contact Stress Mobile-Bearing Total Knee Arthroplasties
J. Bone Joint Surg. Am.,
June 1, 2004;
86(6):
1239 - 1247.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. Bradford, R. Kurland, M. Sankaran, H. Kim, L. A. Pruitt, and M. D. Ries
Early Failure Due to Osteolysis Associated with Contemporary Highly Cross-Linked Ultra-High Molecular Weight Polyethylene. A Case Report
J. Bone Joint Surg. Am.,
May 1, 2004;
86(5):
1051 - 1056.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. J. Yoo, H. J. Kim, and Y.-M. Kim
Damage of an Alumina-on-Alumina Bearing Surface from a Difficult Reduction of a Total Hip Arthroplasty. A Report of Three Cases
J. Bone Joint Surg. Am.,
February 1, 2004;
86(2):
376 - 378.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Y.-H. Kim, J.-S. Kim, S.-H. Oh, and J.-M. Kim
Comparison of Porous-Coated Titanium Femoral Stems with and without Hydroxyapatite Coating
J. Bone Joint Surg. Am.,
September 1, 2003;
85(9):
1682 - 1688.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. M. Claus, C. A. Engh, C. J. Sychterz, J. S. Xenos, K. F. Orishimo, and C. A. Engh
Radiographic Definition of Pelvic Osteolysis Following Total Hip Arthroplasty
J. Bone Joint Surg. Am.,
August 1, 2003;
85(8):
1519 - 1526.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. Heisel, M. Silva, and T. P. Schmalzried
Bearing Surface Options for Total Hip Replacement in Young Patients
J. Bone Joint Surg. Am.,
July 3, 2003;
85(7):
1366 - 1379.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Y.-H. Kim, S.-H. Oh, J.-S. Kim, and K.-H. Koo
Contemporary Total Hip Arthroplasty with and without Cement in Patients with Osteonecrosis of the Femoral Head
J. Bone Joint Surg. Am.,
March 31, 2003;
85(4):
675 - 681.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Y.-H. Kim, S.-H. Oh, and J.-S Kim
Primary Total Hip Arthroplasty with a Second-Generation Cementless Total Hip Prosthesis in Patients Younger Than Fifty Years of Age
J. Bone Joint Surg. Am.,
January 17, 2003;
85(1):
109 - 114.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
K. Takahashi, Y. Yasunaga, T. Hisatome, Y. Ikuta, and M. Ochi
Spontaneous Decrease in Focal Osteolysis in the Unrevised Femur After Revision of a Contralateral Bipolar Hemiarthroplasty: A Case Report
J. Bone Joint Surg. Am.,
January 17, 2003;
85(1):
122 - 125.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. D. Stulberg, R. L. Wixson, A. D. Adams, R. W. Hendrix, and J. B. Bernfield
Monitoring Pelvic Osteolysis Following Total Hip Replacement Surgery: An Algorithm for Surveillance
J. Bone Joint Surg. Am.,
November 25, 2002;
84
(suppl_2):
S116 - S122.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
Y.-H. Kim, H.-K. Kook, and J.-S. Kim
Total Hip Replacement with a Cementless Acetabular Component and a Cemented Femoral Component in Patients Younger than Fifty Years of Age
J. Bone Joint Surg. Am.,
May 1, 2002;
84(5):
770 - 774.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. Puri, R. L. Wixson, S. H. Stern, J. Kohli, R. W. Hendrix, and S. D. Stulberg
Use of Helical Computed Tomography for the Assessment of Acetabular Osteolysis After Total Hip Arthroplasty
J. Bone Joint Surg. Am.,
April 1, 2002;
84(4):
609 - 614.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. J. Millett, M. J. Allen, and M. P.G. Bostrom
Effects of Alendronate on Particle-Induced Osteolysis in a Rat Model
J. Bone Joint Surg. Am.,
February 1, 2002;
84(2):
236 - 249.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. T. Moskal, F. H. Shen, and T. E. Brown
The Fate of Stable Femoral Components Retained During Isolated Acetabular Revision : A Six-to-Twelve-Year Follow-up Study
J. Bone Joint Surg. Am.,
February 1, 2002;
84(2):
250 - 255.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
C. R. Clark and J. Cohen
Abscopal Wear Debris and Total Joint Arthroplasty: Cause for Concern?
J. Bone Joint Surg. Am.,
July 1, 2001;
83(7):
1110 - 1113.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. M. Wright, P. M. Pellicci, E. A. Salvati, B. Ghelman, M. M. Roberts, and J. L. Koh
Bone Density Adjacent to Press-Fit Acetabular Components : A Prospective Analysis with Quantitative Computed Tomography
J. Bone Joint Surg. Am.,
April 1, 2001;
83(4):
529 - 529.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
H. McKellop, F.-w. Shen, B. Lu, P. Campbell, and R. Salovey
Effect of Sterilization Method and Other Modifications on the Wear Resistance of Acetabular Cups Made of Ultra-High Molecular Weight Polyethylene : A Hip-Simulator Study
J. Bone Joint Surg. Am.,
December 1, 2000;
82(12):
1708 - 1708.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
C. R. Clark
Editorial. A Potential Concern in Total Joint Arthroplasty: Systemic Dissemination of Wear Debris
J. Bone Joint Surg. Am.,
April 1, 2000;
82(4):
455 - 455.
[Full Text]
|
 |
|

