This Article
Right arrow Full Text (PDF)
Right arrow Letters to the Editor: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when Letters to the Editor are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrowReprints and Permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by ROSE, R. M.
Right arrow Articles by RADIN, E. L.
Right arrow Search for Related Content
PubMed
Right arrow Articles by ROSE, R. M.
Right arrow Articles by RADIN, E. L.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Technorati  
What's this?
Journal of Bone and Joint Surgery, 1972;54:854-862.
© 1972 by The Journal of Bone and Joint Surgery, Inc


Corrosion-Accelerated Mechanical Failure of a Vitallium Nail-Plate

ROBERT M. ROSE SC.D.1, ALAN L. SCHILLER M.D.1, and ERIC L. RADIN M.D.1

1 From the Department of Metallurgy and Materials Science, Massachusetts institute of Technology, Cambridge, and the Departments of Orthopaedic Surgery and Pathology, Massachusetts General Hospital, Boston

Analysis of a broken Vitallium nail-plate device which had been used to fix an intertrochanteric fracture of the hip revealed that the nail was composed of cast Vitallium and the plate of wrought Vitallium. Little or no plastic flow accompanied the failure of the plate, suggesting gross overloading was not the prime cause of implant fracture, although it would have probably occurred in this case if failure had not interceded. The condition of the fracture surface indicated that essentially brittle failure had occurred, and that the fracture crack had initiated from the region of nail-plate contact, even though such a region is normally in a state of compressive stress. The scanning electron microscope was used to examine all surfaces, and revealed the presence of microscopic cracks nucleated by corrosion pits in the nail-plate contact area. The pits were probably due to electrochemical action aggravated by the dissimilar metal contact, stress concentration, and crevice corrosion effect of the nail-plate contact area inherent in the implant design. These findings suggest that wrought and cast Vitallium should not be used in combination, even when only compressive stresses on the implant are involved.