This Article
Right arrow Full Text (PDF)
Right arrow Letters to the Editor: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when Letters to the Editor are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrowReprints and Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by COHEN, J.
Right arrow Articles by WULFF, J.
Right arrow Search for Related Content
PubMed
Right arrow Articles by COHEN, J.
Right arrow Articles by WULFF, J.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Technorati  
What's this?
Journal of Bone and Joint Surgery, 1972;54:617-628.
© 1972 by The Journal of Bone and Joint Surgery, Inc


Clinical Failure Caused by Corrosion of a Vitallium Plate

CASE REPORT, NEW TESTING METHODS FOR CREVICE CORROSION, AND NEW TECHNIQUES FOR FASHIONING COBALT CHROMIUM ALLOYS TO BE USED IN SURGICAL IMPLANTS

JONATHAN COHEN M.D.1 and JOHN WULFF PH.D.1

1 From the Departments of Orthopaedic Surgery and Pathology, Children's Hospital Medical Center, Boston, and the Department of Metallurgy, Massachusetts institute of Technology, Cambridge

A case of clinical failure of a surgical implant (a Thornton plate and Smith-Petersen nail) caused by crevice corrosion of the plate is presented. In this implant which was composed of wrought and cast Vitallium it was the wrought Vitallium that showed the most significant changes. The metals involved and the corrosion products were analyzed and crevice corrosion studies were carried out using a testing method applicable to the metals from this case and to any metals used for similar applications.

Experiments performed in our laboratory are reviewed in which it was shown that superior resistance to crevice corrosion and superior strength and ductility can be imparted to the currently used cast cobalt-chromium alloy, Haynes Stellite 21, by alteration of fabrication methods, thereby giving it over-all superiority as a surgical implant material.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Technorati Technorati    What's this?


This article has been cited by other articles:


Home page
JBJSHome page
J. J. JACOBS, J. L. GILBERT, and R. M. URBAN
Current Concepts Review - Corrosion of Metal Orthopaedic Implants
J. Bone Joint Surg. Am., February 1, 1998; 80(2): 268 - 82.
[Full Text]