|
 |

|
 |
 
K. D. MOORE, R. L. BARRACK, C. J. SYCHTERZ, J. SAWHNEY, A. M. YANG, and C. A. ENGH
The Effect of Weight-Bearing on the Radiographic Measurement of the Position of the Femoral Head After Total Hip Arthroplasty
J. Bone Joint Surg. Am.,
January 1, 2000;
82(1):
62 - 9.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
B. K. VAUGHN, T. B. DAMERON, T. W. BAUER, Y. MOCHIDA, T. AKISUE, and R. W. EBERLE
Early Osteolysis Following Total Hip Arthroplasty with Use of a Hylamer Liner in Combination with a Modular Ceramic Femoral Head. A Case Report
J. Bone Joint Surg. Am.,
October 1, 1999;
81(10):
1446 - 9.
[Full Text]
|
 |
|

|
 |

|
 |
 
R. H. EMERSON, S. B. SANDERS, W. C. HEAD, and L. HIGGINS
Effect of Circumferential Plasma-Spray Porous Coating on the Rate of Femoral Osteolysis After Total Hip Arthroplasty
J. Bone Joint Surg. Am.,
September 1, 1999;
81(9):
1291 - 8.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
Y. NAKASHIMA, D.-H. SUN, M. C. D. TRINDADE, W. J. MALONEY, S. B. GOODMAN, D. J. SCHURMAN, and R. L. SMITH
Signaling Pathways for Tumor Necrosis Factor-{{alpha}} and Interleukin-6 Expression in Human Macrophages Exposed to Titanium-Alloy Particulate Debris in Vitro
J. Bone Joint Surg. Am.,
May 1, 1999;
81(5):
603 - 15.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
M. A. WIRTH, C. M. AGRAWAL, J. D. MABREY, D. D. DEAN, C. R. BLANCHARD, M. A. MILLER, and C. A. ROCKWOOD
Isolation and Characterization of Polyethylene Wear Debris Associated with Osteolysis Following Total Shoulder Arthroplasty
J. Bone Joint Surg. Am.,
January 1, 1999;
81(1):
29 - 37.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
T. P. SCHMALZRIED and J. J. CALLAGHAN
Current Concepts Review - Wear in Total Hip and Knee Replacements
J. Bone Joint Surg. Am.,
January 1, 1999;
81(1):
115 - 136.
[Full Text]
|
 |
|

