This Article
Right arrow Full Text (PDF)
Right arrow Letters to the Editor: Submit a response
Right arrow Alert me when this article is cited
Right arrow Alert me when Letters to the Editor are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Add to My File Cabinet
Right arrow Download to citation manager
Right arrowReprints and Permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Mauriello, G. E.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Mauriello, G. E.
Social Bookmarking
 Add to CiteULike   Add to Connotea   Add to Del.icio.us   Add to Technorati  
What's this?
Journal of Bone and Joint Surgery, 1968;50:524-534.
© 1968 by The Journal of Bone and Joint Surgery, Inc


Some Electronic Problems of Myoelectric Control of Powered Orthotic and Prosthetic Appliances

George E. Mauriello Ph.D.1

1 New York, N. Y.

In this paper the literature on the myoelectric signal is collected and reviewed for purposes of determining the electronic characteristics from which the optimum specifications for an amplifier used in a myoelectric control system are determined.

An average figure of 5,000 ohms at 10 cycles per second is given as the output impedance of tissue. Accepting a 1 per cent signal loss, the input impedance of the amplifier should be at least 500,000 ohms.

By integrating curves from the literature, it is fourid that the signal energy band is from 25 to 300 cycles per second as determined by the maximum signal/noise ratio in conjuction with a 1 per cent signal loss. The pass band of the amplifier should then be 25 to 300 cycles per second.

Various other electronic problems of myoelectric control are discussed such as type of amplifier, electrodes, alternating current noise pickup, signal variation, and tissue noise.


Add to CiteULike CiteULike   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Technorati Technorati    What's this?


This article has been cited by other articles:


Home page
CLIN PEDIATRHome page
M. Kadkhoda, S. M. K. Chung, and F. O. Adebonojo
Congenital Dislocation of the Hip Diagnostic Screening and Treatment: A Comparative Study of Two Populations of Infants and Children
Clinical Pediatrics, March 1, 1976; 15(3): 239 - 246.
[Abstract] [PDF]