The Journal of Bone and Joint Surgery, Vol 68, Issue 7 1035-1040, Copyright © 1986 by Journal of Bone and Joint Surgery, Inc
The influence of tibial-patellofemoral location on function of the knee in patients with the posterior stabilized condylar knee prosthesis
HE Figgie, VM Goldberg, KG Heiple, HS Moller and NH Gordon
Function of the knee and patellofemoral symptoms were correlated with the
position of the implant in 101 consecutive patients with 116 posterior
stabilized condylar knee prostheses. All of the patients were followed for
a minimum of two and a half years with sequential physical examinations,
radiographs, and functional evaluation of the knee. In sixteen knees (14
per cent), clicking or catching of the patella in terminal extension or
painless crepitation throughout the arc of flexion developed without
lowering the functional knee-evaluation score. Pain or mechanical problems,
or both, that lowered the functional knee-evaluation score occurred in
another fourteen knees (12 per cent), within the first postoperative year.
Of these fourteen, eight required revision solely for patellofemoral
complaints. Critical analysis of the tibial-patellofemoral mechanical axis
identified three surgical variables that were found to markedly affect the
functional result of the prosthesis: the distance from the center line of
the tibial prosthesis to the center line of the tibial plateau, a change in
the position of the joint line of the prosthesis relative to the hip and
ankle, and the patellar height, measured as the perpendicular distance from
the inferior pole of the patellar implant to the joint line of the
prosthesis. Functional knee scores, range of motion, patellofemoral pain or
mechanical symptoms, the need for revision, and the necessity of
manipulation could all be statistically significantly correlated with the
three independent variables. In addition, a range of neutral alignment was
developed.(ABSTRACT TRUNCATED AT 250 WORDS)

CiteULike Connotea Del.icio.us Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
J. Parvizi, T. D. Tarity, M. J. Steinbeck, R. G. Politi, A. Joshi, J. J. Purtill, and P. F. Sharkey
Management of Stiffness Following Total Knee Arthroplasty
J. Bone Joint Surg. Am.,
December 1, 2006;
88(suppl_4):
175 - 181.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Bhan, R. Malhotra, E. K. Kiran, S. Shukla, and M. Bijjawara
A Comparison of Fixed-Bearing and Mobile-Bearing Total Knee Arthroplasty at a Minimum Follow-up of 4.5 Years
J. Bone Joint Surg. Am.,
October 1, 2005;
87(10):
2290 - 2296.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. A. Rand
Extensor Mechanism Complications Following Total Knee Arthroplasty
J. Bone Joint Surg. Am.,
September 1, 2004;
86(9):
2062 - 2072.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. R. Bong and P. E. Di Cesare
Stiffness After Total Knee Arthroplasty
J. Am. Acad. Ortho. Surg.,
May 1, 2004;
12(3):
164 - 171.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. A. Parker, M. J. Dunbar, and C. H. Rorabeck
Extensor Mechanism Failure Associated With Total Knee Arthroplasty: Prevention and Management
J. Am. Acad. Ortho. Surg.,
July 1, 2003;
11(4):
238 - 247.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. C. Pollock, D. J. Ammeen, and G. A. Engh
Synovial Entrapment: A Complication of Posterior Stabilized Total Knee Arthroplasty
J. Bone Joint Surg. Am.,
December 9, 2002;
84(12):
2174 - 2178.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. D. Stulberg, P. Loan, and V. Sarin
Computer-Assisted Navigation in Total Knee Replacement: Results of an Initial Experience in Thirty-five Patients
J. Bone Joint Surg. Am.,
November 25, 2002;
84
(suppl_2):
S90 - S98.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. R. O'Rourke, J. J. Callaghan, D. D. Goetz, P. M. Sullivan, and R. C. Johnston
Osteolysis Associated with a Cemented tModular Posterior-Cruciate-Substituting Total Knee Design : Five to Eight-Year Follow-up
J. Bone Joint Surg. Am.,
August 12, 2002;
84(8):
1362 - 1371.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. L. PEARL and S. KURUTZ
Geometric Analysis of Commonly Used Prosthetic Systems for Proximal Humeral Replacement
J. Bone Joint Surg. Am.,
May 1, 1999;
81(5):
660 - 71.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
C. S. RANAWAT, C. P. LUESSENHOP, and J. A. RODRIGUEZ
The Press-Fit Condylar Modular Total Knee System. Four-to-Six-Year Results with a Posterior-Cruciate-Substituting Design
J. Bone Joint Surg. Am.,
March 1, 1997;
79(3):
342 - 8.
[Abstract]
[Full Text]
|
 |
|
|