|
 |

|
 |
 
R. B. BOURNE, C. H. RORABECK, M. SKUTEK, S. MIKKELSEN, M. WINEMAKER, and D. ROBERTSON
The Harris Design-2 Total Hip Replacement Fixed with So-Called Second-Generation Cementing Techniques. A Ten to Fifteen-Year Follow-up
J. Bone Joint Surg. Am.,
December 1, 1998;
80(12):
1775 - 80.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
M. W. HUNGERFORD, M. A. MONT, R. SCOTT, C. FIORE, D. S. HUNGERFORD, and K. A. KRACKOW
Surface Replacement Hemiarthroplasty for the Treatment of Osteonecrosis of the Femoral Head
J. Bone Joint Surg. Am.,
November 1, 1998;
80(11):
1656 - 64.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
J. J. JACOBS, A. K. SKIPOR, L. M. PATTERSON, N. J. HALLAB, W. G. PAPROSKY, J. BLACK, and J. O. GALANTE
Metal Release in Patients Who Have Had a Primary Total Hip Arthroplasty. A Prospective, Controlled, Longitudinal Study
J. Bone Joint Surg. Am.,
October 1, 1998;
80(10):
1447 - 58.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
T. R. YOON, S. M. ROWE, S. T. JUNG, K. J. SEON, and W. J. MALONEY
Osteolysis in Association with a Total Hip Arthroplasty with Ceramic Bearing Surfaces
J. Bone Joint Surg. Am.,
October 1, 1998;
80(10):
1459 - 67.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
M. T. MANLEY, W. N. CAPELLO, J. A. D'ANTONIO, A. A. EDIDIN, and R. G. T. GEESINK
Fixation of Acetabular Cups without Cement in Total Hip Arthroplasty. A Comparison of Three Different Implant Surfaces at a Minimum Duration of Follow-up of Five Years
J. Bone Joint Surg. Am.,
August 1, 1998;
80(8):
1175 - 85.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
C. F. BURT, K. L. GARVIN, E. T. OTTERBERG, and O. M. JARDON
A Femoral Component Inserted without Cement in Total Hip Arthroplasty. A Study of the Tri-Lock Component with an Average Ten-Year Duration of Follow-up
J. Bone Joint Surg. Am.,
July 1, 1998;
80(7):
952 - 60.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
S. TOKSVIG-LARSEN, L. RYD, and A. LINDSTRAND
Early Inducible Displacement of Tibial Components in Total Knee Prostheses Inserted with and without Cement. A Randomized Study with Roentgen Stereophotogrammetric Analysis
J. Bone Joint Surg. Am.,
January 1, 1998;
80(1):
83 - 9.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. D. BLOEBAUM, N. L. MIHALOPOULUS, J. W. JENSEN, and L. D. DORR
Postmortem Analysis of Bone Growth into Porous-Coated Acetabular Components
J. Bone Joint Surg. Am.,
July 1, 1997;
79(7):
1013 - 22.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
P. A. DEVANE, E. J. ROBINSON, R. B. BOURNE, C. H. RORABECK, N. N. NAYAK, and J. G. HORNE
Measurement of Polyethylene Wear in Acetabular Components Inserted with and without Cement. A Randomized Trial
J. Bone Joint Surg. Am.,
May 1, 1997;
79(5):
682 - 89.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
T. P. SCHMALZRIED, K. H. AKIZUKI, A. N. FEDENKO, and J. MIRRA
The Role of Access of Joint Fluid to Bone in Periarticular Osteolysis: A Report of Four Cases
J. Bone Joint Surg. Am.,
March 1, 1997;
79(3):
447 - 52.
[Full Text]
|
 |
|

|
 |

|
 |
 
J. YAO, G. CS-SZABÓ, J. J. JACOBS, K. E. KUETTNER, and T. T. GLANT
Suppression of Osteoblast Function by Titanium Particles*{{dagger}}
J. Bone Joint Surg. Am.,
January 1, 1997;
79(1):
107 - 12.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
C. J. SYCHTERZ, K. H. MOON, Y. HASHIMOTO, K. M. TEREFENKO, C. A. ENGH JR., and T. W. BAUER
Wear of Polyethylene Cups in Total Hip Arthroplasty. A Study of Specimens Retrieved Post Mortem
J. Bone Joint Surg. Am.,
August 1, 1996;
78(8):
1193 - 1200.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
B. D. MULLIKEN, R. B. BOURNE, C. H. RORABECK, and N. NAYAK
A Tapered Titanium Femoral Stem Inserted without Cement in a Total Hip Arthroplasty. Radiographic Evaluation and Stability
J. Bone Joint Surg. Am.,
August 1, 1996;
78(8):
1214 - 25.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
R. M. URBAN, J. J. JACOBS, D. R. SUMNER, C. L. PETERS, F. R. VOSS, and J. O. GALANTE
The Bone-Implant Interface of Femoral Stems with Non-Circumferential Porous Coating. A Study of Specimens Retrieved at Autopsy
J. Bone Joint Surg. Am.,
July 1, 1996;
78(7):
1068 - 81.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
K. G. SHEA, R. D. BLOEBAUM, J. M. AVENT, G. T. BIRK, and K. M. SAMUELSON
Analysis of Lymph Nodes for Polyethylene Particles in Patients Who Have Had a Primary Joint Replacement*{{dagger}}
J. Bone Joint Surg. Am.,
April 1, 1996;
78(4):
497 - 504.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
M. T. MAI, T. P. SCHMALZRIED, F. J. DOREY, P. A. CAMPBELL, and H. C. AMSTUTZ
The Contribution of Frictional Torque to Loosening at the Cement-Bone Interface in Tharies Hip Replacements
J. Bone Joint Surg. Am.,
April 1, 1996;
78(4):
505 - 11.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
E. B. BENZ, B. SHERBURNE, J. E. HAYEK, K. H. FALCHUK, C. B. SLEDGE, and M. SPECTOR
Lymphadenopathy Associated with Total Joint Prostheses. A Report of Two Cases and a Review of the Literature
J. Bone Joint Surg. Am.,
April 1, 1996;
78(4):
588 - 93.
[Full Text]
|
 |
|
|