|
The Journal of Bone and Joint Surgery, Vol 70, Issue 1 45-50, Copyright © 1988 by Journal of Bone and Joint Surgery, Inc
Telemetric force measurements across the hip after total arthroplasty
DT Davy, GM Kotzar, RH Brown, KG Heiple, VM Goldberg, KG Heiple, J Berilla and AH Burstein
Department of Mechanical and Aerospace Engineering, Case Western Reserve University, Cleveland, Ohio 44104.
A telemeterized total hip prosthesis was implanted in one patient and
force-data were obtained. Thirty-one days postoperatively, the magnitude of
the joint-contact force during double-limb stance was 1.0 times body
weight. During ipsilateral single-limb stance the joint-contact force was
2.1 times body weight, and during the stance phase of gait the peak force
typically was 2.6 to 2.8 times body weight, with the resultant force
located on the anterosuperior portion of the ball. During stair-climbing,
the force was 2.6 times body weight. At peak loads, the angle between the
resultant force and the axis of the neck was 30 to 35 degrees and that
between the resultant force and the plane of the prosthesis was 20 degrees.
During stair-climbing or straight-leg raising, the out-of-plane orientation
of the resultant force increased substantially. These data provide
information concerning the forces that must be sustained by prosthetic hip
joints during a number of common activities of daily living within the
first month after implantation. The results also provide insight into the
progression of early recovery and demonstrate the variety of forces that
are generated during this period.

CiteULike Connotea Del.icio.us Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
C. P. Skurla, G. E. Pluhar, D. J. Frankel, E. L. Egger, and S. P. James
Assessing the dog as a model for human total hip replacement: ANALYSIS OF 38 CANINE CEMENTED FEMORAL COMPONENTS RETRIEVED AT POST-MORTEM
J Bone Joint Surg Br,
January 1, 2005;
87-B(1):
120 - 127.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. N. Sangiorgio, E. Ebramzadeh, D. B. Longjohn, and L. D. Dorr
Effects of Dorsal Flanges on Fixation of a Cemented Total Hip Replacement Femoral Stem
J. Bone Joint Surg. Am.,
April 1, 2004;
86(4):
813 - 820.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. Unver, V. Karatosun, I. Gunal, and S. Angin
Comparison of two different rehabilitation programmes for thrust plate prosthesis: a randomized controlled study
Clinical Rehabilitation,
January 1, 2004;
18(1):
84 - 91.
[Abstract]
[PDF]
|
 |
|

|
 |

|
 |
 
E. Ebramzadeh, S. N. Sangiorgio, D. B. Longjohn, C. F. Buhari, and L. D. Dorr
Initial Stability of Cemented Femoral Stems as a Function of Surface Finish, Collar, and Stem Size
J. Bone Joint Surg. Am.,
January 1, 2004;
86(1):
106 - 115.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. D. Komistek, D. A. Dennis, J. A. Ochoa, B. D. Haas, and C. Hammill
In Vivo Comparison of Hip Separation After Metal-on-Metal or Metal-on-Polyethylene Total Hip Arthroplasty
J. Bone Joint Surg. Am.,
October 10, 2002;
84(10):
1836 - 1841.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
B. A. Morris, D. D. D'Lima, J. Slamin, N. Kovacevic, S. W. Arms, C. P. Townsend, and C. W. Colwell Jr.
e-Knee: Evolution of the Electronic Knee Prosthesis : Telemetry Technology Development
J. Bone Joint Surg. Am.,
October 1, 2001;
83(90021):
S62 - 66.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. J. BRUMBACK, T. R. TOAL, M. S. MURPHY-ZANE, V. P. NOVAK, and S. M. BELKOFF
Immediate Weight-Bearing After Treatment of a Comminuted Fracture of the Femoral Shaft with a Statically Locked Intramedullary Nail
J. Bone Joint Surg. Am.,
November 1, 1999;
81(11):
1538 - 44.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
D. A. LOCH, R. F. KYLE, J. E. BECHTOLD, M. KANE, K. ANDERSON, and R. E. SHERMAN
Forces Required to Initiate Sliding in Second-Generation Intramedullary Nails
J. Bone Joint Surg. Am.,
November 1, 1998;
80(11):
1626 - 31.
[Abstract]
[Full Text]
|
 |
|
|